ATTUNE 2544-01-006Recalls & reported problems
Depuy (Ireland) · ATTUNE FEMORAL IMPACTOR · Class 1
All models of ATTUNE (1471)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This DePuy ATTUNE femoral impactor has 2,544 adverse event reports, with no deaths and 3 injuries. Most reports describe the tool breaking, cracking or splitting, sometimes while it is used to impact the implant. One recall is closed, starting in November 2014, over impactors that broke into small pieces, which could remain in a patient if the break went unnoticed during surgery.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 2,528
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2,528 adverse event reports that list this exact device (DI 10603295130222), received between Jun 23, 2017 and Sep 25, 2026 — classified as 3 injuries, 2,525 malfunctions.
The most frequently reported specific device problem is “Break” (2,195 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Nov 10, 2014Class IITerminated
ATTUNE INTUITION Impactors (cat. nos. 254401003, 254401004, and 254401006) have broken and produced small pieces. Should a fracture occur and not be observed during surgery, there is the potential for these small fractured pieces of the instrument to be left in the patient.
Z-0905-2015 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Break2,195
- Crack524
- Naturally Worn56
- No Apparent Adverse Event25
- Adverse Event Without Identified Device or Use Problem16
- Scratched Material16
- Component Missing12
- Contamination /Decontamination Problem11
Patient problems
- No Clinical Signs, Symptoms or Conditions1,422
- No Consequences Or Impact To Patient895
- Insufficient Information124
- No Information86
- No Code Available32
- Not Applicable18
- No Known Impact Or Consequence To Patient3
- No Patient Involvement1
Recent adverse event reports
The 25 most recent of 2,528 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 25, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50903849 ↗MalfunctionSep 18, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50828630 ↗MalfunctionSep 16, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50785806 ↗MalfunctionJul 13, 2026Break
IT WAS REPORTED THAT DURING A TOTAL KNEE ARTHROPLASTY (TKA), THE FEMORAL IMPACTOR TIP BROKE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25801823 ↗MalfunctionJul 10, 2026Break; Crack
IT WAS REPORTED THAT THE RP IMPACTOR TIP CRACKED/BROKE. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25784354 ↗MalfunctionJul 1, 2026Break
IT WAS REPORTED THAT THE IMPACTED PRODUCTS WERE FOUND FRACTURED DURING ASSEMBLY. THE EVENT DID NOT HAPPENED IN SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25682910 ↗MalfunctionJun 25, 2026Break
BROKEN FEMORAL IMPACTOR REPORTED. PATIENT STATUS/ OUTCOME / CONSEQUENCES. NO PATIENT INVOLVEMENT, WAS OTHER MEDICAL INTERVENTION (E.G. X-RAYS, ADDITIONAL PROCEDURES, PRESCRIPTIONS, OTC, REVISION) REQUIRED: UNKNOWN,
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25628312 ↗MalfunctionJun 22, 2026Break
IT WAS REPORTED THAT FEM IMPACTORS BROKE. PIN DRIVERS ARE STRIPPED. PRESS FIT TIBIA DRILL DULL. 3-5 RING IS CRACKING. FEM DRILL IS DULL. BOX OSTEOTOME ARE BENDING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25578839 ↗MalfunctionJun 22, 2026Break
IT WAS REPORTED THAT FEM IMPACTORS BROKE. PIN DRIVERS ARE STRIPPED. PRESS FIT TIBIA DRILL DULL. 3-5 RING IS CRACKING. FEM DRILL IS DULL. BOX OSTEOTOME ARE BENDING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25578828 ↗MalfunctionJun 22, 2026Break
IT WAS REPORTED THAT FEM IMPACTORS BROKE. PIN DRIVERS ARE STRIPPED. PRESS FIT TIBIA DRILL DULL. 3-5 RING IS CRACKING. FEM DRILL IS DULL. BOX OSTEOTOME ARE BENDING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25578804 ↗MalfunctionJun 22, 2026Break
IT WAS REPORTED THAT FEM IMPACTORS BROKE. PIN DRIVERS ARE STRIPPED. PRESS FIT TIBIA DRILL DULL. 3-5 RING IS CRACKING. FEM DRILL IS DULL. BOX OSTEOTOME ARE BENDING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25578764 ↗MalfunctionJun 10, 2026Crack
IT WAS REPORTED THAT THE FEMORAL AND TIBIAL IMPACTORS BOTH CRACKED DURING SURGERY. THE IMPACTOR DID BREAK INTO SEVERAL PIECES. STAFF OPENED UP A BACKUP UNIVERSAL IMPACTOR. THERE WAS NO SURGICAL DELAY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25457926 ↗MalfunctionJun 9, 2026Break
IT WAS REPORTED THAT IMPACTOR BROKE DURING SECONDARY IMPACTING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25449084 ↗MalfunctionJun 5, 2026Break
IT WAS REPORTED THAT THE FEMORAL IMPACTOR WAS BROKE. THE DEVICE BROKE INTO TWO PIECES. YES THE FRAGMENTS WERE GENERATED AND RETRIEVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25410941 ↗MalfunctionJun 5, 2026Break
IT WAS REPORTED THAT THE IMPACTOR BROKE DURING FEMORAL IMPACTION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25410129 ↗MalfunctionJun 2, 2026Break
IT WAS REPORTED THAT THE IMPACTED PRODUCT HAD AN UNKNOWN ALLEGATION. IT WAS FURTHER REPORTED THAT THE DEVICES WERE FOUND TO BE BROKEN PRIOR TO SHIPMENT. THE EVENT DID NOT OCCUR IN SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25360926 ↗MalfunctionMay 29, 2026Break; Crack
IT WAS REPORTED THAT DURING IMPACTION THE IMPACTOR CRACKED AND SPLIT INTO TWO PIECES. THERE WAS NO REPORTED PATIENT OR USER HARM. THERE WAS NO SIGNIFICANT DELAY. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25336908 ↗MalfunctionMay 27, 2026Break
IT WAS REPORTED THAT DURING AN ARTHROPLASTY SURGICAL PROCEDURE IT WAS OBSERVED THAT THE IMPACTOR BROKE DURING FINAL IMPACTION. IT WAS REPORTED THAT THERE WAS NO DELAY IN THE PROCEDURE DUE TO THE EVENT. IT WAS REPORTED THAT THERE WERE FRAGMENTS GENERATED, AND ALL FRAGMENTS WERE RETRIEVED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25299893 ↗MalfunctionMay 26, 2026Break
IT WAS REPORTED THAT THE ATTUNE FEMORAL IMPACTOR BROKE INTO PIECES WHEN USING TO IMPACT THE DEFINITIVE ATTUNE CEMENTLESS FEMUR IMPLANT USED IN THE CASE. THE POSTERIOR SAW CAPTURE WAS BROKEN WHEN HANDED TO THE SURGEON. THE PLASTIC HOOK THAT SECURES THE CAPTURE TO THE BLOCK FOR USE HAD LOST ITS HOOK PART ON THE TAB THAT HOLDS IT IN A LOCKED POSITION SUCH THAT THE CAPTURE GOOD PULL OR BE SHAKEN OFF THE BLOCK WITHOUT IT HOLDING AS IT SHOULD. THERE WAS NO PATIENT OR USER HARM. THERE WAS NO SIGNIFICANT DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25285375 ↗MalfunctionMay 21, 2026Crack
IT WAS REPORTED THAT THE DEVICE CRACKED UPON FEMORAL IMPACTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25247245 ↗MalfunctionMay 20, 2026No Apparent Adverse Event
PRODUCTS WERE REPORTED FOR UNKNOWN REASON. IT DID NOT HAPPEN IN SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25232608 ↗MalfunctionMay 20, 2026Break
IT WAS REPORTED THAT THE INSTRUMENT WAS BROKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25229413 ↗MalfunctionMay 13, 2026Break; Crack
IT WAS REPORTED THAT DURING THE PROCEDURE THE SP2 TIBIAL IMPACTED BROKE INTO 2 PIECES AND THE ATTUNE FEMORAL IMPACTOR BROKE AND DEVELOPED A LARGE CRACK DURING FEMORAL IMPACTING. THERE WAS NO SURGICAL DELAY AND NO HARM TO PATIENT. ALL PIECES WHERE RETRIEVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25161943 ↗MalfunctionMay 6, 2026Break
IT WAS REPORTED THAT DURING IMPACTION THE ATTUNE FEMORAL IMPACTOR BROKE. ALL PIECES WERE RETRIEVED. IT WAS REPORTED THAT THERE WERE NO DELAYS OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25092122 ↗MalfunctionApr 28, 2026Break
IT WAS REPORTED THAT, DURING SURGERY, THE FEMORAL IMPACTOR BROKE INTO TWO PIECES. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25020822 ↗
FDA clearance and classification
The FDA regulates product code HWA (impactor) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the ATTUNE 2544-01-006
Has the ATTUNE 2544-01-006 been recalled?
Yes. 1 FDA recall record names this device's identifier, and the FDA has closed (terminated) it. The recall, Z-0905-2015 (Class II), began on Nov 10, 2014: ATTUNE INTUITION Impactors (cat. nos. 254401003, 254401004, and 254401006) have broken and produced small pieces. Should a fracture occur and not be observed during surgery, there is the potential for these small… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the ATTUNE 2544-01-006?
The FDA's MAUDE database lists 2,528 adverse event reports for this device, received between Jun 23, 2017 and Sep 25, 2026. In those reports, the most common specific device problems are “Break” (2,195), “Crack” (524) and “Naturally Worn” (56); the most common patient problem is “Not Applicable” (18). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ATTUNE 2544-01-006?
None of the reports are classified as deaths, 3 as injuries and 2,525 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the ATTUNE 2544-01-006 FDA approved?
Devices of this type (product code HWA) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Depuy (Ireland).
Who makes the ATTUNE 2544-01-006, and is it still sold?
The FDA's device database lists Depuy (Ireland) as the company responsible for it (brand ATTUNE, model 2544-01-006). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other IMPACTOR devices
- PINNACLE 2217-50-041Depuy (Ireland) · IMPACTOR4,276 reports
- ATTUNE 2544-01-003Depuy (Ireland) · IMPACTOR1 recall1,491 reports
- ATTUNE 2544-01-017Depuy (Ireland) · IMPACTOR1 recall1,454 reports
- ATTUNE 2544-01-004Depuy (Ireland) · IMPACTOR1 recall873 reports
- DURALOC 2360-71-000Depuy Orthopaedics, Inc. · IMPACTOR715 reports
- SPECIALIST 2581-11-000Depuy (Ireland) · IMPACTOR703 reports
- INHANCE 6205-10-103Depuy (Ireland) · IMPACTOR334 reports
- SELF CENTERING 2055-13-000Depuy (Ireland) · IMPACTOR322 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.