Skip to content

Medica Capillary Tubes recallZ-1370-2026 · Medica Corporation · initiated Dec 31, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medica Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A systematic positive bias of up to 20% was identified across the reportable potassium range, resulting in false elevation results of patient samples when using Medica s capillary tubes.

FDA root cause category: Nonconforming Material/Component

What the company is doing

An URGENT FIELD SAFETY NOTICE was sent to consignees on 01/09/25 informing them that potassium results obtained using affected devices in capillary mode may be falsely elevated. The FSN instructed consignees not to report potassium results obtained in capillary mode using affected lots on the Easy Stat Analyzer. The FSN is to be forwarded to end users and to where product was further distributed. Consignees are to contact their local representative with any questions.

Recall details

Recall number
Z-1370-2026
Status
Open, Classified
Initiated by the company
Dec 31, 2025
Posted by the FDA
Feb 13, 2026
Recalling firm
Medica Corporation
Quantity
1,578 units
Product code
CEM
FDA submissions
K021515
Distribution
International distribution to the countries of Russia, Bulgaria, Kazakhstan, Ukraine, Pakistan, Belarus.

Product description

Medica Capillary Tubes; EasyBloodGas/EasyStat Capillary Tube Kit.

Code information (models, lots, serial numbers)

Model No. 7303, UDI-DI: 00840095607575, Lot No. 25XXX.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates