Skip to content

Snaploc G59683Recalls & reported problems

Cook Incorporated · Snaploc wire guide locking device · Class 2

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00827002596832), received between Jun 10, 2024 and Aug 5, 2025 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Difficult or Delayed Positioning” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 3 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult or Delayed Positioning7
  • Device Slipped6
  • Accessory Incompatible1
  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 5, 2025Break; Break

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE, THE PHYSICIAN USED A COOK SNAPLOC WIRE GUIDE LOCKING DEVICE. IT WAS REPORTED THAT THE LOCKING DEVICE BROKE INTO PIECES WHILST THE ENDOSCOPIST WAS AT CRUCIAL POSITION WITH WIRES IN THE PANCREATIC DUCT. THE WIRES THEN HAD TO BE REMOVED TO REPLACE THE DEVICE. WHILST THE BILIARY PROCEDURE WAS SUCCESSFUL, IT WAS NOT POSSIBLE TO RE-CANNULATE THE PANCREATIC DUCT, WHICH LED TO A SIGNIFICANT PANCREATITIS RISK FOR THE PATIENT. [LOSS OF WIRE GUIDE ACCESS - SUBJECT OF THIS REPORT]. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22714504 ↗
  • MalfunctionMar 28, 2025Difficult or Delayed Positioning; Device Slipped; Difficult or Delayed Positioning; Device Slipped

    DURING ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURES, THE PHYSICIAN USED AN UNKNOWN NUMBER OF COOK SNAPLOC WIRE GUIDE LOCKING DEVICES. IT WAS REPORTED THAT THE DOCTOR NOTED THAT THE WIRE GUIDE WAS SLIPPING AND DIDN'T FEEL SECURE WITH BOTH THE HYDRA JAG 0.35 WIRE GUIDE AND DREAMWIRE 0.35 WIRE GUIDE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21721355 ↗
  • MalfunctionFeb 11, 2025Difficult or Delayed Positioning; Device Slipped; Difficult or Delayed Positioning; Device Slipped

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE, THE PHYSICIAN USED A COOK SNAPLOC WIRE GUIDE LOCKING DEVICE. IT WAS REPORTED THAT THE PHYSICIAN HAD DIFFICULTY IN ACCESSING CORRECT DUCT. AFTER 1ST BALLOON TRAWL LOST POSITIONING AND HAD TO RE CANNULATE [LOSS OF WIRE GUIDE ACCESS]. SEVERAL ATTEMPTS AT RE CANNULATING, HAD ISSUES GETTING IN CORRECT DUCT. FLUSHED WITH STANDARD SYRINGE. LOST POSITION OF THE WIRE THREE (3) TIMES WHEN LOCKED IN WITH SNAPLOC [LOSS OF WIRE GUIDE ACCESS]. CLICK WAS HEARD WHEN LOCKING IN WIRE. THE PHYSICIAN CHANGED THE LOCKING DEVICE MID-PROCEDURE TO WHAT THEY CURRENTLY USE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21359021 ↗
  • MalfunctionDec 13, 2024Difficult or Delayed Positioning; Device Slipped; Difficult or Delayed Positioning; Device Slipped

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE, THE PHYSICIAN USED A COOK SNAPLOC WIRE GUIDE LOCKING DEVICE. IT WAS REPORTED THAT THE WIRE GUIDE WIRE SLIPPED ON NUMEROUS OCCASIONS WHILST IT WAS LOCKED OFF IN THE WIRE LOCK DEVICE. THE PHYSICIAN DIDN'T LOSE ACCESS IN THE CBD [COMMON BILE DUCT] BUT HAD TO WATCH CAREFULLY. THE WIRE SLIPPED WHILST ON ITS OWN WHILST LOCKED AND WAITING FOR THE NURSE TO PREPARE THE NEXT DEVICE. THE WIRE GUIDE ALSO SLIPPED WHILST INSIDE THE SPHINCTEROTOME AND LOCKED USING THE WIRE LOCK. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20926869 ↗
  • MalfunctionNov 12, 2024Difficult or Delayed Positioning; Device Slipped; Difficult or Delayed Positioning; Device Slipped

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE, THE PHYSICIAN USED A COOK SNAPLOC WIRE GUIDE LOCKING DEVICE. IT WAS REPORTED THAT THE WIRE GUIDE SLIPPED WHEN REMOVING THE TOME - THE WIRE MOVED OUT WITH IT EVEN THOUGH IT WAS LOCKED. THE PHYSICIAN WAS USING BOSTON 0.035 STANDARD WIRE GUIDE. A VIDEO WAS PROVIDED OF THE WIRE BEING LOCKED ON BOTH SIDES OF THE SNAPLOC SHOWING THE WIRE SLIPPING WHILE LOCKED. AT THE TIME OF THIS REPORT, OUR ATTEMPTS TO COLLECT ADDITIONAL INFORMATION REGARDING PATIENT OUTCOME HAVE BEEN UNSUCCESSFUL. WHILE THE COMPLAINANT DID NOT SPECIFY IF THE PATIENT EXPERIENCED ANY ADVERSE EFFECTS OR REQUIRED ADDITIONAL MEDICAL PROCEDURES DUE TO THIS EVENT, THE INFORMATION ABLE TO BE COLLECTED DOES NOT REASONABLY SUGGEST THE PATIENT WAS ADVERSELY IMPACTED. ADDITIONAL ATTEMPTS TO GATHER THIS INFORMATION WILL BE MADE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20665201 ↗
  • MalfunctionNov 11, 2024Difficult or Delayed Positioning; Device Slipped; Difficult or Delayed Positioning; Device Slipped

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE, THE PHYSICIAN USED A COOK SNAPLOC WIRE GUIDE LOCKING DEVICE. IT WAS REPORTED THAT THE PHYSICIAN EXPERIENCED AN ISSUE WITH THE LOCK, THE WIRE KEPT SLIPPING. IT¿S IMPORTANT TO NOTE THAT THEY WERE USING THE OLYMPUS VISIGLIDE WIRE, WHICH IS THEIR STANDARD WIRE IN GALWAY, A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20653855 ↗
  • MalfunctionNov 11, 2024Difficult or Delayed Positioning; Device Slipped; Difficult or Delayed Positioning; Device Slipped

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE, THE PHYSICIAN USED A COOK SNAPLOC WIRE GUIDE LOCKING DEVICE. IT WAS REPORTED THAT THE PHYSICIAN EXPERIENCED AN ISSUE WITH THE LOCK¿THE WIRE KEPT SLIPPING. IT¿S IMPORTANT TO NOTE THAT THEY WERE USING THE OLYMPUS VISIGLIDE WIRE, WHICH IS THEIR STANDARD WIRE IN GALWAY, A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20653853 ↗
  • MalfunctionJun 10, 2024Accessory Incompatible; Difficult or Delayed Positioning; Accessory Incompatible; Difficult or Delayed Positioning

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE, THE PHYSICIAN USED A COOK SNAPLOCK WIRE GUIDE LOCKING DEVICE. IT WAS REPORTED THAT "THE WIRE KEPT 'BUCKLING-OUT' [WIRE LOOPING AT THE SNAPLOCK] DURING WIRE EXCHANGES. THE PHYSICIAN SWITCHED TO A COMPETITORS SPHINCTEROTOME AND WIRE GUIDE DEVICE TO COMPLETE THE INTENDED PROCEDURE SUCCESSFULLY." THE INFORMATION PROVIDED INDICATED THAT A COOK WIRE GUIDE WAS USED AND LATER REPLACED WITH A COMPETITORS WIRE GUIDE. THIS IS BEING INTERPRETED AS LOSS OF WIRE GUIDE ACCESS. [SUBJECT OF REPORT]. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19498500 ↗

Other Endoscope channel accessory devices

All Endoscope channel accessory devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.