Skip to content

Axium QC-18-40-3DRecalls & reported problems

Micro Therapeutics, Inc. · QC-18-40-3D BARE AXIUM 3D COIL · Class 2 · Implant

All models of Axium (28)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00836462010512), received between Feb 28, 2020 and Nov 27, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K081465); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Detachment of Device or Device Component1
  • Difficult or Delayed Positioning1
  • Migration1
  • Physical Resistance/Sticking1
  • Separation Failure1
  • Stretched1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 27, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    ADDITIONAL INFORMATION RECEIVED CLARIFIED THAT THE FIRST COIL PREMATURELY DETACHED AND THE PUSHWIRE SEPARATED. CONTRADICTORILY, IT WAS STATED THAT THERE WAS NO KINK/DAMAGE OBSERVED TO THE PUSHWIRE OF EITHER COIL. THE COIL MANAGED TO BE PLACED, BUT IT WAS UNKNOWN IF IT WAS IN THE INTENDED PLACE. THE FIRST COIL HAD PUSHED OUT OF THE INTRODUCER SHEATH SMOOTHLY, BUT THE SECOND COIL WAS STUCK. THE INTRODUCER SHEATH WAS HELD VERTICALLY WHEN THE IMPLANT COILS WERE PULLED BACK INSIDE DURING HYDRATION. THE CAUSE OF THE EVENT WAS NOT DETERMINED. THE PATIENT HAD BEEN UNDERGOING TREATMENT FOR A BASILAR TIP ANEURYSM, WHICH WAS MORE THAN 20MM. THE PATIENT'S VESSEL TORTUOSITY WAS NORMAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20792501 ↗
  • MalfunctionJan 7, 2021Difficult or Delayed Positioning; Migration; Difficult or Delayed Positioning; Migration

    ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE COIL WAS NOT DETACHED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11136281 ↗
  • MalfunctionFeb 28, 2020Stretched; Separation Failure; Physical Resistance/Sticking; Stretched; Separation Failure; Physical Resistance/Sticking

    ADDITIONAL INFORMATION RECEIVED THAT THE PHYSICIAN ATTEMPTED TO DETACH THE COIL 3-4 TIMES WITH USING INSTANT DETACHER AND 1 TIME WITH USING MANUAL DETACHMENT METHOD BUT COIL DID NOT DETACH.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9769559 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K081465. One submission often covers many models, so check the product and lot details in each record.

  • Z-0873-2012 · Mar 1, 2010 · Terminated
    Axium Detachable Coil System QC-2-1-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular ab
  • Z-0874-2012 · Mar 1, 2010 · Terminated
    Axium Detachable Coil System QC-7-30-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular a

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page