Optima Coil System OPTI0220BLKRecalls & reported problems

Balt USA LLC · The Optima Coil System is a series of specialized coils that are inserted into the vasculature under angiographic visualization to embolize intracranial aneurysms and other vascular anomalies. The system consists of an embolization coil implant comprised of platinum and tungsten, affixed to a delivery pusher to facilitate insertion into the hub of a microcatheter. The system is available in various shapes, lengths, and sizes. The devices are to be placed into aneurysms to create blood stasis, reducing flow into the aneurysm and thrombosing the aneurysm. Upon positioning coils into the aneurysm, the coils are detached from the delivery pusher in a serial manner until the aneurysm is occluded. · Class 2 · Implant

All models of Optima Coil System (379)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00840303711858), received between Jul 18, 2025 and May 14, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Premature Separation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Premature Separation2
  • Migration1
  • Stretched1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 14, 2026Premature Separation

    IT WAS REPORTED THAT: "DURING A COIL EMBOLIZATION OF MMA FOR CSDH, THE PHYSICIAN ATTEMPTED TO DEPLOY OPTIBLOCK 2X20 COIL THROUGH A SL-10 MICROCATHETER. THE COIL PUSHED BACK THE SL-10 AS IT WAS ADVANCED OUT OF DISTAL END. THE PHYSICIAN RESHEATHED COIL AND REPOSITIONED MICRO MORE DISTAL. HE THEN RELOADED THE OPTIBLOCK 2X20 COIL AND AFTER DEPLOYING THE FIRST TWO LOOPS OF COIL HE LOST 1 TO 1 AND IT WAS OBSERVED THAT THE COIL WAS NO LONGER ATTACHED TO DELIVERY WIRE. HE USED A SYNCHRO WIRE TO PUSH THE REMAINDER OF THE COIL OUT OF SL-10. THIS WAS NOTED TO BE DIFFICULT BUT THE ENTIRE COIL WAS ULTIMATELY PUSHED INTO VESSEL. COIL WAS PREPPED PER DFU. NO PATIENT INJURY OCCURRED. THE DELIVERY WIRE IS AVAILABLE FOR RETURN." INCIDENT REPORT FORM SUBMITTED FOR THIS COMPLAINT INDICATES THAT NO PATIENT INJURY WAS SUSTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25174268 ↗
  • MalfunctionJul 18, 2025Stretched; Migration; Premature Separation; Stretched; Migration; Premature Separation

    IT WAS REPORTED THAT: "WHEN RESHEATHING THE 2X20 COIL IN THE HEADWAY DUO 156CM MICROCATHETER IN THE MMA, THE COIL STRETCHED AND THEN BROKE OFF THE DELIVERY WIRE. THE COIL WAS NOT ABLE TO BE RECAPTURED AND WAS LEFT IN THE ECA WHERE IT BROKE OFF. FORTUNATELY, NO PATIENT HARM. THE SAME MICROCATHETER WAS USED TO DELIVER COILS FOR THE REMAINDER OF THE CASE." UPDATE 23JUN2025 - ADDITIONAL INFORMATION RECEIVED FROM THE ISSUER: " IS IT KNOWN APPROXIMATELY WHERE THE ACTUAL "BREAK" ALONG THE COIL TOOK PLACE (DID IT OCCUR AT THE DETACHMENT ZONE, DID THE IMPLANT COIL WIND BREAK, ETC.)? THE ACTUAL COIL IMPLANT STRETCHED MIDWAY THROUGH AND THEN BROKE AT THE DETACHMENT ZONE. WHERE IT IS MENTIONED "WHEN RESHEATHING THE 2X20 COIL IN THE HEADWAY DUO 156CM MICROCATHETER IN THE MMA", DOES THIS MEAN THE IMPLANT WAS DEPLOYED FROM THE MC AND THE PHYSICIAN WAS ATTEMPTING TO RETRACT BACK INTO THE MICROCATHETER? THE MICROCATHETER WAS KICKING BACK AS HE WAS TRYING TO DEPLOY THE COIL, SO THERE WAS SOME RESHEATHING OF THE COIL INSIDE THE MICROCATHETER TO REPOSITION. BASED ON THE ANATOMICAL LOCATION OF THE PROCEDURE, WHAT LEVEL OF VESSEL TORTUOSITY WOULD YOU CONSIDER WAS ENCOUNTERED BY THE COIL SYSTEM? THERE WA

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22546351 ↗

Other Device, Neurovascular Embolization devices

All Device, Neurovascular Embolization devices →