Skip to content

Globe Scientific 51635Recalls & reported problems

Globe Scientific Inc. · Capillary tube, heparinized for Medica Corp. 10x100/box · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00840095607575), received between Jan 29, 2026 and Jan 29, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Defective Component” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 31, 2025Open, Classified

    A systematic positive bias of up to 20% was identified across the reportable potassium range, resulting in false elevation results of patient samples when using Medica s capillary tubes.

    Z-1370-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Defective Component1

Patient problems

  • Arrhythmia1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 29, 2026Defective Component

    GLOBE SCIENTIFIC CUSTOMER MEDICA CORPORATION FOR THE INVOLVED PRODUCT (CAPILLARY TUBE COATED WITH A HEPARIN / SALT FORMULATION) REPORTED THE FOLLOWING TO GLOBE SCIENTIFIC, THE IMPORTER OF THE PRODUCT: MEDICA OBSERVED ELEVATED POTASSIUM MEASUREMENT BIAS BY UP TO 20% WHEN USING THE CAPILLARY TUBES BY GLOBE SCIENTIFIC (ITEM 51635) ON THE MEDICA EASYSTAT ANALYZER. THIS CONDITION APPEARS IN ALL LOTS OF CAPILLARY TUBES SHIPPED TO MEDICA IN 2025 AND MAYBE AS FAR BACK AS 2022. PRELIMINARY INVESTIGATION AT MEDICA INDICATES THAT THIS CONDITION IS CAUSED BY THE COATING INSIDE THE GLASS CAPILLARY TUBE. MEDICA IS THE EXCLUSIVE CUSTOMER GLOBE SCIENTIFIC SELLS THIS SPECIFIC PRODUCT SKU TO IN THE MARKETPLACE BASED ON THE PROPRIETARY NATURE OF THE RECIPE USED FOR THE HEPARIN / SALT FORMULATION INVOLVED WITH THE PRODUCTION OF THE PRODUCT.

    Patient problems: Arrhythmia

    MAUDE report 24215772 ↗

Other Tube, Collection, Capillary Blood devices

All Tube, Collection, Capillary Blood devices →

Sources and updates