Prestige Coil System PRES2465COM18Recalls & reported problems
Balt USA LLC · The Prestige Coil System is indicated for arterial and venous embolizations in the peripheral vasculature. It is designed for use with the Ultra Detachment Controller. · Class 2 · Implant
All models of Prestige Coil System (97)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K220699); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Recalls naming this device
- Jan 3, 2024Open, Classified
Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.
Z-1042-2024 · Root cause: Material/Component Contamination · Recall details · FDA record ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K220699. One submission often covers many models, so check the product and lot details in each record.
- Z-1989-2025 · May 8, 2025 · Open, ClassifiedPrestige Coil System REFs/UDI-DI codes: PRES0153CXPPLT 00810068567311 PRES1045CPKPLT 00810068567663 PRES1245CPKPLT 00810068567694 PRES1445CPKPLT 00810068567724 PRES0103CXPPLT 00810
Other Device, Vascular, For Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page