EVO/EVO+VISIAN Implantable Collamer Lens VICM5 12.6 (-11.5) (REF VICM5 12.6)Recalls & reported problems
Staar Surgical Company · VICM5 - Diopter: -11.5; Overall Diameter (Length): 12.6 · Class 3 · Implant
All models of EVO/EVO+VISIAN Implantable Collamer Lens (1030)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00840311300631), received between May 17, 2024 and Aug 26, 2026 — classified as 8 injuries, 2 malfunctions.
8 other recalls cover devices cleared under the same FDA submission (P030016); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem10
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Blurred Vision4
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- Cataract1
- Intraocular Pressure Increased1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
InjuryAug 26, 2026Adverse Event Without Identified Device or Use Problem
A facility representative reported that following an implantable collamer lens (ICL) implantation procedure, the patient experienced over/under correction. Lens was exchanged for a different power lens and problem was resolved. Cause of the event was reported as unknown.
Patient problems: Blurred Vision
MAUDE report 50464693 ↗InjuryDec 26, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE THE PATIENT EXPERIENCED ELEVATED IOP 15 MMHG WITH ANGLE CLOSURE, EXCESSIVE VAULT, SIGNIFICANT REDUCTION OF IRIDOCORNEAL ANGLES AND A SHALLOWING OF THE AC. MEDICATION ADMINISTERED," RESOLVED AFTER TREATMENT," AND LENS REMAINS IMPLANTED. CAUSE OF THE EVENT WAS REPORTED AS DEVICE.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Intraocular Pressure Increased
MAUDE report 23898613 ↗InjuryNov 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THE LENS WAS IMPLANTED UPSIDE DOWN. THERE WAS PATIENT CONTACT. THERE WAS NO PATIENT INJURY. IN A SEPARATE SURGERY THE LENS WAS EXCHANGED WITH THE SAME MODEL/ LENGTH, DIFFERENT POWER AND THE PROBLEM WAS RESOLVED. REPORTEDLY, "THE REASON FOR THE CHANGE IN POWER IS BY PATIENT'S REQUEST." THE CAUSE OF THE EVENT WAS REPORTED AS UNKNOWN. THERE ARE MULTIPLE MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS PATIENT. THIS REPORT IS 1 OF 2 TOTAL REPORTS. BASED ON THE INFORMATION PROVIDED, THE RISK ASSESSMENT FOR OUR PRODUCT, AND OUR EXPERIENCE, WE HAVE DETERMINED THE CAUSE OR CONTRIBUTION TO THE REPORTED ISSUE IS NOT INDICATIVE OF A MANUFACTURING RELATED ISSUE OR PRODUCT DEFICIENCY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23610206 ↗InjurySep 15, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE THE PATIENT EXPERIENCED REFRACTIVE SURPRISE THE LENS EXPLANTED. THE CAUSE OF EVENT WAS UNKNOWN.
Patient problems: Blurred Vision
MAUDE report 23059532 ↗InjuryFeb 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) PROCEDURE THE PATIENT EXPERIENCED LOW VAULTING, BLURRED VISION AND CATARACT DEVELOPED AND REMOVED. LENS EXPLANTED AND REPLACED AND THE PROBLEM WAS RESOLVED. CAUSE OF THE EVENT IS UNKNOWN. THIS REPORT IS 2 OF 2 TOTAL REPORTS. BASED ON THE INFORMATION PROVIDED, THE RISK ASSESSMENT FOR OUR PRODUCT, AND OUR EXPERIENCE, WE HAVE DETERMINED THE CAUSE OR CONTRIBUTION TO THE REPORTED ISSUE IS NOT INDICATIVE OF A MANUFACTURING RELATED ISSUE OR PRODUCT DEFICIENCY.
Patient problems: Blurred Vision; Cataract
MAUDE report 21415437 ↗InjuryJan 10, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED A REFRACTIVE SURPRISE. IN A SEPARATE VISIT THE LENS WAS EXCHANGED FOR THE SAME MODEL, DIFFERENT POWER LENS AND THE PROBLEM WAS RESOLVED. THE CAUSE OF THE EVENT WAS REPORTED AS UNKNOWN. IT HAS BEEN NOTED THAT THE SURGEON SELECTED A DIFFERENT LENS SIZE THAN THAT INITIALLY SUGGESTED BY THE ICL CALCULATION SOFTWARE. THEREFORE, THERE IS NOT ENOUGH SAFETY AND EFFECTIVENESS DATA TO SUPPORT IMPLANTATION IN THIS PATIENT CATEGORY.
Patient problems: Blurred Vision
MAUDE report 21129541 ↗MalfunctionJul 8, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE HEALTH CARE PROFESSIONAL REPORTED THAT PRIOR TO AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THERE WAS PRODUCT NON-CONFORMITY. THE PROBLEM WAS DESCIRBED AS "FOREIGN SUBSTANCE_BLACK DOT". THE LENS REMAINS IMPLANTED. STATUS OF THE EYE IS "CLEAR". THE CAUSE OF EVENT IS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19697452 ↗MalfunctionJul 2, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE REPORTER INDICATED THE SURGEON IMPLANTED A 12.6MM VICM5_ 12.6 IMPLANTABLE COLLAMER LENS OF DIOPTER -11.50 INTO THE LEFT EYE (OS) ON (B)(6) 2024. THE PATIENT EXPERIENCED A LOW VAULT AND SIGNIFICANT REDUCTION OF IRIDOCORNEAL ANGLES. THE LENS WAS IMPLANTED, REMOVED AND REPLACED INTRAOPERATIVELY WITH A LENS OF THE SAME MODEL, LONGER LENGTH AND THE PROBLEM WAS RESOLVED. THE REPORTER INDICATED THE CAUSE OF THE EVENT, THE DEVICE.CAUSE DETAILS PROVIDED: "LENS SIZE OF 12.6 IMPLANTED, AS RECOMMENDED BY THE REINSTEIN FORMULA. THE INTRAOPERATIVE VAULT(MEASURED USING ARTEVO IOCT) WAS LOW (-250 UM) EXCHANGE TO 13.2 WAS PERFORMED WITH VAULT RETURNING TO A REASONABLE HEIGHT -800)".
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; No Clinical Signs, Symptoms or Conditions
MAUDE report 19657928 ↗InjuryJun 4, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE REPORTER INDICATED THAT THE SURGEON IMPLANTED A 12.6MM VICM5 12.6, -11.50 DIOPTER, IMPLANTABLE COLLAMER LENS INTO THE PATIENT'S LEFT EYE (OS) ON (B)(6) 2024. THE LENS WAS REPLACED WITH A LONGER LENGTH LENS DUE TO LOW VAULT ON (B)(6) 2024. THIS RESOLVED THE PROBLEM. CAUSE OF THE EVENT IS REPORTED AS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19457857 ↗InjuryMay 17, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE REPORTER INDICATED THAT THE SURGEON IMPLANTED A 12.6MM VICM5_12.6 IMPLANTABLE COLLAMER LENS OF DIOPTER -11.50 INTO THE PATIENT'S LEFT EYE (OS) ON (B)(6) 2024. ON (B)(6) 2024 THE LENS WAS EXPLANTED DUE TO LOW VAULT WITHOUT ROTATION. ON (B)(6) 2024 A REPLACEMENT LENS OF A LONGER LENGTH WAS IMPLANTED. REPORTEDLY, "PERSISTS LOW VAULT. PATIENT IN TREATMENT OF OCULAR HYPERTENSION." THE PROBLEM WAS NOT RESOLVED. THE CAUSE OF THE EVENT WAS REPORTED AS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19347507 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P030016. One submission often covers many models, so check the product and lot details in each record.
- Z-1480-2026 · Jan 21, 2026 · Open, ClassifiedEVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5
- Z-0468-2024 · Sep 29, 2023 · Open, ClassifiedEVO VISIAN Implantable Collamer Lens, REF: VICMO12.1. Phakic Intraocular lens.
- Z-0464-2024 · Sep 29, 2023 · Open, ClassifiedEVO+VISIAN Implantable Collamer Lens, REF:VICM5_12.1. Phakic Intraocular lens.
- Z-0469-2024 · Sep 29, 2023 · Open, ClassifiedEVO VISIAN Implantable Collamer Lens REF: VICMO13.2. Phakic Intraocular lens.
- Z-0467-2024 · Sep 29, 2023 · Open, ClassifiedEVO +VISIAN Implantable Collamer Lens, REF: VICM5_13.7.Phakic Intraocular lens.
- Z-0465-2024 · Sep 29, 2023 · Open, ClassifiedEVO+VISIAN Implantable Collamer Lens, REF: VICM5_12.6. Phakic Intraocular lens.
- Z-0474-2024 · Sep 29, 2023 · Open, ClassifiedEVO VISIAN Implantable Collamer Lens, REF: VICMO12.6. Phakic Intraocular lens.
- Z-0466-2024 · Sep 29, 2023 · Open, ClassifiedEVO+VISIAN Implantable Collamer Lens, REF: VICM5_13.2. Phakic Intraocular lens.
Other Lens, Intraocular, Phakic devices
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 13.2 (-8.0)Staar Surgical Company · Lens, Intraocular, Phakic1 recall12 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 13.2 (-11.0)Staar Surgical Company · Lens, Intraocular, Phakic1 recall11 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 13.2 (-5.5)Staar Surgical Company · Lens, Intraocular, Phakic1 recall11 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 13.2 (-8.5)Staar Surgical Company · Lens, Intraocular, Phakic1 recall11 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 12.6 (-8.0)Staar Surgical Company · Lens, Intraocular, Phakic1 recall9 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 13.2 (-9.0)Staar Surgical Company · Lens, Intraocular, Phakic1 recall9 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 12.6 (-9.0)Staar Surgical Company · Lens, Intraocular, Phakic1 recall9 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 12.6 (-9.5)Staar Surgical Company · Lens, Intraocular, Phakic1 recall8 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page