EVO/EVO+VISIAN Implantable Collamer Lens VICM5 12.6 (-3.0) (REF VICM5 12.6)Recalls & reported problems

Staar Surgical Company · VICM5 - Diopter: -3.0; Overall Diameter (Length): 12.6 · Class 3 · Implant

All models of EVO/EVO+VISIAN Implantable Collamer Lens (1030)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00840311300464), received between May 9, 2024 and Sep 16, 2026 — classified as 6 injuries.

8 other recalls cover devices cleared under the same FDA submission (P030016); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem6

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
  • Blurred Vision2
  • No Clinical Signs, Symptoms or Conditions1
  • Visual Disturbances1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjurySep 16, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 50795919 ↗
  • InjuryJun 29, 2026Adverse Event Without Identified Device or Use Problem

    THE REPORTER INDICATED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED OVER/UNDER CORRECTION. THE LENS WAS EXCHANGED FOR A STRONGER POWER, SAME MODEL AND LENGTH LENS. THE PROBLEM WAS RESOLVED. CAUSE OF THE EVENT WAS REPORTED AS PATIENT RELATED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 25660172 ↗
  • InjuryJan 23, 2026Adverse Event Without Identified Device or Use Problem

    A HEALTHCARE PROFESSIONAL REPORTED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED A TRANSIENT LOSS OF BCVA. THE LENS WAS EXCHANGED FOR A SAME LENGTH LENS OF A DIFFERENT POWER & THE PROBLEM WAS RESOLVED. CAUSE OF THE EVENT WAS REPORTED AS UNKNOWN. ADDITIONAL INFORMATION WAS REQUESTED. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME. THIS FILE WILL BE RE-OPENED IF ADDITIONAL INFORMATION IS PROVIDED.

    Patient problems: Blurred Vision

    MAUDE report 24162511 ↗
  • InjuryJun 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED AN UNDER REFRACTION. THE LENS WAS LATER EXCHANGED FOR A SAME SIZE, SAME MODEL, DIFFERENT POWER LENS AND THE PROBLEM WAS RESOLVED. THERE ARE MULTIPLE MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS PATIENT. THIS REPORT IS 1 OF 2 TOTAL REPORTS. BASED ON THE INFORMATION PROVIDED, THE RISK ASSESSMENT FOR OUR PRODUCT, AND OUR EXPERIENCE, WE HAVE DETERMINED THE CAUSE OR CONTRIBUTION TO THE REPORTED ISSUE IS NOT INDICATIVE OF A MANUFACTURING RELATED ISSUE OR PRODUCT DEFICIENCY.

    Patient problems: Visual Disturbances

    MAUDE report 22303230 ↗
  • InjuryMar 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A FACILITY REPRESENTATIVE REPORTED THAT AN IMPLANTABLE COLLAMER LENS WAS IMPLANTED INTO THE PATIENT'S EYE. THE PATIENT EXPERIENCED REFRACTIVE CHANGE OVER TIME. THE LENS WAS EXCHANGED WITH A DIFFERENT POWER LENS AND THE PROBLEM WAS RESOLVED. THE CAUSE OF THE EVENT WAS REPORTED AS UNKNOWN.

    Patient problems: Blurred Vision

    MAUDE report 21635769 ↗
  • InjuryMay 9, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE REPORTER INDICATED THAT A 12.6MM VICM5_12.6 IMPLANTABLE COLLAMER LENS OF -3.00 DIOPTER WAS IMPLANTED INTO THE PATIENTS LEFT EYE (OS) ON (B)(6) 2023. THE PATIENT EXPERIENCED LOW VAULT. ON (B)(6) 2023 THE LENS WAS EXCHANGED FOR A LONGER LENGTH LENS OF DIFFERENT POWER. THIS EXCHANGE RESOLVED THE PROBLEM. THE CAUSE OF THE EVENT WAS REPORTED AS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19281394 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P030016. One submission often covers many models, so check the product and lot details in each record.

  • Z-1480-2026 · Jan 21, 2026 · Open, Classified
    EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5
  • Z-0466-2024 · Sep 29, 2023 · Open, Classified
    EVO+VISIAN Implantable Collamer Lens, REF: VICM5_13.2. Phakic Intraocular lens.
  • Z-0464-2024 · Sep 29, 2023 · Open, Classified
    EVO+VISIAN Implantable Collamer Lens, REF:VICM5_12.1. Phakic Intraocular lens.
  • Z-0469-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens REF: VICMO13.2. Phakic Intraocular lens.
  • Z-0474-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens, REF: VICMO12.6. Phakic Intraocular lens.
  • Z-0467-2024 · Sep 29, 2023 · Open, Classified
    EVO +VISIAN Implantable Collamer Lens, REF: VICM5_13.7.Phakic Intraocular lens.
  • Z-0468-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens, REF: VICMO12.1. Phakic Intraocular lens.
  • Z-0465-2024 · Sep 29, 2023 · Open, Classified
    EVO+VISIAN Implantable Collamer Lens, REF: VICM5_12.6. Phakic Intraocular lens.

Other Lens, Intraocular, Phakic devices

All Lens, Intraocular, Phakic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page