Blue Rhino G57717Recalls & reported problems
Cook Incorporated · Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Tray · Class 2
All models of Blue Rhino (16)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00827002577176), received between Oct 22, 2021 and Jan 20, 2022 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Gas/Air Leak” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
3 other recalls cover devices cleared under the same FDA submission (K193133); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Apr 7, 2023Class IOpen, Classified
There were reports from customers that the tracheostomy in the sets and trays is not making a secure connection with the 15mm cap and other 15mm circuit components and accessories. This resulted in an unsecure connection between the device connector and circuit components, cap, or accessories.
Z-1428-2023 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Gas/Air Leak2
- Material Rupture1
Patient problems
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJan 20, 2022Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED A (B)(6) MALE PATIENT HAD AN 8.5 FR TAPER GUARD TRACHEOSTOMY TUBE FROM A BLUE RHINO G2-MULTI PERCUTANEOUS TRACHEOSTOMY INTRODUCER TRAY PLACED WITH NO ISSUES. THE PATIENT HAD COVID-19 AND HAD BEEN ON A VENT FOR ABOUT 4 WEEKS. THE PATIENT IMMEDIATELY HAD AN AIR LEAK OF 50 ML, BUT THE LEAK WAS NOT AUDIBLE. THREE DAYS AFTER PLACEMENT, THE AIR LEAK WAS UP TO 200 ML AND AUDIBLE. THE PRESSURE IN THE CUFF MEASURED 30, AND NEITHER INFLATING UP TO 45 NOR CHANGING HEAD POSITION RESOLVED THE AIR LEAK. ON THE SAME DAY, THE PATIENT UNDERWENT AN ADDITIONAL PROCEDURE IN WHICH THE TUBE WAS REMOVED AND REPLACED WITH A DIFFERENT 8 FRENCH CUFFED TRACHEOSTOMY TUBE. THIS WAS DONE AT BEDSIDE AND REQUIRED SEDATION AND PARALYSIS. UPON REMOVAL, THE BALLOON WAS INTACT. CARE FOR THE TUBE WAS DESCRIBED AS SUCTIONING AS NEEDED. ADDITIONALLY, THE TRACH WAS SUTURED IN LOOSELY WITH 2 SUTURES AND HELD IN PLACE WITH TRACH STRAPS. THE PHYSICIAN ALSO PROVIDED THE FOLLOWING COMMENT REGARDING THE ISSUE: "SOMETIMES REPLACING THE 8.5 FR WITH AN 8 FR STILL RESULTS IN SOME AIR LEAK THAT IS NOT NORMALLY SEEN WHEN JUST STARTING PATIENTS OFF WITH AN 8 FR." NO OTHER ADVERSE EFFECTS WERE REPORTED. THE ADDITIONAL
Patient problems: Insufficient Information
MAUDE report 13308882 ↗MalfunctionOct 22, 2021Material Rupture; Gas/Air Leak; Material Rupture; Gas/Air Leak
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12682552 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K193133. One submission often covers many models, so check the product and lot details in each record.
- Z-1779-2026 · Mar 5, 2026 · Open, ClassifiedCOOK MEDICAL Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Set: Reference Part Number C-PTIS-100-HC-G-EU, Order Number G57685; Reference Part Number C-PTIS-100-HC-G-NA
- Z-1780-2026 · Mar 5, 2026 · Open, ClassifiedCOOK MEDICAL Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Tray: Reference Part Number C-PTISY-100-HC-G-NA, Order Number G57703; Reference Part Number C-PTISY-100-UNL-
- Z-1427-2023 · Apr 7, 2023 · Open, ClassifiedBlue Rhino G2-Multi Percutaneous Tracheostomy Introducer Set, Reference Part Numbers C-PTIS-100-HC-G-EU-FLEX7.5 (G57695), C-PTIS-100-HC-G-EU-FLEX8.5 (G57696), C-PTIS-100-HC-G-NA-FL
Other Tube tracheostomy and tube cuff devices
- Shiley 8CN85HCovidien LP · Tube tracheostomy and tube cuff1 recall2 deaths354 reports
- Shiley 6CN75HCovidien LP · Tube tracheostomy and tube cuff1 recall3 deaths240 reports
- Shiley 7CN80RCovidien LP · Tube tracheostomy and tube cuff2 recalls148 reports
- Bivona 67PFSS35Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall166 reports
- Bivona 670180Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall140 reports
- Bivona 670170Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall10 deaths132 reports
- Shiley 6CN75RCovidien LP · Tube tracheostomy and tube cuff1 recall1 death121 reports
- Bivona 67PFSS40Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall115 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page