ASSURE WCD System 80014-001Recalls & reported problems

Kestra Medical Technologies, Inc. · The ASSURE Wearable Cardioverter Defibrillator System · Class 3

All models of ASSURE WCD System (3)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This ASSURE wearable defibrillator is worn by patients to monitor the heart and deliver shocks when needed. Many of the 162 reports describe electrical problems, inappropriate shocks, and service error codes on the monitor. The reports mention 10 deaths and 2 injuries, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
162
Deaths reported
10
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 162 adverse event reports that list this exact device (DI 00840241600047), received between Mar 2, 2022 and Nov 19, 2025 — classified as 10 deaths, 2 injuries, 150 malfunctions.

The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (96 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 5 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Electrical /Electronic Property Problem96
  • Use of Device Problem42
  • Inappropriate/Inadequate Shock/Stimulation35
  • Adverse Event Without Identified Device or Use Problem13
  • Connection Problem7
  • Failure to Power Up7
  • Insufficient Device Problem Information5
  • Component Misassembled2

Patient problems

  • No Clinical Signs, Symptoms or Conditions114
  • Shock from Patient Lead(s)39
  • Insufficient Information7
  • Asystole2
  • Heart Failure/Congestive Heart Failure2

Recent adverse event reports

The 25 most recent of 162 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 19, 2025Inappropriate/Inadequate Shock/Stimulation; Use of Device Problem; Inappropriate/Inadequate Shock/Stimulation; Use of Device Problem

    PATIENT'S DOCTOR FROM ER REPORTED PATIENT ARRIVED VIA AMBULANCE AND STATED THEY RECEIVED THERAPEUTIC SHOCK THREE TIMES. THE REPORTED INJURY WAS NOT LIFE-THREATENING, DID NOT RESULT IN PERMANENT DAMAGE, OR NECESSITATE MEDICAL INTERVENTION TO PRECLUDE PERMANENT DAMAGE. HOWEVER, AN UNDIVERTED SHOCK TO A CONSCIOUS PATIENT COULD RESULT IN SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23596142 ↗
  • MalfunctionJul 16, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    PATIENT CALLED IN TO REPORT SERVICE ERROR CODE WITH WRENCH APPEARING ON MONITOR SCREEN. PATIENT REBOOTED THE ENTIRE WCD SYSTEM BUT THE ISSUE WAS NOT RESOLVED. THE UNRESOLVABLE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19750557 ↗
  • MalfunctionJan 11, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    SERVICE ERROR CODE WAS RECEIVED ON THE CUSTOMER SUPPORT HELPLINE QUEUE. PATIENT CONFIRMED RECEIVING A WRENCH ICON. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18499533 ↗
  • MalfunctionJan 8, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    SERVICE ERROR CODE WAS RECEIVED ON THE CUSTOMER SUPPORT HELPLINE QUEUE. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18471037 ↗
  • MalfunctionNov 30, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    PATIENT CALLED IN TO REPORT THAT IT ALERTED THE PATIENT TO "CALL THE HELPLINE, YOUR OUR DEVICE NEEDS SERVICE." THE PATIENT STATED THAT THERE IS A WRENCH ON THE MONITOR WITH THE SERVICE ERROR CODE. CUSTOMER CARE ADVISED THE PATIENT TO REBOOT THE SYSTEM. THE ISSUE WAS RESOLVED. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18238474 ↗
  • MalfunctionNov 7, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    SERVICE ERROR CODE WAS RECEIVED ON THE CUSTOMER SUPPORT HELPLINE QUEUE. THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18085883 ↗
  • MalfunctionNov 4, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    CUSTOMER SUPPORT RECEIVED A SERVICE REQUIRED CODE FROM A CALL BY PATIENT'S FAMILY CAREGIVER THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18070338 ↗
  • MalfunctionNov 4, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    CUSTOMER SUPPORT RECEIVED A SERVICE REQUIRED CODE ON THE HELPLINE QUEUE. CUSTOMER SUPPORT CONFIRMED THE PATIENT IS UNABLE TO REBOOT THE SYSTEM WITHOUT REGENERATING THE SERVICE CODE. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18070301 ↗
  • MalfunctionNov 4, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    CUSTOMER SUPPORT RECEIVED A SERVICE REQUIRED CODE ON THE HELPLINE QUEUE CUSTOMER SUPPORT WAS ABLE TO CONFIRM THE PATIENT CLEARED THE CODE BY RE-BOOTING THE DEVICE. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18070289 ↗
  • MalfunctionOct 23, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    SERVICE ERROR CODE WAS RECEIVED ON THE CUSTOMER SUPPORT HELPLINE QUEUE. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17986198 ↗
  • MalfunctionOct 16, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    SERVICE ERROR CODE WAS RECEIVED ON THE CUSTOMER SUPPORT HELPLINE QUEUE. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17946920 ↗
  • MalfunctionOct 16, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    SERVICE ERROR CODE WAS RECEIVED ON THE CUSTOMER SUPPORT HELPLINE QUEUE. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17946890 ↗
  • MalfunctionOct 16, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    SERVICE ERROR CODE WAS RECEIVED ON THE CUSTOMER SUPPORT HELPLINE QUEUE. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17946884 ↗
  • MalfunctionOct 16, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    SERVICE ERROR CODE WAS RECEIVED ON THE CUSTOMER SUPPORT HELPLINE QUEUE. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17946881 ↗
  • MalfunctionOct 10, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    SERVICE ERROR CODE WAS RECEIVED ON THE CUSTOMER SUPPORT HELPLINE QUEUE. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17909631 ↗
  • MalfunctionOct 10, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    SERVICE ERROR CODE WAS RECEIVED ON THE CUSTOMER SUPPORT HELPLINE QUEUE. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17908833 ↗
  • MalfunctionOct 3, 2023Inappropriate/Inadequate Shock/Stimulation; Inappropriate/Inadequate Shock/Stimulation

    PATIENT RECEIVED A THERAPEUTIC SHOCK WHILE WEARING THE WCD IN HOSPITAL FOLLOWING AN UNRELATED SURGERY. SPOUSE REPORTED THE PATIENT WAS LIGHTLY SEDATED AND RESTRAINED WHILE ON A VENTILATOR AND UNABLE TO DIVERT THE SHOCK. SPOUSE REPORTED THAT NURSES WERE INDEPENDENTLY AWARE OF PATIENT'S HIGH HEART RATE OVER AN EXTENDED PERIOD OF TIME BUT WOULD NOT HAVE SHOCKED THE PATIENT. CUSTOMER CARE ADVISED THE PATIENT'S SPOUSE THAT THE WCD IS NOT INTENDED FOR IN-HOSPITAL USE AND ADVISED THE SPOUSE SPEAK TO PATIENT'S HEALTHCARE PROVIDER ABOUT REMOVING THE SYSTEM.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 17865670 ↗
  • MalfunctionSep 26, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    SERVICE ERROR CODE WAS RECEIVED ON THE CUSTOMER SUPPORT HELPLINE QUEUE. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17821957 ↗
  • MalfunctionSep 26, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    SERVICE ERROR CODE WAS RECEIVED ON THE CUSTOMER SUPPORT HELPLINE QUEUE. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17821504 ↗
  • MalfunctionSep 26, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    SERVICE ERROR CODE WAS RECEIVED ON THE CUSTOMER SUPPORT HELPLINE QUEUE. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17821503 ↗
  • MalfunctionSep 18, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    SERVICE ERROR CODE WAS RECEIVED ON THE CUSTOMER SUPPORT HELPLINE QUEUE. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17769794 ↗
  • MalfunctionSep 18, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    CUSTOMER CARE RECEIVED A SERVICE REQUIRED NOTIFICATION ON THE HELPLINE QUEUE. CUSTOMER CARE CALLED THE PATIENT WHO SAID HE RECEIVED TWO ALERTS. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17769775 ↗
  • DeathSep 14, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    THE PATIENT'S FRIEND STEPPED OUT OF THE ROOM TO GET SOME ICE AND ON RETURN THE PATIENT WAS FOUND UNRESPONSIVE. WHEN ASKED IF A FAMILY MEMBER KNEW WHETHER THE PATIENT WAS WEARING THE WCD AT THE TIME, THE ANSWER WAS AFFIRMATIVE. THE WCD WAS CUT OFF THE PATIENT BY THE PERSONS THAT RESPONDED TO THE 911 CALL -- AND THE RESPONDERS DISPOSED OF THE WCD. THE FAMILY MEMBER REPORTED CAUSE OF DEATH AS CONGESTIVE HEART FAILURE. BECAUSE THE WCD WAS DESTROYED, THE PATIENT'S EVENT LOGS AND EPISODE REPORTS ARE NOT AVAILABLE FOR FURTHER INVESTIGATION..

    Patient problems: Heart Failure/Congestive Heart Failure

    MAUDE report 17751295 ↗
  • MalfunctionSep 14, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    A SERVICE REQUIRED CODE WAS RECEIVED ON THE CUSTOMER SUPPORT HELPLINE QUEUE. THE SERVICE CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17750820 ↗
  • MalfunctionSep 13, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    SERVICE ERROR CODE WAS RECEIVED ON THE CUSTOMER SUPPORT HELPLINE QUEUE. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17741450 ↗

Other Wearable Automated External Defibrillator devices

All Wearable Automated External Defibrillator devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page