Skip to content

Prestige Coil System PRES1240COM18Recalls & reported problems

Balt USA LLC · The Prestige Coil System is indicated for arterial and venous embolizations in the peripheral vasculature. It is designed for use with the Ultra Detachment Controller. · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K220699); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Jan 3, 2024Open, Classified

    Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.

    Z-1042-2024 · Root cause: Material/Component Contamination · FDA record ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K220699. One submission often covers many models, so check the product and lot details in each record.

  • Z-1989-2025 · May 8, 2025 · Open, Classified
    Prestige Coil System REFs/UDI-DI codes: PRES0153CXPPLT 00810068567311 PRES1045CPKPLT 00810068567663 PRES1245CPKPLT 00810068567694 PRES1445CPKPLT 00810068567724 PRES0103CXPPLT 00810

Other Device, Vascular, For Promoting Embolization devices

All Device, Vascular, For Promoting Embolization devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page