Skip to content

Viant Medical, LLC 511172Recalls & reported problems

Viant Medical · Metal Handle Offset Cup Impactor · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
15
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 00840096400045), received between Jan 28, 2020 and Feb 13, 2026 — classified as 1 injury, 14 malfunctions.

The most frequently reported specific device problem is “Mechanical Jam” (11 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Jam11
  • Failure to Disconnect6
  • Mechanical Problem5
  • Break4
  • Damaged Thread1
  • Device Slipped1
  • Fracture1
  • Loss of or Failure to Bond1

Patient problems

  • No Clinical Signs, Symptoms or Conditions11
  • Insufficient Information3
  • No Consequences Or Impact To Patient1
  • Pain1

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 13, 2026Loss of or Failure to Bond; Break; Material Integrity Problem; No Apparent Adverse Event

    IT WAS REPORTED DURING A TOTAL HIP REPLACEMENT THAT THE WELD ON THE HANDLE HAS BROKEN CAUSING THE HANDLE TO UNSCREW AND NOT BE FUNCTIONAL. NO CONSEQUENCES OR IMPACT TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Pain

    MAUDE report 24339034 ↗
  • MalfunctionAug 7, 2025Mechanical Problem; Device Slipped; Unintended Movement; Mechanical Problem; Device Slipped; Unintended Movement

    IT WAS REPORTED DURING A THA SURGERY FOR OA ON HIP JOINT ON AN UNKNOWN PATIENT THAT A CUP LOCKING PROBLEM OCCURRED. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH MORE THAN 30 MINUTES SURGICAL DELAY. NO CONSEQUENCES OR IMPACT TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22741808 ↗
  • MalfunctionMar 12, 2025Mechanical Problem; Mechanical Jam; Mechanical Problem; Mechanical Jam

    IT WAS REPORTED DURING AN UNKNOWN PROCEDURE ON A MALE PATIENT THAT THE KNOB ON CUP INSERTER IS STIFF WHICH MAKING THE RELEASE OF THE CUP DIFFICULT AFTER IMPACTING. THERE WAS NO SURGICAL DELAY AND NO CONSEQUENCES OR IMPACT TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21584661 ↗
  • MalfunctionNov 1, 2024Mechanical Problem; Mechanical Jam; Mechanical Problem; Mechanical Jam

    IT WAS REPORTED DURING AN UNKNOWN PROCEDURE ON A MALE PATIENT THAT THE HANDLE WOULDN'T DISENGAGE FROM THE CUP ONCE THE CUP WAS PLACED. NO KNOWN DELAY AND CONSEQUENCES OR IMPACT TO PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 20592524 ↗
  • MalfunctionOct 4, 2024Mechanical Problem; Mechanical Jam; Mechanical Problem; Mechanical Jam

    IT WAS REPORTED DURING AN UNKNOWN PROCEDURE ON A MALE PATIENT THAT THE PIECE THAT TIGHTENS ONTO THE OFFSET IMPACTOR TO CONNECT THE CUP WOULD NOT COME LOOSE TO TAKE THE HANDLE OFF THE CUP. NO KNOWN DELAY AND CONSEQUENCES OR IMPACT TO PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 20375212 ↗
  • MalfunctionMar 28, 2024Mechanical Problem; Mechanical Jam; Damaged Thread; Mechanical Problem; Mechanical Jam; Damaged Thread

    IT WAS REPORTED DURING A HIP SURGERY ON AN UNKNOWN PATIENT THAT THE OFFSET CUP IMPACTOR THREADS ARE DAMAGED. IT WILL NOT RELEASE THE CUP AFTER IMPACTION / INSERTION AND RISKS THE CUP MOVING FROM OPTIMAL POSITIONING. THE SURGEON TOOK THEIR TIME TO REMOVE THE CUP FROM THE IMPACTOR RESULTING IN A FIVE (5) MINUTE DELAY IN SURGERY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO CONSEQUENCES OR IMPACT TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18999357 ↗
  • MalfunctionMar 14, 2024Break; Material Deformation; Material Twisted/Bent; Mechanical Jam; Break; Material Deformation; Material Twisted/Bent; Mechanical Jam

    IT WAS REPORTED DURING A THA (TOTAL HIP ARTHROPLASTY) FOR OA (OSTEOARTHRITIS) OF THE HIP JOINT ON A 74 YEAR OLD FEMALE PATIENT THAT THE SURGEON ATTACHED THE CUP TO THE PRODUCT AND STRUCK IT. WHEN THE SCREW WAS TURNED TO REMOVE THE CUP FROM THE PRODUCT, THE SCREW DID NOT MOVE. SINCE THE SCREW WOULD NOT MOVE, THE SURGEON PULLED THE CUP OUT OF ACETABULUM TO CHECK IT AND FOUND THAT THE METAL AT THE JOINT WAS DEFORMED. THE DEFORMED AREA WAS SPREAD OUTWARD AND COULD NOT BE USED. ANOTHER SIMILAR PRODUCT (STRAIGHT IMPACTOR) WAS USED TO COMPLETE THE PROCEDURE SUCCESSFULLY WITH SURGERY TIME EXTENDED BY ABOUT 10 MINUTES. NO CONSEQUENCES OR IMPACT TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18906249 ↗
  • MalfunctionSep 19, 2023Failure to Disconnect; Mechanical Jam; Failure to Disconnect; Mechanical Jam

    IT WAS REPORTED DURING AN UNKNOWN PATIENT PROCEDURE THAT THE IMPACTOR WAS DIFFICULT TO REMOVED/UNSCREW FROM THE IMPLANT. NO CONSEQUENCES OR IMPACT TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17775756 ↗
  • MalfunctionSep 19, 2023Break; Fracture; Break; Fracture

    IT WAS REPORTED DURING A TOTAL HIP REPLACEMENT THAT WHEN IMPACTING THE BACK OF THE HANDLE, THE DRIVE SHAFT THAT SCREWS INTO THE CUP SHEARED. A SIMILAR DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH THE SAME IMPLANT SINCE IT WAS NOT SEATED IN THE PATIENT YET. THERE WAS A DELAY OF A COUPLE MINUTES TO GET AND OPEN ANOTHER TRAY, BUT THERE WERE NO CONSEQUENCES OR IMPACT TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17775751 ↗
  • MalfunctionJul 5, 2023Failure to Disconnect; Mechanical Jam; Failure to Disconnect; Mechanical Jam

    IT WAS REPORTED DURING A TOTAL HIP REPLACEMENT THAT THE OFFSET CUP INSERTER WAS DIFFICULT TO REMOVE FROM THE ACETABULAR COMPONENT AFTER IMPACTION. IT IS UNKNOWN IF THERE WERE ANY HEALTH CONSEQUENCES OR IMPACT TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17260159 ↗
  • MalfunctionJul 5, 2023Failure to Disconnect; Mechanical Jam; Failure to Disconnect; Mechanical Jam

    IT WAS REPORTED DURING A TOTAL HIP REPLACEMENT THAT IT IS DIFFICULT TO REMOVE CUP FROM OFFSET IMPACTOR AFTER IMPLANTING. IT IS UNKNOWN IF THERE WERE ANY HEALTH CONSEQUENCES OR IMPACT TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17259688 ↗
  • MalfunctionJul 5, 2023Failure to Disconnect; Mechanical Jam; Failure to Disconnect; Mechanical Jam

    IT WAS REPORTED DURING A TOTAL HIP REPLACEMENT THAT THE ACETABULAR COMPONENT DIFFICULT TO REMOVE AFTER IMPACTION. IT IS UNKNOWN IF THERE WERE ANY HEALTH CONSEQUENCES OR IMPACT TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17256419 ↗
  • MalfunctionJun 23, 2023Failure to Disconnect; Mechanical Jam; Failure to Disconnect; Mechanical Jam

    IT WAS REPORTED DURING TOTAL HIP REPLACEMENT ON AN UNKNOWN PATIENT PROCEDURE THAT THE OFFSET CUP INSERTER GETS STUCK ON THE REAL IMPLANT AND IT'S VERY DIFFICULT TO UNSCREW. THERE WAS A SURGICAL DELAY OF 10 SECONDS TO GET PLIERS TO HELP LOOSEN THE DEVICE DUE TO THE REPORTED MALFUNCTION. THERE WERE NO HEALTH CONSEQUENCES OR IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17195244 ↗
  • MalfunctionJun 15, 2023Failure to Disconnect; Mechanical Jam; Failure to Disconnect; Mechanical Jam

    IT WAS REPORTED DURING AN UNKNOWN PROCEDURE THAT THE CURVED CUP INSERTER IS CROSS THREADED. THE CUP WOULD NOT DISENGAGE FROM THE INSERTER. THE CUP WAS PULLED OUT AND REMOVED, AND PUT ON A STRAIGHT HANDLE FOR INSERTION. IT IS UNKNOWN IF THERE WERE ANY HEALTH CONSEQUENCES OR IMPACT.

    Patient problems: Insufficient Information

    MAUDE report 17140085 ↗
  • MalfunctionJan 28, 2020Break; Break

    IT WAS REPORTED THAT THE LATCH BROKE ON OFFSET CUP IMPACT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9636362 ↗

Other Impactor devices

All Impactor devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page