TULA 72290097Recalls & reported problems

Smith & Nephew, Inc. · TULA TDS 1.14 GROMMET (STR007011) · Class 3 · Implant

All models of TULA (9)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00885556817742), received between Sep 25, 2024 and Oct 31, 2025 — classified as 3 injuries.

The most frequently reported specific device problem is “Premature Activation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Premature Activation2
  • Device Dislodged or Dislocated1
  • Unexpected Shutdown1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Unspecified Mental, Emotional or Behavioral Problem1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 31, 2025Premature Activation; Premature Activation

    IT WAS REPORTED THAT, DURING A TYMPANOSTOMY, ONE (1) TULA TDS DEPLOYED THE TUBE PREMATURELY. THE PROCEDURE WAS CANCELLED AND RESCHEDULED FOR THE OR. NO FURTHER COMPLICATIONS WERE REPORTED. THE PATIENT IS FINE WITH AN OR PLACED TUBE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23442205 ↗
  • InjuryJan 28, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT AFTER AN UNLITERAL TYMPANOSTOMY ON (B)(6) 2024. THE PATIENT HAD A FOLLOW-UP VISIT. DURING THE POST OP EXAM, IT WAS OBSERVED THAT THE TUBE HAD MEDIALIZED, THOUGH THE EXACT TIME WAS UNCLEAR. THE PATIENT HAS BEEN SCHEDULED FOR A TUBE RETRIEVAL IN THE OR. NO FURTHER COMPLICATIONS WERE REPORTED. PATIENT CURRENT STATUS IS FINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21261140 ↗
  • InjurySep 25, 2024Premature Activation; Unexpected Shutdown; Premature Activation; Unexpected Shutdown

    IT WAS REPORTED THAT DURING A TYMPANOSTOMY, DURING THE IONTOPHORESIS, THE SURGEON DOUBLE CLICKED ON BOTH CHANNEL BUTTONS TO ENTER "REDUCE MODE" AND THE FAULT INDICATOR LIT UP AND THE CONTROL UNIT STOPPED WORKING. THE SURGEON INSISTED TO USE ANOTHER CONTROL UNIT AND UTILIZED A SPARE TO COMPLETE THE IONTOPHORESIS. AFTER IT WAS COMPLETED, THE SURGEON DEPLOYED THE TUBE DELIVERY SYSTEM IN WHICH A "SHORT SHOT" OCCURRED. THE SURGEON PROMPTLY REMOVED THE TUBE FROM THE EAR CANAL, AND UTILIZED A REPLACEMENT TDS DEVICE. THEY TRIED TO MOVE ON TO THE NEXT EAR AND DUE TO THE PATIENT´S TEMPERAMENT, THE PROCEDURE WAS ABORTED AND RE-SCHEDULED IN THE OPERATING ROOM. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: Unspecified Mental, Emotional or Behavioral Problem

    MAUDE report 20301458 ↗

Other Tympanostomy tube delivery product with drug devices

All Tympanostomy tube delivery product with drug devices →