Skip to content

Vault C ACDF System 37-IN-0428Recalls & reported problems

Precision Spine, Inc. · Streamline Insertion Tip · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00840019916882), received between Jan 6, 2023 and Jan 6, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Difficult to Remove” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Difficult to Remove1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 6, 2023Difficult to Remove

    IT WAS REPORTED THAT A PROCEDURE WAS PERFORMED ON (B)(6) 2022, UTILIZING THE VAULT C ACDF SYSTEM. THE SURGEON INSERTED THE ASSEMBLED 14 X 12 X 07MM ANTERIOR PLATE (37-PA-4207) AND 14 X 12 X 07MM 7° POSTERIOR PEEK CAGE (37-CL-4207). WHEN HE ATTEMPTED TO REMOVED THE INSERTER/GUIDE TIP FROM THE PLATE CAGE ASSEMBLY, THE STREAMLINE INSERTION TIP (37-IN-0428) WOULD NOT RELEASE, RESULTING IN THE PLATE BEING PULLED OFF OF THE CAGE. THE DOCTOR THEN HAD TO REMOVE THE CAGE FROM THE DISC SPACE AND REPLACE WITH A NEW PLATE/CAGE ASSEMBLY. THERE WAS NO PATIENT INJURY BUT THERE WAS A FIVE (5) MINUTE DELAY TO THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16112007 ↗

Other Guide, Surgical, Instrument devices

All Guide, Surgical, Instrument devices →