EVO/EVO+VISIAN Implantable Collamer Lens VICM5 12.6 (-3.5) (REF VICM5 12.6)Recalls & reported problems
Staar Surgical Company · VICM5 - Diopter: -3.5; Overall Diameter (Length): 12.6 · Class 3 · Implant
All models of EVO/EVO+VISIAN Implantable Collamer Lens (1030)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 18
- Deaths reported
- 0
- Injuries reported
- 14
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 00840311300471), received between Oct 17, 2023 and Jun 3, 2026 — classified as 14 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Misfocusing” (2 reports).
8 other recalls cover devices cleared under the same FDA submission (P030016); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem13
- No Apparent Adverse Event3
- Misfocusing2
- Off-Label Use1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Toxic Anterior Segment Syndrome (TASS)5
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available3
- Blurred Vision2
- Halo1
- Inflammation1
- Visual Disturbances1
Recent adverse event reports
The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.
InjuryJun 3, 2026Adverse Event Without Identified Device or Use Problem
A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED MILD TOXIC ANTERIOR SEGMENT SYNDROME. TREATMENT WAS PERFORMED AND PROBLEM RESOLVED. LENS REMAINS IMPLANTED. CAUSE OF THE EVENT WAS REPORTED AS OTHER: TASS. THERE ARE MULTIPLE MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS PATIENT. THIS REPORT IS 1 OF 2 TOTAL REPORTS. BASED ON THE INFORMATION PROVIDED, THE RISK ASSESSMENT FOR OUR PRODUCT, AND OUR EXPERIENCE, WE HAVE DETERMINED THE CAUSE OR CONTRIBUTION TO THE REPORTED ISSUE IS NOT INDICATIVE OF A MANUFACTURING RELATED ISSUE OR PRODUCT DEFICIENCY. NO FURTHER INFORMATION HAS BEEN PROVIDED. IF ADDITIONAL INFORMATION IS RECEIVED, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED.
Patient problems: Inflammation
MAUDE report 25379558 ↗InjuryApr 20, 2026Adverse Event Without Identified Device or Use Problem
A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED OVER/UNDER CORRECTION. THE LENS WAS EXCHANGED WITH A DIFFERENT DIOPTER LENS AND PROBLEM WAS RESOLVED. CAUSE OF THE EVENT WAS REPORTED AS UNKNOWN.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 24931426 ↗InjuryApr 17, 2026Adverse Event Without Identified Device or Use Problem
A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED AN OVER/UNDER CORRECTION. THE LENS WAS LATER EXCHANGED FOR A SAME LENGTH, DIFFERENT POWER LENS AND THE PROBLEM WAS RESOLVED. CAUSE OF THE EVENT IS UNKNOWN. THERE ARE MULTIPLE MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS PATIENT. THIS REPORT IS 1 OF 2 TOTAL REPORTS. BASED ON THE INFORMATION PROVIDED, THE RISK ASSESSMENT FOR OUR PRODUCT, AND OUR EXPERIENCE, WE HAVE DETERMINED THE CAUSE OR CONTRIBUTION TO THE REPORTED ISSUE IS NOT INDICATIVE OF A MANUFACTURING RELATED ISSUE OR PRODUCT DEFICIENCY.
Patient problems: Visual Disturbances
MAUDE report 24915993 ↗InjuryMar 23, 2026Adverse Event Without Identified Device or Use Problem
A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED OVER/UNDER CORRECTION. LENS EXHANGED FOR DIFFERENT DIOPTER LENS. PROBLEM RESOLVED. CAUSE OF THE EVENT WAS NOT REPORTED.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 24665585 ↗MalfunctionNov 14, 2024No Apparent Adverse Event; No Apparent Adverse Event
A FACILITY REPRESENTATIVE REPORTED THAT AN IMPLANTABLE COLLAMER LENS WAS IMPLANTED INTO THE PATIENT'S EYE. THE PATIENT EXPERIENCED LOW VAULTING. THE LENS WAS INTRAOPERATIVELY REMOVED. NEXT OPERATION DATE IS NOT YET DECIDED. THE CAUSE OF THE EVENT WAS REPORTED AS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20688290 ↗InjurySep 11, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A HEALTHCARE PROFESSIONAL REPORTED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED HALOS. LENS WAS EXPLANTED. THE PROBLEM WAS RESOLVED. CAUSE OF THE EVENT WAS REPORTED AS PATIENT RELATED. THERE ARE MULTIPLE MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS PATIENT. THIS REPORT IS 1 OF 2 TOTAL REPORTS. BASED ON THE INFORMATION PROVIDED, THE RISK ASSESSMENT FOR OUR PRODUCT, AND OUR EXPERIENCE, WE HAVE DETERMINED THE CAUSE OR CONTRIBUTION TO THE REPORTED ISSUE IS NOT INDICATIVE OF A MANUFACTURING RELATED ISSUE OR PRODUCT DEFICIENCY. ADDITIONAL INFORMATION WAS REQUESTED. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME. THIS FILE WILL BE RE-OPENED IF ADDITIONAL INFORMATION IS PROVIDED.
Patient problems: Halo
MAUDE report 20196628 ↗MalfunctionAug 28, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A HEALTHCARE PROFESSIONAL REPORTED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED EXCESSIVE VAULT AND SIGNIFICANT REDUCTION IN IRIDOCORNEAL ANGLES. LENS REMAINS IMPLANTED. PROBLEM NOT RESOLVED. THERE ARE MULTIPLE MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS PATIENT. THIS REPORT IS 1 OF 2 TOTAL REPORTS. BASED ON THE INFORMATION PROVIDED, THE RISK ASSESSMENT FOR OUR PRODUCT, AND OUR EXPERIENCE, WE HAVE DETERMINED THE CAUSE OR CONTRIBUTION TO THE REPORTED ISSUE IS NOT INDICATIVE OF A MANUFACTURING RELATED ISSUE OR PRODUCT DEFICIENCY. CAUSE OF THE EVENT WAS REPORTED AS DEVICE; DEVICE DID NOT FAIL TO PERFORM AS INTENDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20088037 ↗MalfunctionAug 28, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A HEALTHCARE PROFESSIONAL REPORTED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED EXCESSIVE VAULT AND SIGNIFICANT REDUCTION IN IRIDOCORNEAL ANGLES. LENS REMAINS IMPLANTED. PROBLEM NOT RESOLVED. THERE ARE MULTIPLE MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS PATIENT. THIS REPORT IS 1 OF 2 TOTAL REPORTS. BASED ON THE INFORMATION PROVIDED, THE RISK ASSESSMENT FOR OUR PRODUCT, AND OUR EXPERIENCE, WE HAVE DETERMINED THE CAUSE OR CONTRIBUTION TO THE REPORTED ISSUE IS NOT INDICATIVE OF A MANUFACTURING RELATED ISSUE OR PRODUCT DEFICIENCY. CAUSE OF THE EVENT WAS REPORTED AS DEVICE; DEVICE DID NOT FAIL TO PERFORM AS INTENDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20088002 ↗InjuryJun 21, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE REPORTER INDICATED THAT A 12.6MM VICM5_ 12.6 IMPLANTABLE COLLAMER LENS OF -3.50 DIOPTER WAS IMPLANTED INTO THE PATIENTS RIGHT EYE (OD) ON (B)(6) 2024.THE PATIENT EXPERIENCED EXCESSIVE VAULT. ON (B)(6) 2024 THE LENS WAS EXCHANGED FOR A SHORTER LENGTH LENS AND THE PROBLEM RESOLVED. IN THE REPORTERS OPINION THE CAUSE OF THE EVENT WAS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19580484 ↗InjuryMay 14, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE REPORTER INDICATED THAT A 12.6MM VICM5 -3.50D IMPLANTABLE COLLAMER LENS WAS IMPLANTED INTO THE PATIENT'S RIGHT EYE (OD) ON (B)(6) 2024. CONCOMITANT PRODUCTS USED INTRAOPERATIVELY INCLUDE OPEGAN VISCOELASTIC AND THE MSI INJECTOR SYSTEM. TOXIC ANTERIOR SEGMENT SYNDROME (TASS) WAS DIAGNOSED. NO ADDITIONAL DIAGNOSTICS WERE PERFORMED. ANTIBIOTIC AND STEROID OPHTHALMIC DROPS WERE PRESCRIBED. DIFFUSE DEPOSITS ON THE POSTERIOR CORNEA AND INFLAMMATORY OPACITY ON THE LENS SURFACE HAVE IMPROVED "BUT STILL SLIGHTLY CLOUDY". PER LAST VISIT CONDUCTED ON (B)(6) 2024, VISUAL ACUITY WAS REPORTED TO BE GOOD, THE BCVA WAS 20/13 EQUIVALENT TO PRE-OP. PER SAME VISIT IOP WAS 13 MMHG AND VAULT WAS 1.4 CT. THE LENS REMAINS IMPLANTED.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Toxic Anterior Segment Syndrome (TASS)
MAUDE report 19312318 ↗InjuryMay 6, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE REPORTER INDICATED THAT THE SURGEON IMPLANTED A 12.6MM VICM5 -3.50D IMPLANTABLE COLLAMER LENS INTO THE PATIENT'S RIGHT EYE (OD) ON (B)(6) 2024. THE PATIENT UNDERWENT BILATERAL SEQUENTIAL SURGERY. CONCOMITANT PRODUCTS USED INTRAOPERATIVELY INCLUDE OPEGAN VISCOELASTIC AND THE MSI INJECTOR SYSTEM. DIAGNOSTICS WERE NOT PERFORMED; HOWEVER, ON DAY 3 FIBRIN WAS OBSERVED BY OCT "SO TASS SUSPECTED". STEROID EYE DROPS WERE PRESCRIBED AND SYMPTOMS SUBSIDED. REPORTEDLY BY DAY 7 THE PROBLEM RESOLVED AND THE LENS REMAINS IMPLANTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NONE HAS BEEN FORTHCOMING.
Patient problems: Toxic Anterior Segment Syndrome (TASS)
MAUDE report 19256122 ↗InjuryMay 6, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE REPORTER INDICATED THAT A 12.6MM VICM5 -3.50D IMPLANTABLE COLLAMER LENS WAS IMPLANTED INTO THE PATIENT'S RIGHT EYE (OD) ON (B)(6) 2024. CONCOMITANT PRODUCTS USED INTRAOPERATIVELY INCLUDE OPEGAN VISCOELASTIC AND THE MSI INJECTOR SYSTEM. TOXIC ANTERIOR SEGMENT SYNDROME (TASS) WAS DIAGNOSED. DIAGNOSTICS WERE REPORTEDLY PERFORMED WITH RESULTS OF "FOUND DEPOSITS ON THE IMPLANTED LENS." PATIENT COMPLAINED OF BLURRED VISION. ON DAY 3 THE LENS WAS EXPLANTED AND THE PATIENT REMAINS UNDER OBSERVATION. IN THE REPORTER'S OPINION THE CAUSE OF THE EVENT IS UNKNOWN. THE LENS WAS RETURNED ON (B)(6) 2024.
Patient problems: Toxic Anterior Segment Syndrome (TASS)
MAUDE report 19255919 ↗InjuryApr 19, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE REPORTER INDICATED THAT A 12.6MM VICM5 12.6 IMPLANTABLE COLLAMER LENS OF -3.50 DIOPTER WAS IMPLANTED INTO THE PATIENT'S RIGHT EYE (OD) ON (B)(6) 2023. PATIENT COMPLAINED OF DIMINISHED VISION. ON (B)(6) 2023 THE LENS WAS REMOVED AND ON THE SAME DAY A REPLACEMENT LENS VICM5 12.6 OF -3.00 DIOPTER WAS IMPLANTED. THIS REPLACEMENT LENS RESOLVED THE PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19149694 ↗InjuryApr 4, 2024No Apparent Adverse Event; No Apparent Adverse Event
THE REPORTER INDICATED THAT A 12.6MM VICM512.6 IMPLANTABLE COLLAMER LENS OF -3.50 DIOPTER WAS IMPLANTED INTO THE PATIENT'S LEFT EYE (OS) ON 21-FEB-2024. CONCOMITANT PRODUCTS USED INTRAOPERATIVELY INCLUDE OPEGAN VISCOELASTIC AND THE MSI INJECTOR SYSTEM. ON THE SAME DAY, TOXIC ANTERIOR SEGMENT SYNDROME (TASS) WAS OBSERVED. MANAGEMENT INCLUDED STEROID "MEDICATION, EYE DROPS" AND SUBCONJUNCTIVAL INJECTION. ON 07-MAR-2024 BCVA WAS NOTED AT 20/13 WITH THE ISSUE RESOLVED. IN THE REPORTER'S OPINION THE CAUSE OF THE EVENT IS UNKNOWN. THE LENS REMAINS IMPLANTED. THE PROBLEM WAS RESOLVED.
Patient problems: Toxic Anterior Segment Syndrome (TASS)
MAUDE report 19047485 ↗InjuryMar 28, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE REPORTER INDICATED THAT A 12.6MM VICM5_12.6 IMPLANTABLE COLLAMER LENS OF -3.5 DIOPTER WAS IMPLANTED INTO THE "PATIENT'S" RIGHT EYE (OD) ON (B)(6) 2024. FIBRIN (+) IN THE ANTERIOR CHAMBER AND NUMEROUS DEPOSITS ON THE FRONT AND BACK SURFACE OF THE LENS WAS REPORTED. THE ANTERIOR CHAMBER WAS WASHED AND SUBCONJUNCTIVAL INJECTION OF STEROIDS WAS GIVEN. LENS REMAINS IMPLANTED.
Patient problems: Toxic Anterior Segment Syndrome (TASS)
MAUDE report 18999953 ↗InjuryFeb 27, 2024Misfocusing; Misfocusing
THE REPORTER INDICATED THAT A 12.6MM, VICM5_12.6, -3.5 DIOPTER, IMPLANTABLE COLLAMER LENS WAS IMPLANTED INTO THE PATIENT'S RIGHT EYE (OD) ON (B)(6) 2023. ON (B)(6) /2023 THE LENS WAS REPLACED WITH A SAME SIZE , DIFFERENT DIOPTER TORIC LENS DUE TO REFRACTIVE SURPRISE. THE PROBLEM WAS RESOLVED.
Patient problems: Blurred Vision
MAUDE report 18788738 ↗InjuryNov 10, 2023Misfocusing; Off-Label Use; Misfocusing; Off-Label Use
THE REPORTER INDICATED THAT A 12.6MM VICM5 12.6 IMPLANTABLE COLLAMER LENS OF -3.50 DIOPTER WAS IMPLANTED INTO THE PATIENT'S RIGHT EYE (OD) ON (B)(6) 2023. REFRACTIVE CHANGE OVER TIME WAS OBSERVED. ON (B)(6) 2023 THE LENS WAS EXCHANGED FOR A SPHERICAL LENS OF THE SAME SIZE AND LRI WAS PERFORMED FOR ASTIGMATISM. THIS RESOLVED THE PROBLEM.
Patient problems: Blurred Vision
MAUDE report 18116294 ↗MalfunctionOct 17, 2023No Apparent Adverse Event; No Apparent Adverse Event
A FAULTY 12.6MM VICM5_12.6 IMPLANTABLE COLLAMER LENS OF -3.50 DIOPTER WAS REPORTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NONE HAS BEEN FORTHCOMING. IF ADDITIONAL INFORMATION IS RECEIVED A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17954474 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P030016. One submission often covers many models, so check the product and lot details in each record.
- Z-1480-2026 · Jan 21, 2026 · Open, ClassifiedEVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5
- Z-0466-2024 · Sep 29, 2023 · Open, ClassifiedEVO+VISIAN Implantable Collamer Lens, REF: VICM5_13.2. Phakic Intraocular lens.
- Z-0468-2024 · Sep 29, 2023 · Open, ClassifiedEVO VISIAN Implantable Collamer Lens, REF: VICMO12.1. Phakic Intraocular lens.
- Z-0474-2024 · Sep 29, 2023 · Open, ClassifiedEVO VISIAN Implantable Collamer Lens, REF: VICMO12.6. Phakic Intraocular lens.
- Z-0467-2024 · Sep 29, 2023 · Open, ClassifiedEVO +VISIAN Implantable Collamer Lens, REF: VICM5_13.7.Phakic Intraocular lens.
- Z-0469-2024 · Sep 29, 2023 · Open, ClassifiedEVO VISIAN Implantable Collamer Lens REF: VICMO13.2. Phakic Intraocular lens.
- Z-0465-2024 · Sep 29, 2023 · Open, ClassifiedEVO+VISIAN Implantable Collamer Lens, REF: VICM5_12.6. Phakic Intraocular lens.
- Z-0464-2024 · Sep 29, 2023 · Open, ClassifiedEVO+VISIAN Implantable Collamer Lens, REF:VICM5_12.1. Phakic Intraocular lens.
Other Lens, Intraocular, Phakic devices
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 13.2 (-8.0)Staar Surgical Company · Lens, Intraocular, Phakic1 recall12 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 13.2 (-11.0)Staar Surgical Company · Lens, Intraocular, Phakic1 recall11 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 13.2 (-5.5)Staar Surgical Company · Lens, Intraocular, Phakic1 recall11 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 13.2 (-8.5)Staar Surgical Company · Lens, Intraocular, Phakic1 recall11 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 12.6 (-8.0)Staar Surgical Company · Lens, Intraocular, Phakic1 recall9 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 13.2 (-9.0)Staar Surgical Company · Lens, Intraocular, Phakic1 recall9 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 12.6 (-9.0)Staar Surgical Company · Lens, Intraocular, Phakic1 recall9 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 12.6 (-9.5)Staar Surgical Company · Lens, Intraocular, Phakic1 recall8 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page