Skip to content

Reform Ti Modular 39-SK-7550Recalls & reported problems

Precision Spine, Inc. · Ø7.5 x 50mm Modular Bone Screw, Cannulated · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00840019933704), received between Apr 13, 2022 and Apr 13, 2022 — classified as 1 injury.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 13, 2022Break

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A PROCEDURE PERFORMED ON (B)(6) 2021, UTILIZING THE REFORM TI MIS CT PEDICLE SCREW SYSTEM. A REVISION WAS PERFORMED ON (B)(6) 2022 TO ADDRESS A BROKEN Ø7.5MM X 50MM REFORM TI MODULAR CANNULATED SCREW (39-SK-7550) THAT WAS IDENTIFIED DURING ROUTINE FOLLOW-UP X-RAYS. THE BREAK OCCURRED AT THE NECK JUST BELOW THE SCREW HEAD. THE BROKEN SCREW WAS REMOVED AND REPLACED WITH A 9.5MM X 50MM, AND THE SCREW FROM THE ADJACENT LEVEL WAS REMOVED AND REPLACED WITH AN 8.5MM X 50MM SCREW. IT WAS NOTED THAT FUSION HAD NOT BEEN ACHIEVED AND THE PATIENT WAS COMPLIANT.

    Patient problems: Insufficient Information

    MAUDE report 14103508 ↗

Other Appliance, Fixation, Spinal Interlaminal devices

All Appliance, Fixation, Spinal Interlaminal devices →