Balt USA LLC · The Prestige Coil System is indicated for arterial and venous embolizations in the peripheral vasculature. It is designed for use with the Ultra Detachment Controller. · Class 2 · Implant

All models of Prestige Coil System (97)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00840303708124), received between Jul 31, 2026 and Jul 31, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Premature Separation” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K220699); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 3, 2024Open, Classified

    Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.

    Z-1042-2024 · Root cause: Material/Component Contamination · Recall details · FDA record ↗

Most reported problems

Device problems

  • Premature Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 31, 2026Premature Separation

    It was reported that:"During the procedure to place an endoprosthesis for a right common iliac artery aneurysm, the aneurysm was visualized via arteriography. Selective cannulation of the right iliac artery was performed using a guidewire; the catheter was not advanced further. VERT, Internal Mammary, and Simmons 1 catheters were used. Puncture of the left common femoral artery was performed, followed by placement of a Simmons catheter and cannulation of the contralateral right common iliac artery, without advancing the catheter into the right internal iliac artery. The 5Fr sheath in the right common femoral artery was exchanged for a long Sentran sheath. Two guidewires were passed through the Sentran sheath: a Lunderquist wire into the aorta and a hydrophilic wire into the aneurysm. A 5Fr catheter was placed over the hydrophilic wire, and a micro-guidewire and microcatheter were introduced through the 5Fr catheter. A Medtronic endoprosthesis was deployed, and the aneurysm was excluded. A coil was passed through the microcatheter; however, it was damaged and broke inside the catheter. The microcatheter was withdrawn, and an attempt was made to advance the coil using a saline flush,

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50105322 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K220699. One submission often covers many models, so check the product and lot details in each record.

  • Z-1989-2025 · May 8, 2025 · Open, Classified
    Prestige Coil System REFs/UDI-DI codes: PRES0153CXPPLT 00810068567311 PRES1045CPKPLT 00810068567663 PRES1245CPKPLT 00810068567694 PRES1445CPKPLT 00810068567724 PRES0103CXPPLT 00810

Other Device, Vascular, For Promoting Embolization devices

All Device, Vascular, For Promoting Embolization devices →