Skip to content

Prestige Coil System PRESPACK30V02Recalls & reported problems

Balt USA LLC · The Prestige Packing Coil consists of an implantable embolization coil comprised of a platinum-tungsten alloy attached to a proximal stainless steel hypo-tube and distal body coil delivery pusher with a radiopaque distal positioning marker and a proximal fluorosafe marker. The Prestige Packing Coil delivery pusher is 185cm in length. It is designed for use with the Ultra Detachment Controller. The Ultra Detachment Controller comes pre-loaded with batteries and is a sterile, handheld, single-patient-use device which should not to be reused, re-sterilized, opened or tampered with. The Prestige Packing Coil and Ultra Detachment Controller(s) are sold separately. The Prestige Packing Coil achieves detachment by an internal heater element, which is powered by the Ultra Detachment Controller. · Class 2 · Implant

All models of Prestige Coil System (97)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00840303712046), received between Aug 4, 2025 and Aug 4, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Separation Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Separation Problem1
  • Stretched1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 4, 2025Stretched; Separation Problem; Stretched; Separation Problem

    IT WAS REPORTED THAT: "PRESPACK30 WAS UTILIZED IN A PELVIC EMBO COIL TRACKED AND PACKED GREAT GOT GREEN LIGHT ON ULTRA AND COIL DID NOT FULLY DETACH. BELIEVE THE PUSHER BECAME ENTANGLED WITH COIL AND PULLED COIL BACK. WE THEN TRIED TO SNARE THROUGH 4 FRENCH CATHETER, WAS ABLE TO GET ABOUT 15CM PULLED BACK AND THEN COIL STRETCHED. STARTED TO PULL STRETCH COIL AND REST OF COIL CAME OUT OF BODY. NO INJURY TO PATIENT UTILIZED (2) 2X12 AND 2X8 AND 2X4 WITH OUT ISSUE. " UPDATE 08JUL2025 - ADDITIONAL INFORMATION RECEIVED FROM THE ISSUER: "1. HOW WOULD THE TORTUOSITY BE DESCRIBED IN THIS PROCEDURE? SOME TORTUOUS VESSEL 2. CAN YOU CONFIRM IF THE SUPPLEMENTAL ACCESSORIES WILL BE RETURNED? (XCEL, MICROCATHETER, ETC) IF SO, IT WOULD BE APPRECIATED. YES 3. CAN YOU CONFIRM IF THE IMPLANT DETACHED AT ANY POINT IN THE PROCEDURE? YES BUT SEEMED TO BE ENTANGLED". UPDATE 16JUL2025 - ADDITIONAL INFORMATION RECEIVED FROM THE ISSUER: 1. WAS THE DETACHMENT OF THE IMPLANT COIL ACHIEVED FROM THE PHYSICIAN PRESSING THE DETACHMENT BUTTON ON THE CONTROLLER? I BELIEVE SO, THINK THE PUSHER WAS ENTANGLED IN COIL WHICH CAUSED COIL TO BE PULLED BACK INCIDENT REPORT FORM SUBMITTED FOR THIS COMPLAINT INDICATES THAT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22701778 ↗

Other Device, Vascular, For Promoting Embolization devices

All Device, Vascular, For Promoting Embolization devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page