Memory G57271Recalls & reported problems
Cook Incorporated · Memory Soft Wire Basket · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00827002572713), received between Jun 11, 2021 and Jun 11, 2021 — classified as 1 injury.
The most frequently reported specific device problem is “Material Split, Cut or Torn” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Material Split, Cut or Torn1
Patient problems
- Foreign Body In Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJun 11, 2021Material Split, Cut or Torn; Material Split, Cut or Torn
DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) , THE PHYSICIAN USED A MEMORY SOFT WIRE BASKET. DURING THE PROCEDURE, A STONE IN THE BILE DUCT NEEDED TO BE LITHOTRIPTER. THE HANDLE OF THE MSB-2X4-D WAS CUT OFF AND THE SHEATH WAS REMOVED. WIRES WERE PULLED INTO THE COOK TTCL-10 (LITHOTRIPTOR CABLE) AND WERE CLAMPED ONTO TO THE SLH-1(SOEHENDRA LITHOTRIPTOR HANDLE). TTCL-10 WAS CONNECTED TO THE SLH-1 VIA LUERLOCK CONNECTION. THE ASSISTANT BEGAN TO CRACK THE STONE WITH SINGLE AND SHORT TURNS OF THE SLH-1. AFTER EVERY TURNING MOVEMENT SHE WAITED A SECOND AND MADE THE NEXT. AFTER A WHILE, THE WIRE OF THE MSB CRACKED INSTEAD OF THE STONE. THE BASKET WAS STILL IN PLACE AND COULD NOT BE RETRIEVED BY THE ENDOSCOPIC DEPARTMENT. PATIENT WAS BROUGHT INTO THE OR THEATRE AND BASKET WAS REMOVED VIA SURGERY. CLARIFICATION WAS RECEIVED THAT THE WIRES DID NOT RIP AT THE TIP OR AT THE HANDLE. THEY RIPPED INSIDE OF THE CONQUEST TTCL-10 [BASKET DRIVE WIRES IN THE TTCL]. IT WAS ALSO REPORTED THAT THE SURGERY WAS NOT PLANNED [AS PART OF THE ORIGINAL TREATMENT]. THE PATIENT IS (B)(6) YEARS OLD. HER GENERAL CONSTITUTION IS BEYOND LOWER LIMITS. THE SURGERY HAS NOT BEEN A POSITIVE EFFECT ON TH
Patient problems: Foreign Body In Patient
MAUDE report 11982037 ↗
FDA clearance and classification
The FDA regulates product code LQR (dislodger, stone, biliary) under 21 CFR 876.5010 as a Class II device. 21 CFR 876.5010 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Memory G57271
Has the Memory G57271 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00827002572713), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Memory G57271?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Jun 11, 2021. That report lists device problem “Material Split, Cut or Torn” (1); patient problem “Foreign Body In Patient” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Memory G57271?
None of the reports are classified as deaths, 1 as an injury and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Memory G57271, and is it still sold?
The FDA's device database lists Cook Incorporated as the company responsible for it (brand Memory, model G57271). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other DISLODGER, STONE, BILIARY devices
- Stonetome M00535110Boston Scientific Corporation · DISLODGER, STONE, BILIARY74 reports
- Memory Basket G25149Cook Incorporated · DISLODGER, STONE, BILIARY74 reports
- Stonetome M00535150Boston Scientific Corporation · DISLODGER, STONE, BILIARY66 reports
- Cook G36250Cook Incorporated · DISLODGER, STONE, BILIARY1 recall10 reports
- NCompass G36251Cook Incorporated · DISLODGER, STONE, BILIARY1 recall8 reports
- Strange G09108Cook Incorporated · DISLODGER, STONE, BILIARY1 recall1 report
- Single Use Retrieval Nitinol Basket V FG-V451POlympus Medical Systems Corp. · DISLODGER, STONE, BILIARY31 reports
- SpyGlass™ Retrieval Basket M00546550Boston Scientific Corporation · DISLODGER, STONE, BILIARY25 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page