Viant Medical, LLC 510912Recalls & reported problems

Viant Medical · Metal Handle Offset Cup Impactor · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00840096401011), received between Apr 24, 2025 and Apr 24, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Connection Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Connection Problem1
  • Failure to Disconnect1
  • Mechanical Jam1
  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 24, 2025Mechanical Problem; Failure to Disconnect; Connection Problem; Mechanical Jam; Mechanical Problem; Failure to Disconnect; Connection Problem; Mechanical Jam

    IT WAS REPORTED DURING THE TOTAL HIP ARTHROPLASTY (THA) CUP PLACEMENT, THE CUP COULD NOT BE REMOVED FROM THE OFFSET IMPACTOR AND COULD NOT BE IMPLANTED INTO THE ACETABULUM. THE PHYSICIAN THEN REAMED UPWARD WITH A 46MM REAMER AND IMPLANTED A 46MM CUP USING A STRAIGHT IMPACTOR. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO IMPACT AND MEDICAL INTERVENTION REQUIRED. THE PROCEDURE WAS DELAYED 60 MINUTES DUE TO THE INCIDENT. THE USER FACILITY INDICATED THE PHYSICIAN HAD TIGHTENED THE CUP TOO HARD WHEN ATTACHING IT AND THE OFFSET IMPACTOR WAS NOT DAMAGED. NO CONSEQUENCES OR IMPACT TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21903186 ↗

Other Impactor devices