Reform TI Pedicle Screw System 39-SB-5545Recalls & reported problems
Precision Spine, Inc. · Modular Bone Screw, 5.5mm x 45mm, Gold · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00840019931878), received between Oct 13, 2023 and Oct 13, 2023 — classified as 2 injuries.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Fracture2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryOct 13, 2023Break; Fracture; Break; Fracture
IT WAS REPORTED THAT THE PATIENT UNDERWENT A PROCEDURE TO ADDRESS AN L1 FRACTURE ON (B)(6) 2023. THE SURGEON PUT SCREWS AT T12 AND AT L2 AND DID A LAMINECTOMY AT L1, UTILIZING THE REFORM TI MODULAR CANNULATED PEDICLE SCREW SYSTEM. SUBSEQUENTLY, A REVISION PROCEDURE WAS PERFORMED ON SEPTEMBER 18, 2023, TO ADDRESS TWO BROKEN Ø5.5MM X 45MM REFORM TI MODULAR NON-CANNULATED SCREWS (B)(6). THE TOP TWO SCREWS IN THE CONSTRUCT WERE FOUND TO BE BROKEN AND WERE REMOVED WITH FORCEPS, THE BOTTOM TWO SCREW WERE ALSO REMOVED WITH A DRIVER AND T-HANDLE. NO NEW HARDWARE WAS IMPLANTED AS THE SURGEON FELT THERE WAS ADEQUATE FUSION ACROSS THE FRACTURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17931185 ↗InjuryOct 13, 2023Break; Fracture; Break; Fracture
IT WAS REPORTED THAT THE PATIENT UNDERWENT A PROCEDURE TO ADDRESS AN L1 FRACTURE ON (B)(6) 2023. THE SURGEON PUT SCREWS AT T12 AND AT L2 AND DID A LAMINECTOMY AT L1, UTILIZING THE REFORM TI MODULAR CANNULATED PEDICLE SCREW SYSTEM. SUBSEQUENTLY, A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2023, TO ADDRESS TWO BROKEN Ø5.5MM X 45MM REFORM TI MODULAR NON-CANNULATED SCREWS (B)(6). THE TOP TWO SCREWS IN THE CONSTRUCT WERE FOUND TO BE BROKEN AND WERE REMOVED WITH FORCEPS, THE BOTTOM TWO SCREW WERE ALSO REMOVED WITH A DRIVER AND T-HANDLE. NO NEW HARDWARE WAS IMPLANTED AS THE SURGEON FELT THERE WAS ADEQUATE FUSION ACROSS THE FRACTURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17931184 ↗
Other Thoracolumbosacral Pedicle Screw System devices
- EXPEDIUM VERSE 199721000Medos International Sàrl · Orthosis, spinal pedicle fixation, for degenerative disc disease387 reports
- Sequoia® 3301-1Zimmer Spine, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease184 reports
- CD HORIZON® Spinal System 8590855Medtronic Sofamor Danek, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease179 reports
- Vitality® 07.02011.001Biomet Spine LLC · Orthosis, spinal pedicle fixation, for degenerative disc disease3 recalls0 reports
- EXPEDIUM VERSE 199721001SMedos International Sàrl · Thoracolumbosacral pedicle screw system124 reports
- LineSider LS-N6SA7550Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls1 report
- LineSider LS-N6SA6545Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls0 reports
- EXPEDIUM VERSE 199721000SMedos International Sàrl · Thoracolumbosacral pedicle screw system97 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page