Skip to content

Reform Ti Modular 39-SK-6550Recalls & reported problems

Precision Spine, Inc. · Ø6.5 x 50mm Modular Bone Screw, Cannulated · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00840019933636), received between Oct 20, 2023 and Oct 20, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture1

Patient problems

  • Inadequate Pain Relief1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 20, 2023Fracture; Fracture

    IT WAS REPORTED THAT THE PATIENT UNDERWENT INITIAL L5/S1 IMPLANTATION OF MODULAR CANNULATED SCREWS O (B)(6) 2023, IN AUSTRALIA. SUBSEQUENTLY, THE PATIENT PRESENTED WITH PAIN, AND IT WAS IDENTIFIED THAT THE SCREWS IN S1 HAD BROKEN. REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2023, DURING WHICH IT WAS NOTED THAT LEFT S1 WAS BROKEN MID-SHAFT (DISTAL TO THE DIAMETER TRANSITION). THE PROXIMAL END OF 39-SK-6550 WAS EXPLANTED, THE DISTAL FRAGMENT REMAINS IN THE PATIENT. A 39-SK-7580 SHANK (WITH TULIP AND SET SCREW) WERE IMPLANTED AT THE ILIAC CREST FOR ADDITIONAL STRENGTH AND TO AVOID THE FRAGMENT. THE PROXIMAL END OF 39-SK-6545 WAS EXPLANTED AND THE DISTAL FRAGMENT REMAINS IN THE PATIENT. A 39-SK-7570 SHANK (WITH TULIP AND SET SCREW) WERE IMPLANTED AT THE ILIAC CREST ON THE RIGHT SIDE FOR ADDITIONAL STRENGTH AND TO AVOID THE FRAGMENT. THE L5 SCREWS WERE REMOVED AND REPLACED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 17976975 ↗

Other Thoracolumbosacral Pedicle Screw System devices

All Thoracolumbosacral Pedicle Screw System devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page