Balt USA LLC · The Optima Coil System is a series of specialized coils that are inserted into the vasculature under angiographic visualization to embolize intracranial aneurysms and other vascular anomalies. The system consists of an embolization coil implant comprised of platinum and tungsten, affixed to a delivery pusher to facilitate insertion into the hub of a microcatheter. The system is available in various shapes, lengths, and sizes. The devices are to be placed into aneurysms to create blood stasis, reducing flow into the aneurysm and thrombosing the aneurysm. Upon positioning coils into the aneurysm, the coils are detached from the delivery pusher in a serial manner until the aneurysm is occluded. · Class 2 · Implant
All models of Optima Coil System (379)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00840303713715), received between Mar 6, 2026 and Mar 30, 2026 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Premature Separation” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Premature Separation2
- Failure to Advance1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 30, 2026Failure to Advance; Premature Separation
IT WAS REPORTED THAT: "MMA EMBOLIZATION. PHYSICIAN HAD A 6FR BENCHMARK, PLATO 17 MICROCATHETER, AND A 0.014" WIRE. FRONTAL BRANCH OF RIGHT MMA SELECTED WITH DISTAL ACCESS OF PLATO 17. PHYSICIAN BEGAN TO DEPLOY PAX, FELT SOME RESISTANCE WHILE PACKING AND THEN UPON ATTEMPTING TO MANIPULATE COIL HE FELT THE RESISTANCE GONE AND THE COIL WAS NO LONGER ATTACHED. HE PUSHED THE REST OF THE COIL OUT OF THE MICROCATHETER WIRE A WIRE THROUGH THE PLATO 17 AND IT WAS IN MMA AND THE IMAX." UPDATE 18MAR2026 - ADDITIONAL INFORMATION RECEIVED FROM THE ISSUER: "1.DID THE PROCEDURE INVOLVE TORTUOUS ANATOMY? NOT UNUSUALLY TORTUOUS. 2. WILL THE MICROCATHETER BE AVAILABLE FOR RETURN? NO 3. WHERE DID THE IMPLANT PREMATURELY DETACH? (INSIDE THE AN, INSIDE THE MC?, ETC) ABOUT 10CM OF THE COIL WAS IN THE ANTERIOR MMA BRANCH. THE REMAINING 20CM WAS WITHIN THE PLATO 17 MICROCATHETER AT TIME OF PREMATURE DETACHMENT 4. WERE ANY ATTEMPTS MADE TO DETACH THE IMPLANT COIL USING THE DETACHMENT CONTROLLER? NO 5. CAN YOU CONFIRM IF THE DEVICE WAS IMPLANTED? YES. IT WAS PUSHED OUT WITH A WIRE AND WAS PARTIALLY IN MMA AND IMAX ARTERIES." INCIDENT REPORT FORM SUBMITTED FOR THIS COMPLAINT INDICATES THAT NO PATIENT INJURY
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24716771 ↗MalfunctionMar 6, 2026Premature Separation
IT WAS REPORTED THAT: "THE PHYSICIAN WAS TREATING AN MMA WITHOUT PARTICLES, A COIL ONLY TECHNIQUE. HE WAS COILING THROUGH A PHENOM17 MICROCATHETER. INITIALLY THE COIL TRAVELLED DISTALLY AS EXPECTED, BUT WAS NOT BREAKING PROXIMALLY WHATSOEVER AND THE COIL WAS SO LONG THAT IT WAS KICKING THE CATHETER BACK ALL THE WAY INTO THE IMAX. EVENTUALLY WHILE MANEUVERING THE COIL TO TRY TO GET IT TO BREAK, THE COIL PREMATURELY DETACHED. AFTER MULTIPLE ATTEMPTS TO PUSH IT OUT OF THE MICROCATHETER WITH A WIRE, THE COIL ENDED UP STAYING IN THE IMAX." INCIDENT REPORT FORM SUBMITTED FOR THIS COMPLAINT INDICATES THAT NO PATIENT INJURY WAS SUSTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24539276 ↗
Other Device, Neurovascular Embolization devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page