Kestra Medical Technologies, Inc. · The ASSURE Wearable Cardioverter Defibrillator System · Class 3
All models of ASSURE WCD System (3)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This ASSURE wearable defibrillator is worn by patients to monitor the heart and deliver shocks when needed. Many of the 162 reports describe electrical problems, inappropriate shocks, and service error codes on the monitor. The reports mention 10 deaths and 2 injuries, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 162
- Deaths reported
- 10
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 162 adverse event reports that list this exact device (DI 00840241600047), received between Mar 2, 2022 and Nov 19, 2025 — classified as 10 deaths, 2 injuries, 150 malfunctions.
The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (96 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Electrical /Electronic Property Problem96
- Use of Device Problem42
- Inappropriate/Inadequate Shock/Stimulation35
- Adverse Event Without Identified Device or Use Problem13
- Connection Problem7
- Failure to Power Up7
- Insufficient Device Problem Information5
- Component Misassembled2
Patient problems
- No Clinical Signs, Symptoms or Conditions114
- Shock from Patient Lead(s)39
- Insufficient Information7
- Asystole2
- Heart Failure/Congestive Heart Failure2
Recent adverse event reports
The 25 most recent of 162 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 19, 2025Inappropriate/Inadequate Shock/Stimulation; Use of Device Problem; Inappropriate/Inadequate Shock/Stimulation; Use of Device Problem
PATIENT'S DOCTOR FROM ER REPORTED PATIENT ARRIVED VIA AMBULANCE AND STATED THEY RECEIVED THERAPEUTIC SHOCK THREE TIMES. THE REPORTED INJURY WAS NOT LIFE-THREATENING, DID NOT RESULT IN PERMANENT DAMAGE, OR NECESSITATE MEDICAL INTERVENTION TO PRECLUDE PERMANENT DAMAGE. HOWEVER, AN UNDIVERTED SHOCK TO A CONSCIOUS PATIENT COULD RESULT IN SERIOUS INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23596142 ↗MalfunctionJul 16, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
PATIENT CALLED IN TO REPORT SERVICE ERROR CODE WITH WRENCH APPEARING ON MONITOR SCREEN. PATIENT REBOOTED THE ENTIRE WCD SYSTEM BUT THE ISSUE WAS NOT RESOLVED. THE UNRESOLVABLE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19750557 ↗MalfunctionJan 11, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
SERVICE ERROR CODE WAS RECEIVED ON THE CUSTOMER SUPPORT HELPLINE QUEUE. PATIENT CONFIRMED RECEIVING A WRENCH ICON. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18499533 ↗MalfunctionJan 8, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
SERVICE ERROR CODE WAS RECEIVED ON THE CUSTOMER SUPPORT HELPLINE QUEUE. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18471037 ↗MalfunctionNov 30, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
PATIENT CALLED IN TO REPORT THAT IT ALERTED THE PATIENT TO "CALL THE HELPLINE, YOUR OUR DEVICE NEEDS SERVICE." THE PATIENT STATED THAT THERE IS A WRENCH ON THE MONITOR WITH THE SERVICE ERROR CODE. CUSTOMER CARE ADVISED THE PATIENT TO REBOOT THE SYSTEM. THE ISSUE WAS RESOLVED. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18238474 ↗MalfunctionNov 7, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
SERVICE ERROR CODE WAS RECEIVED ON THE CUSTOMER SUPPORT HELPLINE QUEUE. THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18085883 ↗MalfunctionNov 4, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
CUSTOMER SUPPORT RECEIVED A SERVICE REQUIRED CODE FROM A CALL BY PATIENT'S FAMILY CAREGIVER THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18070338 ↗MalfunctionNov 4, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
CUSTOMER SUPPORT RECEIVED A SERVICE REQUIRED CODE ON THE HELPLINE QUEUE. CUSTOMER SUPPORT CONFIRMED THE PATIENT IS UNABLE TO REBOOT THE SYSTEM WITHOUT REGENERATING THE SERVICE CODE. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18070301 ↗MalfunctionNov 4, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
CUSTOMER SUPPORT RECEIVED A SERVICE REQUIRED CODE ON THE HELPLINE QUEUE CUSTOMER SUPPORT WAS ABLE TO CONFIRM THE PATIENT CLEARED THE CODE BY RE-BOOTING THE DEVICE. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18070289 ↗MalfunctionOct 23, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
SERVICE ERROR CODE WAS RECEIVED ON THE CUSTOMER SUPPORT HELPLINE QUEUE. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17986198 ↗MalfunctionOct 16, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
SERVICE ERROR CODE WAS RECEIVED ON THE CUSTOMER SUPPORT HELPLINE QUEUE. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17946920 ↗MalfunctionOct 16, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
SERVICE ERROR CODE WAS RECEIVED ON THE CUSTOMER SUPPORT HELPLINE QUEUE. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17946890 ↗MalfunctionOct 16, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
SERVICE ERROR CODE WAS RECEIVED ON THE CUSTOMER SUPPORT HELPLINE QUEUE. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17946884 ↗MalfunctionOct 16, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
SERVICE ERROR CODE WAS RECEIVED ON THE CUSTOMER SUPPORT HELPLINE QUEUE. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17946881 ↗MalfunctionOct 10, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
SERVICE ERROR CODE WAS RECEIVED ON THE CUSTOMER SUPPORT HELPLINE QUEUE. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17909631 ↗MalfunctionOct 10, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
SERVICE ERROR CODE WAS RECEIVED ON THE CUSTOMER SUPPORT HELPLINE QUEUE. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17908833 ↗MalfunctionOct 3, 2023Inappropriate/Inadequate Shock/Stimulation; Inappropriate/Inadequate Shock/Stimulation
PATIENT RECEIVED A THERAPEUTIC SHOCK WHILE WEARING THE WCD IN HOSPITAL FOLLOWING AN UNRELATED SURGERY. SPOUSE REPORTED THE PATIENT WAS LIGHTLY SEDATED AND RESTRAINED WHILE ON A VENTILATOR AND UNABLE TO DIVERT THE SHOCK. SPOUSE REPORTED THAT NURSES WERE INDEPENDENTLY AWARE OF PATIENT'S HIGH HEART RATE OVER AN EXTENDED PERIOD OF TIME BUT WOULD NOT HAVE SHOCKED THE PATIENT. CUSTOMER CARE ADVISED THE PATIENT'S SPOUSE THAT THE WCD IS NOT INTENDED FOR IN-HOSPITAL USE AND ADVISED THE SPOUSE SPEAK TO PATIENT'S HEALTHCARE PROVIDER ABOUT REMOVING THE SYSTEM.
Patient problems: Shock from Patient Lead(s)
MAUDE report 17865670 ↗MalfunctionSep 26, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
SERVICE ERROR CODE WAS RECEIVED ON THE CUSTOMER SUPPORT HELPLINE QUEUE. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17821957 ↗MalfunctionSep 26, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
SERVICE ERROR CODE WAS RECEIVED ON THE CUSTOMER SUPPORT HELPLINE QUEUE. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17821504 ↗MalfunctionSep 26, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
SERVICE ERROR CODE WAS RECEIVED ON THE CUSTOMER SUPPORT HELPLINE QUEUE. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17821503 ↗MalfunctionSep 18, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
SERVICE ERROR CODE WAS RECEIVED ON THE CUSTOMER SUPPORT HELPLINE QUEUE. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17769794 ↗MalfunctionSep 18, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
CUSTOMER CARE RECEIVED A SERVICE REQUIRED NOTIFICATION ON THE HELPLINE QUEUE. CUSTOMER CARE CALLED THE PATIENT WHO SAID HE RECEIVED TWO ALERTS. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17769775 ↗DeathSep 14, 2023Insufficient Device Problem Information; Insufficient Device Problem Information
THE PATIENT'S FRIEND STEPPED OUT OF THE ROOM TO GET SOME ICE AND ON RETURN THE PATIENT WAS FOUND UNRESPONSIVE. WHEN ASKED IF A FAMILY MEMBER KNEW WHETHER THE PATIENT WAS WEARING THE WCD AT THE TIME, THE ANSWER WAS AFFIRMATIVE. THE WCD WAS CUT OFF THE PATIENT BY THE PERSONS THAT RESPONDED TO THE 911 CALL -- AND THE RESPONDERS DISPOSED OF THE WCD. THE FAMILY MEMBER REPORTED CAUSE OF DEATH AS CONGESTIVE HEART FAILURE. BECAUSE THE WCD WAS DESTROYED, THE PATIENT'S EVENT LOGS AND EPISODE REPORTS ARE NOT AVAILABLE FOR FURTHER INVESTIGATION..
Patient problems: Heart Failure/Congestive Heart Failure
MAUDE report 17751295 ↗MalfunctionSep 14, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
A SERVICE REQUIRED CODE WAS RECEIVED ON THE CUSTOMER SUPPORT HELPLINE QUEUE. THE SERVICE CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17750820 ↗MalfunctionSep 13, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
SERVICE ERROR CODE WAS RECEIVED ON THE CUSTOMER SUPPORT HELPLINE QUEUE. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT. HOWEVER, THE SERVICE REQUIRED EVENT CODE INDICATES A WCD SYSTEM ISSUE THAT IF IT WERE TO RECUR COULD RESULT IN THE WCD FAILING TO PROVIDE NEEDED THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17741450 ↗
Other Wearable Automated External Defibrillator devices
- LifeVest Wearable Defibrillator LifeVest 4000 MonitorZoll Manufacturing Corporation · Wearable Automated External Defibrillator2,039 deaths36,128 reports
- LifeVest Wearable Defibrillator LifeVest 4000 Electrode BeltZoll Manufacturing Corporation · Wearable Automated External Defibrillator44 deaths15,404 reports
- LifeVest Wearable Defibrillator LifeVest 4000 Battery PackZoll Manufacturing Corporation · Wearable Automated External Defibrillator1 death9,929 reports
- LifeVest Wearable Defibrillator LifeVest 4000 Battery ChargerZoll Manufacturing Corporation · Wearable Automated External Defibrillator1 death6,327 reports
- LifeVest Wearable Defibrillator LifeVest 4000 Electrode BeltZoll Manufacturing Corporation · Wearable Automated External Defibrillator14 deaths4,460 reports
- LifeVest Wearable Defibrillator LifeVest 4000 Battery ChargerZoll Manufacturing Corporation · Wearable Automated External Defibrillator1 death1,296 reports
- ASSURE WCD System 80014-002Kestra Medical Technologies, Inc. · Wearable Automated External Defibrillator31 deaths930 reports
- LifeVest Wearable Defibrillator LifeVest 5100 Battery PackZoll Manufacturing Corporation · Wearable Automated External Defibrillator557 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page