Skip to content

Axium QC-6-20-3DRecalls & reported problems

Micro Therapeutics, Inc. · QC-6-20-3D BARE AXIUM 3D COIL · Class 2 · Implant

All models of Axium (28)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00836462010390), received between Nov 20, 2019 and Apr 3, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Separation Failure” (2 reports).

2 other recalls cover devices cleared under the same FDA submission (K081465); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Separation Failure2

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 3, 2024Separation Failure; Separation Failure

    MEDTRONIC RECEIVED A REPORT THAT THE SPRING COILS (LOT 225631674 AND LOT 226051945) WERE STRETCHED AND SUCCESSFULLY IMPLANTED AFTER REPLACEMENT. THE SPRING COIL (LOT 225062459) WAS NOT DETACHED SMOOTHLY, BUT IT WAS SUCCESSFULLY DETACHED AND IMPLANTED AFTER REPLACEMENT. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. THE DEVICE AND ANY ACCESSORIES WERE PREPARED AS INDICATED IN THE IFU. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF A CORONARY FISTULA. IT WAS NOTED THE PATIENT'S BLOOD FLOW WAS NORMAL AND VESSEL TORTUOSITY WAS NORMAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19032986 ↗
  • MalfunctionNov 20, 2019Separation Failure; Separation Failure

    MEDTRONIC RECEIVED INFORMATION THAT THERE WAS A NON-DETACHMENT ISSUE WITH THE AXIUM IMPLANT COIL. THE PHYSICIAN DID NOT ATTEMPT TO DETACH THE COIL WITH THE INSTANT DETACHER (ID). THEY ATTEMPTED TO DETACH THE COIL WITH THE MANUAL METHOD (BREAKING OF THE HYPO-TUBE, AN ALTERNATIVE DETACHMENT METHOD PRESENTED IN THE INSTRUCTIONS FOR USE) 3 (THREE) TIMES BUT WITHOUT SUCCESS. NO PATIENT INJURY WAS REPORTED AS A RESULT OF THE EVENT. THE PATIENT WAS UNDERGOING EMBOLIZATION TREATMENT OF AN UNRUPTURED FUSIFORM ANEURYSM WITH A 6.5MM DIAMETER. THE PATIENT'S BLOOD FLOW WAS NORMAL AND THE VASCULATURE WAS NORMAL IN TORTUOSITY. THE DOCTOR DIDN'T PREPARE THE INSTANT DETACHER (ID) BEFORE THE SURGERY. USED MICROCATHETER TO DELIVER THE SPRING COIL. A HEADWAY17 ( MICROVENTION) WAS USED. NO DAMAGE WAS NOTED TO THE PUSHWIRE. THE PATIENT WAS BEING TREATED FOR AN UNRUPTURED, FUSIFORM, BASILAR ARTERY, DISSECTING ANEURYSM. THE MAX DIAMETER WAS 6.5MM. THE BLOOD FLOW AND THE VESSEL ANATOMY WERE BOTH NORMAL. NO PATIENT INJURY OCCURRED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9348411 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K081465. One submission often covers many models, so check the product and lot details in each record.

  • Z-0873-2012 · Mar 1, 2010 · Terminated
    Axium Detachable Coil System QC-2-1-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular ab
  • Z-0874-2012 · Mar 1, 2010 · Terminated
    Axium Detachable Coil System QC-7-30-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular a

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →

Sources and updates