TB 020024Recalls & reported problems
Oscor Inc. · Class 2
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00836559009924), received between Sep 4, 2018 and Sep 4, 2018 — classified as 1 injury.
The most frequently reported specific device problem is “Fracture” (1 report).
1 other recall cover devices cleared under the same FDA submission (K861403); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Fracture1
Patient problems
- Heart Failure1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjurySep 4, 2018Fracture; Fracture
THE CUSTOMER REPORTED THE PATIENT REPORTED FEELING WEAK; WE ESTABLISHED LACK OF STIMULATION FROM THE LEADS. PER THE EKG LINE ON AVBIII MONITOR, WITH REPLACEMENT RHYTHM IN NARROW QRS, FREQUENCY 40/MIN, PATIENT ABLE TO SUSTAIN CIRCULATION. DISTURBANCE IN CIRCULATION OCCURRED DIRECTLY AFTER ASSUMING SITTING POSITION. MULTIPLE TIMES, INCLUDING DURING ECHO CONTROL, THE ELECTRODE WAS REPOSITIONED WITHOUT ACHIEVING EFFECTIVE STIMULATION. LEAD FIXATION CHECKED AT GENERATOR AND BATTERY CHANGED WITH NO EFFECT. ASSISTANCE FROM CARDIOLOGY CLINIC ON-CALL DOCTOR; WE DISCOVERED A DEFECT IN THE LEADS PROVIDING IMPULSE FOR THE ELECTRODE. WE REMOVED THE DAMAGED FRAGMENT, RECONNECTED TO THE GENERATOR, OBTAINING EFFECTIVE STIMULATION CIRCA 80/MIN. PATIENT CONSCIOUS, ABLE TO SUSTAIN BREATHING AND CIRCULATION, NO CONSCIOUSNESS DISTURBANCES. INCREASING ISUPREL DOSAGE MANAGED TO ACHIEVED PATIENT'S OWN FREQUENCY AT AROUND 90/MIN. WE PERFORMED PRECAUTIONARY CHEST RTG. IN THE EVENING, FREQUENCY DOWN AGAIN TO 40/MIN, THE PATIENT REPORTED WEAKNESS. WE DISCOVERED A FRACTURE OF THE IMPULSE-TRANSFERRING LEAD. CARDIOLOGY CLINIC DOCTOR - REMOVED THE FAULTY FRAGMENT, ACHIEVING STIMULATION. ON (B)(6) 2014 WE CHANGED
Patient problems: Heart Failure
MAUDE report 7842766 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K861403. One submission often covers many models, so check the product and lot details in each record.
- Z-0391-2019 · Sep 26, 2018 · TerminatedTB Series Bipolar Pacing Leads, French Size 4F, 5F, 6F for Curve Type Straight, Atrial J, 60 degree Curve, and Right Heart. TB is packaged for sterile single use in a double tray
Other Electrode, Pacemaker, Temporary devices
- N/A 006173PC. R. Bard, Inc. · ELECTRODE, PACEMAKER, TEMPORARY161 reports
- SWAN-GANZ D200F7Edwards Lifesciences LLC · ELECTRODE, PACEMAKER, TEMPORARY1 recall0 reports
- ETHICON TPW32Ethicon, LLC · ELECTRODE, PACEMAKER, TEMPORARY1 death46 reports
- ETHICON TPW20Ethicon, LLC · ELECTRODE, PACEMAKER, TEMPORARY39 reports
- N/A 520007PC. R. Bard, Inc. · ELECTRODE, PACEMAKER, TEMPORARY33 reports
- Pacel™ 401763St. Jude Medical Cardiovascular Division · ELECTRODE, PACEMAKER, TEMPORARY21 reports
- Pacel™ 401771St. Jude Medical Cardiovascular Division · ELECTRODE, PACEMAKER, TEMPORARY1 death16 reports
- SWAN-GANZ PACING TD CATHETER D205F7Edwards Lifesciences LLC · Electrode, Pacemaker, Temporary13 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.