EVO/EVO+VISIAN Implantable Collamer Lens VICM5 12.1 (-8.0) (REF VICM5 12.1)Recalls & reported problems
Staar Surgical Company · VICM5 - Diopter: -8.0; Overall Diameter (Length): 12.1 · Class 3 · Implant
All models of EVO/EVO+VISIAN Implantable Collamer Lens (1030)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 10
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00840311300235), received between Apr 8, 2023 and May 3, 2026 — classified as 10 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Inadequacy of Device Shape and/or Size” (2 reports).
8 other recalls cover devices cleared under the same FDA submission (P030016); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem9
- Inadequacy of Device Shape and/or Size2
- No Apparent Adverse Event2
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Blurred Vision3
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- Toxic Anterior Segment Syndrome (TASS)1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
InjuryMay 3, 2026Adverse Event Without Identified Device or Use Problem
A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED AND OVER/UNDER CORRECTION. THE LENS WAS LATER EXCHANGED FOR A SAME SIZE, DIFFERENT POWER LENS AND THE PROBLEM WAS RESOLVED. CAUSE OF THE EVENT IS UNKNOWN. THERE ARE MULTIPLE MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS PATIENT. THIS REPORT IS 1 OF 2 TOTAL REPORTS. BASED ON THE INFORMATION PROVIDED, THE RISK ASSESSMENT FOR OUR PRODUCT, AND OUR EXPERIENCE, WE HAVE DETERMINED THE CAUSE OR CONTRIBUTION TO THE REPORTED ISSUE IS NOT INDICATIVE OF A MANUFACTURING RELATED ISSUE OR PRODUCT DEFICIENCY.
Patient problems: Blurred Vision
MAUDE report 25062432 ↗InjuryMay 1, 2026Adverse Event Without Identified Device or Use Problem
A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED AND OVER/UNDER CORRECTION. THE LENS WAS LATER EXCHANGED FOR A SAME SIZE, DIFFERENT POWER LENS AND THE PROBLEM WAS RESOLVED. CAUSE OF THE EVENT IS UNKNOWN.
Patient problems: Blurred Vision
MAUDE report 25062285 ↗InjuryApr 20, 2026Adverse Event Without Identified Device or Use Problem
A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED GLARE/HALOES. THE LENS WAS EXCHANGED. PROBLEM RESOLVED. THERE ARE MULTIPLE MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS PATIENT. THIS REPORT IS 1 OF 2 TOTAL REPORTS. BASED ON THE INFORMATION PROVIDED, THE RISK ASSESSMENT FOR OUR PRODUCT, AND OUR EXPERIENCE, WE HAVE DETERMINED THE CAUSE OR CONTRIBUTION TO THE REPORTED ISSUE IS NOT INDICATIVE OF A MANUFACTURING RELATED ISSUE OR PRODUCT DEFICIENCY. CAUSE OF THE EVENT WAS NOT REPORTED.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 24931619 ↗InjuryApr 20, 2026Adverse Event Without Identified Device or Use Problem
A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED GLARE/HALOES. THE LENS WAS EXCHANGED. PROBLEM RESOLVED. THERE ARE MULTIPLE MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS PATIENT. THIS REPORT IS 1 OF 2 TOTAL REPORTS. BASED ON THE INFORMATION PROVIDED, THE RISK ASSESSMENT FOR OUR PRODUCT, AND OUR EXPERIENCE, WE HAVE DETERMINED THE CAUSE OR CONTRIBUTION TO THE REPORTED ISSUE IS NOT INDICATIVE OF A MANUFACTURING RELATED ISSUE OR PRODUCT DEFICIENCY. CAUSE OF THE EVENT WAS NOT REPORTED.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 24931408 ↗InjurySep 24, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED EXCESSIVE VAULT. IN A SEPARATE VISIT THE LENS WAS EXPLANTED. THE PROBLEM RESOLVED. CAUSE OF THE EVENT WAS REPORTED AS UNKNOWN. THERE ARE MULTIPLE MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS PATIENT. THIS REPORT IS 2 OF 2 TOTAL REPORTS. BASED ON THE INFORMATION PROVIDED, THE RISK ASSESSMENT FOR OUR PRODUCT, AND OUR EXPERIENCE, WE HAVE DETERMINED THAT THE CAUSE OR CONTRIBUTION TO THE REPORTED ISSUE IS NOT INDICATIVE OF A MANUFACTURING RELATED ISSUE OR PRODUCT DEFICIENCY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20297615 ↗InjuryJun 10, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE REPORTER INDICATED THAT A 12.1MM VICM5_12.1 IMPLANTABLE COLLAMER LENS OF -8.00 DIOPTER WAS IMPLANTED INTO THE PATIENTS RIGHT EYE (OD) ON (B)(6) 2023. ON (B)(6) 2023 THE LENS WAS EXPLANTED DUE TO LOW VAULT WITHOUT ROTATION. A REPLACEMENT LENS OF LONGER LENGTH WAS LATER IMPLANTED AND THE PROBLEM RESOLVED. THE CAUSE OF THE EVENT WAS REPORTED AS PATIENT-RELATED FACTOR. HOWEVER, IT WAS ALSO REPORTED THAT THE DEVICE FAILED TO PERFORM AS INTENDED. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NONE HAS BEEN FORTHCOMING. IF ADDITIONAL INFORMATION IS RECEIVED A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19505755 ↗InjuryApr 9, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE REPORTER INDICATED THAT A 12.1MM VICM5 -8.00D IMPLANTABLE COLLAMER LENS WAS IMPLANTED INTO THE PATIENT'S RIGHT EYE (OD) ON AN UNKNOWN DATE. INTRAOPERATIVE CONCOMITANT PRODUCTS USED, INCLUDING OVD AND INJECTOR TYPE, WERE NOT PROVIDED. DAY 1 ONSET OF TOXIC ANTERIOR SEGMENT SYNDROME (TASS) WAS NOTED. DIAGNOSTIC CULTURES WERE NOT PERFORMED, ANTIBIOTICS AND STEROID OPHTHALMIC DROPS AS WELL AS ORAL STEROIDS FOR FIVE DAYS WERE PRESCRIBED. AS REPORTED ON 21-MAR-2024, THE VISION IMPROVED TO 1.5 (~20/15). THE REPORTER DOES NOT STATE A CAUSE FOR THE EVENT. THE LENS REMAINS IMPLANTED.
Patient problems: Toxic Anterior Segment Syndrome (TASS)
MAUDE report 19075298 ↗InjuryMar 28, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE REPORTER INDICATED THAT A 12.1MM VICM5_12.1 IMPLANTABLE COLLAMER LENS OF -8.00 DIOPTER WAS IMPLANTED INTO THE PATIENT'S RIGHT EYE (OD) ON (B)(6) 2023. THE PATIENT EXPERIENCED REFRACTIVE CHANGE OVERTIME. ON (B)(6) 2024 THE LENS WAS EXCHANGED FOR A DIFFERENT POWER LENS AND THE PROBLEM RESOLVED. CAUSE OF EVENT WAS UNKNOWN.
Patient problems: Blurred Vision
MAUDE report 18991698 ↗InjuryDec 1, 2023Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
THE REPORTER INDICATED THAT THE SURGEON IMPLANTED A 12.1MM VICM5_12.1; -8.00 DIOPTER IMPLANTABLE COLLAMER LENS INTO THE PATIENT'S RIGHT EYE (OD) ON (B)(6) 2023. THE PATIENT EXPERIENCED LOW VAULTING. ON 23-NOV-2023 THE LENS WAS EXCHANGED WITH A LONGER LENGTH LENS AND THIS RESOLVED THE PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18248707 ↗MalfunctionAug 2, 2023Adverse Event Without Identified Device or Use Problem; No Apparent Adverse Event; Adverse Event Without Identified Device or Use Problem; No Apparent Adverse Event
THE REPORTER INDICATED THAT A 12.1MM VICM512.1 IMPLANTABLE COLLAMER LENS OF A -8.0 DIOPTER LENS TEAR/BREAK DURING LOADING. REPORTEDLY, "UNSTERILE." ON (B)(6) 2023, THE REPLACEMENT LENS WAS IMPLANTED AND THE PROBLEM WAS RESOLVED. STATUS OF THE EYE IS, "QUIET."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17447292 ↗MalfunctionApr 28, 2023No Apparent Adverse Event; No Apparent Adverse Event
THE REPORTER INDICATED THE SURGEON IMPLANTED A 12.1MM VICM5_12.1 IMPLANTABLE COLLAMER LENS, -08.00 DIOPTER, IN THE PATIENTS LEFT EYE (OS), ON (B)(6) 2023. THIS WAS THE REPLACEMENT LENS FOR MFR. REPORT # 2023826-2023-01499, BUT THE PROBLEM WAS NOT RESOLVED, THERE IS A POSSIBILITY OF REPLACEMENT. THE LENS REMAINED IMPLANTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NONE HAS BEEN FORTHCOMING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16833350 ↗InjuryApr 8, 2023Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
THE REPORTER INDICATED THAT THE SURGEON IMPLANTED A 12.1MM VICM5_12.1 IMPLANTABLE COLLAMER LENS OF DIOPTER -8.0 INTO THE PATIENT'S RIGHT EYE (OD) ON (B)(6) 2023. THE LENS WAS EXCHANGED ON (B)(6) 2023 FOR A LONGER LENGTH LENS DUE TO LOW VAULT. THIS RESOLVED THE PROBLEM. THE REPORTER DID NOT GIVE A CAUSE FOR THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16704985 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P030016. One submission often covers many models, so check the product and lot details in each record.
- Z-1480-2026 · Jan 21, 2026 · Open, ClassifiedEVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5
- Z-0464-2024 · Sep 29, 2023 · Open, ClassifiedEVO+VISIAN Implantable Collamer Lens, REF:VICM5_12.1. Phakic Intraocular lens.
- Z-0466-2024 · Sep 29, 2023 · Open, ClassifiedEVO+VISIAN Implantable Collamer Lens, REF: VICM5_13.2. Phakic Intraocular lens.
- Z-0467-2024 · Sep 29, 2023 · Open, ClassifiedEVO +VISIAN Implantable Collamer Lens, REF: VICM5_13.7.Phakic Intraocular lens.
- Z-0469-2024 · Sep 29, 2023 · Open, ClassifiedEVO VISIAN Implantable Collamer Lens REF: VICMO13.2. Phakic Intraocular lens.
- Z-0468-2024 · Sep 29, 2023 · Open, ClassifiedEVO VISIAN Implantable Collamer Lens, REF: VICMO12.1. Phakic Intraocular lens.
- Z-0474-2024 · Sep 29, 2023 · Open, ClassifiedEVO VISIAN Implantable Collamer Lens, REF: VICMO12.6. Phakic Intraocular lens.
- Z-0465-2024 · Sep 29, 2023 · Open, ClassifiedEVO+VISIAN Implantable Collamer Lens, REF: VICM5_12.6. Phakic Intraocular lens.
Other Lens, Intraocular, Phakic devices
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 13.2 (-8.0)Staar Surgical Company · Lens, Intraocular, Phakic1 recall12 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 13.2 (-11.0)Staar Surgical Company · Lens, Intraocular, Phakic1 recall11 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 13.2 (-5.5)Staar Surgical Company · Lens, Intraocular, Phakic1 recall11 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 13.2 (-8.5)Staar Surgical Company · Lens, Intraocular, Phakic1 recall11 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 12.6 (-8.0)Staar Surgical Company · Lens, Intraocular, Phakic1 recall9 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 13.2 (-9.0)Staar Surgical Company · Lens, Intraocular, Phakic1 recall9 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 12.6 (-9.0)Staar Surgical Company · Lens, Intraocular, Phakic1 recall9 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 12.6 (-9.5)Staar Surgical Company · Lens, Intraocular, Phakic1 recall8 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page