Skip to content

Prestige Coil System PRESPACK65V02Recalls & reported problems

Balt USA LLC · The Prestige Packing Coil consists of an implantable embolization coil comprised of a platinum-tungsten alloy attached to a proximal stainless steel hypo-tube and distal body coil delivery pusher with a radiopaque distal positioning marker and a proximal fluorosafe marker. The Prestige Packing Coil delivery pusher is 185cm in length. It is designed for use with the Ultra Detachment Controller. The Ultra Detachment Controller comes pre-loaded with batteries and is a sterile, handheld, single-patient-use device which should not to be reused, re-sterilized, opened or tampered with. The Prestige Packing Coil and Ultra Detachment Controller(s) are sold separately. The Prestige Packing Coil achieves detachment by an internal heater element, which is powered by the Ultra Detachment Controller. · Class 2 · Implant

All models of Prestige Coil System (97)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00840303712114), received between Feb 6, 2026 and Apr 24, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Premature Separation” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Premature Separation3
  • Failure to Advance2
  • Difficult to Advance1
  • Stretched1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 24, 2026Failure to Advance; Premature Separation

    IT WAS REPORTED THAT: "TWO COMPLEX PRESTIGE COILS WENT INTO A 4.5MM VENOUS COLLATERAL VESSEL AT (B)(6). THE PHYSICIAN WANTED TO PUT IN TWO PACKING COILS BEHIND THROUGH A PROGREAT 2.0 MICRO. THEY FELT RESISTANCE WHEN ADVANCING THROUGH THE MICRO CATHETER AND THE COIL PREDETACHED. THEY COIL COULD HAVE EITHER KNOTTED OR GOTTEN PINCHED IN THE GLIDECATH BASE CATHETER. SAME THING HAPPENED WITH ANOTHER PACKING COIL. SNARED COIL JUST FINE, NO HARM TO THE PATIENT." INCIDENT REPORT FORM SUBMITTED FOR THIS COMPLAINT INDICATES THAT NO PATIENT INJURY WAS SUSTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24983848 ↗
  • MalfunctionApr 24, 2026Failure to Advance; Premature Separation

    IT WAS REPORTED THAT: "TWO COMPLEX PRESTIGE COILS WENT INTO A 4.5MM VENOUS COLLATERAL VESSEL AT DUKE. THE PHYSICIAN WANTED TO PUT IN TWO PACKING COILS BEHIND THROUGH A PROGREAT 2.0 MICRO. THEY FELT RESISTANCE WHEN ADVANCING THROUGH THE MICRO CATHETER AND THE COIL PREDETACHED. THEY COIL COULD HAVE EITHER KNOTTED OR GOTTEN PINCHED IN THE GLIDECATH BASE CATHETER. SAME THING HAPPENED WITH ANOTHER PACKING COIL. SNARED COIL JUST FINE, NO HARM TO THE PATIENT." INCIDENT REPORT FORM SUBMITTED FOR THIS COMPLAINT INDICATES THAT NO PATIENT INJURY WAS SUSTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24983824 ↗
  • MalfunctionFeb 6, 2026Stretched; Difficult to Advance; Premature Separation

    IT WAS REPORTED THAT: "THIS COIL UNRAVELED. WHEN TRYING TO RETRACT AND REPOSITION THE DOCTOR FELT THE COIL DETACH, HE TRIED TO CONTINUE TO PUSH IT INTO THE ANEURYSM AND WHEN HE MET RESISTANCE HE TRIED TO REMOVE THE COIL AND IT BEGAN UNRAVELING. THEY TRIED TO SNARE AND WHEN IT FAILED THE COIL BROKE IN THE CATHETER. THEY FLUSHED THE REST OF THE COIL INTO THE VESSEL AND KEPT THE MICRO AND WHAT WE WERE ABLE TO REMOVE OF THE COIL. THE CASE ENDED UP FINE." INCIDENT REPORT FORM SUBMITTED FOR THIS COMPLAINT INDICATES THAT NO PATIENT INJURY WAS SUSTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24284257 ↗

Other Device, Vascular, For Promoting Embolization devices

All Device, Vascular, For Promoting Embolization devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page