ERBEFLO CleverCap® 20325-248Recalls & reported problems
Erbe Medical, LLC · Hybrid CO2 Tubing/Cap Set for Fujifilm® Scopes · Class 2
All models of ERBEFLO CleverCap (8)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04065655001006), received between Apr 14, 2026 and Jun 18, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Mar 31, 2026Open, Classified
Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported third-party connector use conditions, which may lead to water aspiration, which may progress to aspiration pneumonia, acute respiratory distress syndrome, or respiratory failure.
Z-2062-2026 · Root cause: Component change control · Recall details · FDA record ↗
Most reported problems
Device problems
- Leak/Splash3
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 18, 2026Leak/Splash
IT WAS REPORTED THAT SOMETIMES THE HYBRID CO2 TUBING/CAP SETS FOR FUJIFILM® LEAK FROM THE IRRIGATION LINE AND SCOPE BUTTONS WHEN CO2 IS ACTIVATED AND IRRIGATION IS INACTIVE. THE SETS ARE BEING USED WITH AN ERBE CO2 INSUFFLATOR [MODEL ECO 2 PLUS, PART NUMBER (P/N) 2N100-068, SERIAL NUMBER (S/N) INFORMATION NOT PROVIDED (NI); ERBE IRRIGATION PUMP (MODEL EIP 2, P/N 10325-000, S/N NI); FUJIFILM® SCOPE (P/N NI, SN NI), AND FUJI DISPOSABLE BUTTONS (P/N NI, LOT NUMBER NI). WHEN THE SET IS CHANGED, THE ISSUE IS RESOLVED. NO PATIENT INJURIES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25547991 ↗MalfunctionJun 18, 2026Leak/Splash
IT WAS REPORTED THAT THE HYBRID CO2 TUBING/CAP SETS FOR FUJIFILM® CONSISTENTLY AND STEADILY LEAKED FROM THE DISTAL END OF THE SCOPE. NO PATIENT INJURIES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25547900 ↗MalfunctionApr 14, 2026Leak/Splash
IT WAS REPORTED THAT AN INCIDENT OCCURRED WITH A HYBRID CO2 TUBING/CAP SET FOR FUJIFILM SCOPES BEING USED FOR AN UPPER GI PROCEDURE. THE SET WAS USED WITH A 700 SERIES FUJI SCOPE, MV STRATUS CO2 INSUFFLATOR, AND AN ERBE IRRIGATION PUMP MODEL EIP 2 FOR AN UPPER GI PROCEDURE. SPECIFICALLY, LEAKING OCCURRED FROM BOTH SCOPE BUTTONS AS WELL AS OUT THE DISTAL END OF THE SCOPE. UPON CHANGING THE TUBING/CAP SET, THE LEAKING CEASED. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24875243 ↗
Other Pump, Air, Non-Manual, For Endoscope devices
- OLYMPUS OFP-2 (ROW)Keymed (Medical & Industrial Equipment) Limited · Pump, Air, Non-Manual, For Endoscope296 reports
- Olympus MAJ-1651Keymed (Medical & Industrial Equipment) Limited · Pump, Air, Non-Manual, For Endoscope125 reports
- Olympus WB991046OLYMPUS Winter & Ibe GmbH · Pump, Air, Non-Manual, For Endoscope81 reports
- ERBEFLO CleverCap® 20325-239Erbe Medical, LLC · Pump, Air, Non-Manual, For Endoscope1 recall24 reports
- OLYMPUS OFP-2 (US)Keymed (Medical & Industrial Equipment) Limited · Pump, Air, Non-Manual, For Endoscope1 recall22 reports
- ERBEFLO CleverCap® 20325-206Erbe Medical, LLC · Pump, Air, Non-Manual, For Endoscope1 recall5 reports
- ERBEFLO CleverCap® 20325-240Erbe Medical, LLC · Pump, Air, Non-Manual, For Endoscope1 recall3 reports
- ERBEFLO CleverCap® 20325-208Erbe Medical, LLC · Pump, Air, Non-Manual, For Endoscope1 recall1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page