Reform Ti Modular 39-SK-6540Recalls & reported problems
Precision Spine, Inc. · Ø6.5 x 40mm Modular Bone Screw, Cannulated · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00840019933612), received between Dec 21, 2022 and Jul 20, 2023 — classified as 2 injuries.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJul 20, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT A PROCEDURE WAS PERFORMED ON (B)(6) 2023, BY DR. (B)(6) UTILIZING THE REFORM TI MODULAR PEDICLE SCREW SYSTEM. THE SCRUB TECH ASSEMBLED THE Ø6.5MM X 40MM REFORM TI MODULAR CANNULATED SCREW (39-SK-6540) AND THE REFORM TI MODULAR TOWER TULIP (39-MT-0403) ON THE BACK TABLE. FOLLOWING ASSEMBLY, SHE PUSHED AND PULLED ON THE TULIP PRIOR TO LOADING ON THE DRIVER. THE SURGEON THEN INSERTED THE K-WIRE WITH A MAZAR ROBOT ON THE RIGHT SIDE AND INSERTED SCREWS IN BOTH L4 & L5 AND THEN INSERTED LEFT SIDE SCREWS. THE ROD WAS PLACED ON THE LEFT SIDE, LOCKING CAPS WERE ASSEMBLED AND TORQUED. THE PHYSICIAN ASSISTANT WAS THEN USING THE GOLD SHAFT HEIGHT ADJUSTER AND WHEN THE HEIGHT ADJUSTER WAS PULLED OFF, THE TULIP HAD DISASSEMBLED AND WAS STILL IN THE ADJUSTER. THE PEDICLE SCREW WAS THEN REMOVED AND REPLACED WITH A NEW SCREW/TULIP ASSEMBLY OF THE SAME SIZE AND THE PROCEDURE WAS COMPLETED WITH NO FURTHER INCIDENT. THERE WAS NO PATIENT INJURY, BUT A 2 1/2 MINUTE DELAY TO THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17365404 ↗InjuryDec 21, 2022Use of Device Problem
IT WAS REPORTED THAT A ONE-LEVEL PROCEDURE WAS PERFORMED ON AN UNKNOWN DATE, UTILIZING THE REFORM MIS TI CT PEDICLE SCREW SYSTEM. IT WAS NOTED DURING FINAL RADIOGRAPHS TAKEN DURING THE PROCEDURE SHOW THAT ONE OF THE MODULAR POLYAXIAL TULIP ASSEMBLY REFORM MIS PEDICLE SCREW ((B)(4)) WAS NOT ASSEMBLED TO THE SCREW. THE SCREW WAS THEN REMOVED AND REPLACED WITH A NEW SCREW/TULIP ASSEMBLY. THERE WAS NO PATIENT INJURY BUT A 20-30 MINUTE DELAY RESULTED. THE DISASSEMBLED PRODUCT WAS DISCARDED IN THE OPERATING ROOM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16024223 ↗
Other Thoracolumbosacral Pedicle Screw System devices
- EXPEDIUM VERSE 199721000Medos International Sàrl · Orthosis, spinal pedicle fixation, for degenerative disc disease387 reports
- Sequoia® 3301-1Zimmer Spine, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease184 reports
- CD HORIZON® Spinal System 8590855Medtronic Sofamor Danek, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease179 reports
- Vitality® 07.02011.001Biomet Spine LLC · Orthosis, spinal pedicle fixation, for degenerative disc disease3 recalls0 reports
- EXPEDIUM VERSE 199721001SMedos International Sàrl · Thoracolumbosacral pedicle screw system124 reports
- LineSider LS-N6SA7550Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls1 report
- LineSider LS-N6SA6545Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls0 reports
- EXPEDIUM VERSE 199721000SMedos International Sàrl · Thoracolumbosacral pedicle screw system97 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page