Cook G58413Recalls & reported problems
Cook Incorporated · Arterial Pressure Monitoring Tray · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00827002584136), received between Nov 8, 2023 and May 6, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Obstruction of Flow” (3 reports).
1 other recall cover devices cleared under the same FDA submission (K180846); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Obstruction of Flow3
- Unraveled Material1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Insufficient Information1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 6, 2025Obstruction of Flow; Obstruction of Flow
IT WAS REPORTED THAT THE CATHETER INCLUDED IN THE ARTERIAL PRESSURE MONITORING TRAY OCCLUDED DURING AN UNKNOWN PROCEDURE FOR A PATIENT OF UNKNOWN AGE AND GENDER. UPON INSERTION OF THE CATHETER INTO THE FEMORAL ARTERY, THE 2.5 FR BY 5 CM CATHETER WAS FOUND TO BE DEFECTIVE AND COULD NOT BE FLUSHED. HOWEVER, AFTER USING A NEW DEVICE FROM A DIFFERENT KIT OF THE SAME SIZE AND LENGTH, THE LINE FLUSHED WITHOUT ANY ISSUES. IT IS UNKNOWN IF IMAGING WAS USED FOR DEVICE PLACEMENT. THE DEVICE WAS RETURNED FOR EVALUATION ON 29APR2025. UPON PRELIMINARY DFA FINDINGS AND ANALYSIS, IT WAS FOUND THAT THE CATHETER WAS FULL OF BIOLOGICAL MATTER AND COULD NOT BE FLUSHED BY A SYRINGE FILLED WITH TAP WATER. THE DEVICE WAS PLACED IN AN ULTRASOUND BATH FOR 15 MINUTES AND SOME CONGEALED BIOLOGICAL MATTER WAS CLEARED. THE CATHETER WAS THEN ABLE TO BE FLUSHED SUCCESSFULLY. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21970033 ↗MalfunctionApr 25, 2025Obstruction of Flow; Obstruction of Flow
IT WAS REPORTED THAT THE CATHETER INCLUDED IN THE ARTERIAL PRESSURE MONITORING TRAY OCCLUDED DURING AN UNKNOWN PROCEDURE FOR A PATIENT OF UNKNOWN AGE AND GENDER. THE USER WAS NOT ABLE TO FLUSH THE CATHETER. AT THIS TIME, NO ADVERSE EFFECTS OR ADDITIONAL PROCEDURES FOR THE PATIENT WERE REPORTED DUE TO THIS OCCURRENCE. ADDITIONAL INFORMATION REGARDING THE EVENT AND PATIENT OUTCOME HAS BEEN REQUESTED BUT IS CURRENTLY UNAVAILABLE.
Patient problems: Insufficient Information
MAUDE report 21911778 ↗MalfunctionDec 3, 2024Unraveled Material; Unraveled Material
IT WAS REPORTED THAT THE WIRE GUIDE INCLUDED IN THE ARTERIAL PRESSURE MONITORING TRAY UNRAVELED. THE PROVIDER PULLED OUT THE WIRE GUIDE TO TEST PRIOR TO PATIENT CONTACT AND FOUND THAT IT WAS UNCOILED AND UNUSABLE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ADDITIONAL INFORMATION REGARDING THE EVENT HAS BEEN REQUESTED BUT IS CURRENTLY UNAVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20835988 ↗MalfunctionNov 8, 2023Obstruction of Flow; Obstruction of Flow
NO ADDITIONAL INFORMATION REGARDING PATIENT AND/OR EVENT DETAILS HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18094603 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K180846. One submission often covers many models, so check the product and lot details in each record.
- Z-1778-2026 · Mar 5, 2026 · Open, ClassifiedCOOK MEDICAL Arterial Pressure Monitoring Tray, Reference Part Number APMY-2.5-15-0-2.5 , Order Number G58412; Reference Part Number APMY-4.0-21J-1-12, Order Number G58419
Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H74908526031Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page