Reform Ti Modular 39-SK-6545Recalls & reported problems

Precision Spine, Inc. · Ø6.5 x 45mm Modular Bone Screw, Cannulated · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00840019933629), received between May 17, 2023 and Mar 18, 2024 — classified as 3 injuries.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2
  • Detachment of Device or Device Component1

Patient problems

  • Implant Pain1
  • Inadequate Pain Relief1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 18, 2024Fracture; Fracture

    IT WAS REPORTED THAT A L4/L5/S2A1 PROCEDURE WAS PERFORMED IN AUSTRALIA ON (B)(6) 2022, UTILIZING THE REFORM TI MODULAR CANNULATED PEDICLE SCREW SYSTEM. SUBSEQUENTLY, THE PATIENT PRESENTED WITH PAIN AND RADIOGRAPHS SHOWED COMPRESSION OF THE CONSTRUCT AND A BROKEN SCREW IN L4, RIGHT SIDE. REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2024, DURING WHICH THE BROKEN 45MM REFORM TI MODULAR CANNULATED SCREW (B)(6) WAS CONFIRMED AND IT WAS NOTED THAT THE LEFT SIDE L4 SCREW SHOWED SIGNIFICANT LOOSENING. BOTH THE PROXIMAL AND DISTAL ENDS OF THE BROKEN SCREW WERE REMOVED AND REPLACED ALONG WITH THE LOOSE SCREW ON THE RIGHT SIDE. IN ADDITION, SCREWS WERE ALSO PLACED IN THE TWO LEVELS ABOVE WITH THE REMAINDER OF THE ORIGINAL CONSTRUCT LEFT IN PLACE. IMAGE PROVIDED SHOWS CAGE MIGRATION IN THE DISC SPACE JUST UNDER THE BROKEN SCREW.

    Patient problems: Implant Pain

    MAUDE report 18924238 ↗
  • InjuryOct 20, 2023Fracture; Fracture

    IT WAS REPORTED THAT THE PATIENT UNDERWENT INITIAL L5/S1 IMPLANTATION OF MODULAR CANNULATED SCREWS ON (B)(6) 2023, IN AUSTRALIA. SUBSEQUENTLY, THE PATIENT PRESENTED WITH PAIN, AND IT WAS IDENTIFIED THAT THE SCREWS IN S1 HAD BROKEN. REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2023 DURING WHICH IT WAS NOTED THAT LEFT S1 WAS BROKEN MID-SHAFT (DISTAL TO THE DIAMETER TRANSITION). THE PROXIMAL END OF 39-SK-6550 WAS EXPLANTED, THE DISTAL FRAGMENT REMAINS IN THE PATIENT. A 39-SK-7580 SHANK (WITH TULIP AND SET SCREW) WERE IMPLANTED AT THE ILIAC CREST FOR ADDITIONAL STRENGTH AND TO AVOID THE FRAGMENT. NO NEW SCREWS IMPLANTED AT S1 ON EITHER SIDE. THE RIGHT S1 SHANK WAS BROKEN AT APPROXIMATELY THE DIAMETER-TRANSITION. THE PROXIMAL END OF 39-SK-6545 WAS EXPLANTED AND THE DISTAL FRAGMENT REMAINS IN THE PATIENT. A 39-SK-7570 SHANK (WITH TULIP AND SET SCREW) WERE IMPLANTED AT THE ILIAC CREST ON THE RIGHT SIDE FOR ADDITIONAL STRENGTH AND TO AVOID THE FRAGMENT. THE L5 SCREWS WERE REMOVED AND REPLACED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 17976976 ↗
  • InjuryMay 17, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT A PROCEDURE WAS PERFORMED ON (B)(6) 2023, UTILIZING THE REFORM TI MIS PEDICLE SCREW SYSTEM. THE MODULAR POLYAXIAL TULIP ASSEMBLY (64-MT-0403) WAS POPPED ONTO THE Ø6.5MM X 45MM REFORM TI MODULAR CANNULATED SCREW (39-SK-6545) IN THE CADDY. THE TECH THEN PULLED ON THE BONE SCREW TO CHECK FOR PROPER ASSEMBLY. THE ASSEMBLED SCREW AND TULIP WAS THEN INSERTED WITH THE POLYAXIAL SCREWDRIVER AND WHEN THE SCREWDRIVER WAS REMOVED FROM THE SCREW THE TULIP HEAD POPPED OFF THE SCREW. AT THIS TIME A NEW K-WIRE WAS INSERTED AND THE BONE SCREW WAS BACKED OUT OF THE PEDICLE. ANOTHER SCREW AND TULIP WERE ASSEMBLED AND IMPLANTED WITH NO FURTHER ISSUE. THERE WAS A FIVE (5) MINUTE DELAY TO THE PROCEDURE DUE TO THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16949974 ↗

Other Appliance, Fixation, Spinal Interlaminal devices

All Appliance, Fixation, Spinal Interlaminal devices →