Hydra SIT-575Recalls & reported problems

Boston Scientific Corporation · Irrigation Tubing System · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00840253108418), received between Jul 11, 2024 and Jan 2, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Reflux within Device” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Reflux within Device5
  • Use of Device Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 2, 2025Reflux within Device; Use of Device Problem; Reflux within Device; Use of Device Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT HYDRA IRRIGATION TUBING WAS USED IN THE COLON DURING A COLONOSCOPY PROCEDURE PERFORMED FOR SCREENING ON (B)(6) 2024. DURING THE PROCEDURE, THERE WAS BACKFLOW INTO THE IRRIGATION TUBE, THE COLOR OF THE FLUID WAS CAUSED BY BLOOD. THE PROCEDURE WAS COMPLETED WITH THIS DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21056109 ↗
  • MalfunctionNov 19, 2024Reflux within Device; Use of Device Problem; Reflux within Device; Use of Device Problem

    IT WAS REPORTED, TO BOSTON SCIENTIFIC CORPORATION. THAT HYDRA IRRIGATION TUBING WAS USED IN THE COLON, DURING A COLONOSCOPY PROCEDURE. PERFORMED FOR SCREENING ON (B)(6) 2024. DURING THE PROCEDURE, BACKFLOW OF BLOOD COLORED FLUID INTO THE IRRIGATION TUBE WAS OBSERVED. THE PROCEDURE WAS COMPLETED WITH THIS DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED, AS A RESULT OF THIS EVENT. NOTE: IT WAS REPORTED, THAT THE CUSTOMER CONNECTED THE ENDOWASHER DIRECTLY TO THE LUER CONNECTOR OF THE IRRIGATION CHANNEL OF THE ENDOSCOPE. HOWEVER, THE HYDRA INSTRUCTIONS FOR USE (IFU) STATE, TO CONNECT THE IRRIGATION TUBING, THE APPROPRIATE IRRIGATION SCOPE CONNECTOR THAT IS ATTACHED TO THE AUXILIARY WATER JET PORT. THE PHYSICIAN DID NOT FOLLOW THE STEPS CITED IN THE IFU.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20722572 ↗
  • MalfunctionSep 30, 2024Reflux within Device; Reflux within Device

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT HYDRA IRRIGATION TUBING WAS USED IN A CHOLANGIOSCOPY PROCEDURE PERFORMED IN COMMON BILE DUCT. DURING THE PROCEDURE, THE VALVE THAT WAS NOT WORKING. THE FLUSHING FLUID RAN BACK INTO THE SYSTEM. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20334804 ↗
  • MalfunctionSep 23, 2024Reflux within Device; Reflux within Device

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT HYDRA IRRIGATION TUBING WAS USED IN A CHOLANGIOSCOPY PROCEDURE PERFORMED IN COMMON BILE DUCT ON AN UNKNOWN PROCEDURE DATE. DURING THE PROCEDURE, THE ONE-WAY CHECK VALVE DID NOT FUNCTION PROPERLY. THE FLUSHING FLUID RAN BACK INTO THE SYSTEM. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20281198 ↗
  • MalfunctionJul 11, 2024Reflux within Device; Reflux within Device

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT HYDRA IRRIGATION TUBING WAS USED IN AN UNKNOWN PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE WATER FLEW BACK DURING IRRIGATION. IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED. THERE WERE NO PATIENT COMPLICATION AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19719263 ↗

Other PUMP, AIR, NON-MANUAL, FOR ENDOSCOPE devices

All PUMP, AIR, NON-MANUAL, FOR ENDOSCOPE devices →