Skip to content

Optima Coil System OPTI0435BLK (REF N/A)Recalls & reported problems

Balt USA LLC · The Optima Coil System is a series of specialized coils that are inserted into the vasculature under angiographic visualization to embolize intracranial aneurysms and other vascular anomalies. The system consists of an embolization coil implant comprised of platinum and tungsten, affixed to a delivery pusher to facilitate insertion into the hub of a microcatheter. The system is available in various shapes, lengths, and sizes. The devices are to be placed into aneurysms to create blood stasis, reducing flow into the aneurysm and thrombosing the aneurysm. Upon positioning coils into the aneurysm, the coils are detached from the delivery pusher in a serial manner until the aneurysm is occluded · Class 2 · Implant

All models of Optima Coil System (379)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00840303712480), received between Mar 11, 2026 and Mar 11, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Premature Separation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Premature Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 11, 2026Premature Separation

    IT WAS REPORTED THAT: "THE PHYSICIAN WAS TREATING A DURAL AVF WITH OPTIBLOCK COILS. WE HAD SUCCESSFULLY DELIVERED (10) OPTIBLOCKS PRIOR TO ISSUE. ON THE 11TH COIL, WE WERE ATTEMPTING TO USE A 4MMX35CM OPTIBLOCK. WE HAD TO ADUST THE COIL POSITION SEVERAL TIMES AS WE COULD NOT GET THE COIL TO SIT EXACTLY WHERE WE WANTED. AFTER MULTIPLE REPOSITIONS, THE COIL PREMATURELY DETATCHED AND WE HAD TO REMOVE THE ENTIRE COIL. THERE WAS NOT A LOT OF VESSEL TOURTOSITY AND THE ONLY THING TO NOTE WOULD BE THE HANDFUL OF TIMES WE HAD TO REPOSITION THE COIL." INCIDENT REPORT FORM SUBMITTED FOR THIS COMPLAINT INDICATES THAT NO PATIENT INJURY WAS SUSTAINED. UPDATE 27FEB2026 - ADDITIONAL INFORMATION RECEIVED FROM THE ISSUER: "COIL WAS AT TREATMENT LOCATION WHEN THIS OCCURRED. THE PHYSICIAN USED A 3CC SYRINGE ON THE BACK OF THE MICRO TO SUCTION BACK COIL UNTIL HE GOT THE COIL ALL INTO MICRO THEN REMOVED THE MICRO CATHETER."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24570775 ↗

Other Device, Vascular, For Promoting Embolization devices

All Device, Vascular, For Promoting Embolization devices →

Sources and updates