Skip to content

TULA TDS114GRecalls & reported problems

Smith & Nephew, Inc. · TULA TDS 1.14 GROMMET · Class 3 · Implant

All models of TULA (9)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00840128600047), received between Sep 10, 2021 and Sep 26, 2024 — classified as 3 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Activation Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Activation Problem1
  • Adverse Event Without Identified Device or Use Problem1
  • Device Dislodged or Dislocated1
  • Firing Problem1
  • Material Protrusion/Extrusion1
  • Unintended System Motion1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Failure of Implant1
  • Foreign Body In Patient1
  • Unspecified Infection1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 26, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT DURING A TYMPANOSTOMY, WHILE TUBE PLACEMENT; THE SURGEON REPORTED A TUBE MEDIALIZATION WITH THE SECOND TDS DEVICE. THE SURGEON REPORTED HE WAS ABLE TO SUCCESSFULLY PLACE A BACK UP DEVICE IN THE TYMPANIC MEMBRANE. THERE WAS A DELAY LESS THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20307244 ↗
  • InjuryNov 10, 2023Activation Problem; Activation Problem

    IT WAS REPORTED THAT DURING A TYMPANOSTOMY PROCEDURE, UPON DEPLOYING THE TUBE OF THE TULA DEVICE, THE SURGEON NOTICED HE DID NOT HEAR AN AUDIBLE CLICK AND HE FELT THAT THE BLADE DID NOT EXTRUDE FULLY TO MAKE AN INCISION AND THE TUBE SHORT SHOT (B)(4). UPON RETRIEVAL OF THE TUBE HE TRIED AGAIN WITH THE SECOND TDS FROM THE TULA KIT AND IT HAD THE SAME RESULT (B)(4). HE THEN TRIED A THIRD TIME WITH A TDS FROM A DIFFERENT KIT AND IT ALSO FAILED (B)(4). AFTER THE THIRD ATTEMPT THE PROCEDURE WAS ABORTED AND RESCHEDULED FOR THE OR. NO ADDITIONAL COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18114903 ↗
  • InjuryJul 27, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING A TYMPANOSTOMY PROCEDURE, THE TULA DELIVERY SYSTEM COULD NOT BE SUCCESSFULLY DEPLOYED. THE FAILURE OCCURRED IN A CLEFT PATIENT THAT HAD VERY SMALL AND UNDEVELOPED ANATOMY. THE EARDRUM WAS ANGLED SO IT WAS NOT POSSIBLE TO GET DIRECT APPOSITION OF THE TDS DEVICE FOR A SUCCESSFUL DEPLOYMENT. AFTER 2 ATTEMPTS ON THE SAME EAR USING A UNILATERAL KIT, THE PROCEDURE WAS ABORTED. A SURGERY IN THE OR WAS SCHEDULED. THE OUTCOME IS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17410474 ↗
  • InjuryNov 5, 2022Unintended System Motion; Material Protrusion/Extrusion; Unintended System Motion; Material Protrusion/Extrusion

    IT WAS REPORTED THAT TWO WEEKS AFTER A TYMPANOSTOMY PROCEDURE USING A TULA DEVICE, AT THE PATIENT FOLLOW UP IT WAS NOTICED THAT PATIENT EXPERIENCED AN EARLY TUBE EXTRUSION. PATIENT EAR HAD ACTIVE INFECTION. NO ADDITIONAL COMPLICATIONS WERE REPORTED.

    Patient problems: Failure of Implant; Unspecified Infection

    MAUDE report 15740866 ↗
  • MalfunctionSep 10, 2021Firing Problem; Firing Problem

    IT WAS REPORTED THAT DURING AN OFFICE-BASED ENT PROCEDURE, A TYMPANOSTOMY TUBE PLACED BY THE TULA TUBE DELIVERY SYSTEM WAS MEDIALIZED INTO THE MIDDLE EAR UPON ACTUATION OF THE DEVICE. THE ENT SURGEON STATED THAT DURING DEPLOYMENT, THE PATIENT WAS MOVING, AND SURGEON HAD POOR VISUALIZATION, AND MAYBE OVERCOMPENSATED WITH PRESSURE ON THE DRUM. THE ENT SURGEON STATED THAT THE EAR CANAL LOOKED SMALLER AND THAT THE CHILD WAS FEISTY BUT NOTHING UNUSUAL WAS NOTICED ABOUT THE TISSUE IN THE TARGET LOCATION, TUBE OR THE DEVICE. A DELAY OF APPROXIMATELY 10 MINUTES WAS REPORTED WHILE THE SURGEON ATTEMPTED TO RETRIEVE THE MEDIALIZED TUBE, WHICH ULTIMATELY WAS NOT SUCCESSFUL. A NON-SMITH & NEPHEW TYMPANOSTOMY TUBE (SILICONE ARMSTRONG GROMMET) WAS PLACED USING TRADITIONAL TECHNIQUES, AND THE TULA TYMPANOSTOMY TUBE WAS LEFT RETAINED IN THE MIDDLE EAR CAVITY. NO PATIENT INJURY OR COMPLICATIONS WERE REPORTED. THE PATIENT IS DOING WELL, AND STANDARD POST-SURGICAL FOLLOW-UPS ARE PLANNED. THE SURGEON WILL FOLLOW GUIDELINES THAT SUGGEST TUBE MAY BE LEFT IN THE MIDDLE EAR AND WILL EXPLANT IF CHILD NEEDS ANOTHER TUBE SURGERY IN WHICH CASE AN OR PROCEDURE WILL BE RECOMMENDED UNDER GENERAL ANESTHESIA FOR TH

    Patient problems: Foreign Body In Patient

    MAUDE report 12456624 ↗

Other Tympanostomy tube delivery product with drug devices

All Tympanostomy tube delivery product with drug devices →

Sources and updates