EVO/EVO+VISIAN Implantable Collamer Lens VICM5 13.2 (-5.5) (REF VICM5 13.2)Recalls & reported problems
Staar Surgical Company · VICM5 - Diopter: -5.5; Overall Diameter (Length): 13.2 · Class 3 · Implant
All models of EVO/EVO+VISIAN Implantable Collamer Lens (1030)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 16
- Deaths reported
- 0
- Injuries reported
- 14
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 00840311300846), received between Mar 10, 2023 and Mar 30, 2026 — classified as 14 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Inadequacy of Device Shape and/or Size” (5 reports).
8 other recalls cover devices cleared under the same FDA submission (P030016); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem10
- Inadequacy of Device Shape and/or Size5
- Misfocusing3
- Device Dislodged or Dislocated1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Blurred Vision4
- Uveitis2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Intraocular Pressure Increased1
- Toxic Anterior Segment Syndrome (TASS)1
- Visual Disturbances1
Recent adverse event reports
The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.
InjuryMar 30, 2026Adverse Event Without Identified Device or Use Problem
A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED A REFRACTIVE SURPRISE. THE LENS WAS EXCHANGED FOR A SAME LENGTH LENS OF A DIFFERENT POWER, AND THE PROBLEM WAS RESOLVED. CAUSE OF THE EVENT WAS REPORT AS PATIENT RELATED FACTOR AND "OTHER", "PATIENT WAS UNABLE TO TOLERATE BLENDED VISION." AND "HE IS BLENDED/MONOVISION." ADDITIONAL INFORMATION WAS REQUESTED. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME. THIS FILE WILL BE RE-OPENED IF ADDITIONAL INFORMATION IS PROVIDED.
Patient problems: Visual Disturbances
MAUDE report 24721495 ↗InjuryDec 12, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED REFRACTIVE SURPRISE. IN A SEPARATE VISIT THE LENS WAS EXCHANGED FOR A SAME SIZE LENS BUT DIFFERENT POWER. THE REPLACEMENT LENS RESOLVED THE PROBLEM. CAUSE REPORTED AS UNKNOWN.
Patient problems: Blurred Vision
MAUDE report 23792238 ↗InjuryNov 25, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED A REFRACTIVE SURPRISE. IN A SEPARATE VISIT THE LENS WAS EXCHANGED FOR A SAME MODEL/SIZE BUT DIFFERENT POWER LENS. THE PROBLEM WAS RESOLVED. CAUSE REPORTED AS UNKNOWN. THIS REPORT IS 1 OF 2 TOTAL REPORTS. BASED ON THE INFORMATION PROVIDED, THE RISK ASSESSMENT FOR OUR PRODUCT, AND OUR EXPERIENCE, WE HAVE DETERMINED THE CAUSE OR CONTRIBUTION TO THE REPORTED ISSUE IS NOT INDICATIVE OF A MANUFACTURING RELATED ISSUE OR PRODUCT DEFICIENCY.
Patient problems: Blurred Vision
MAUDE report 23649867 ↗InjuryNov 5, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A FACILITY REPRESENTATIVE REPORTED FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED EXCESSIVE VAULT. THE LENS WAS REMOVED AND EXCHANGED INTRAOPERATIVELY FOR A SHORTER LENGTH LENS. CAUSE OF THE EVENT WAS REPORTED AS UNKNOWN. THE PROBLEM WAS RESOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20606100 ↗InjuryAug 9, 2024Misfocusing; Adverse Event Without Identified Device or Use Problem; Misfocusing; Adverse Event Without Identified Device or Use Problem
A FACILITY REPRESENTATIVE REPORTED THAT AN IMPLANTABLE COLLAMER LENS WAS IMPLANTED INTO THE PATIENT'S EYE. THE PATIENT EXPERIENCED REFRACTIVE SURPRISE. THE LENS WAS EXCHANGED WITH THE SAME LENGTH LENS BUT TORIC MODEL. THIS RESOLVED THE PROBLEM.
Patient problems: Blurred Vision
MAUDE report 19946875 ↗InjuryJun 29, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE REPORTER INDICATED THAT THE SURGEON IMPLANTED A 13.2MM VICM5_13.2; -5.50 DIOPTER IMPLANTABLE COLLAMER LENS INTO THE PATIENT'S RIGHT EYE (OD) ON (B)(6) 2024. ON (B)(6) 2024 THE LENS WAS EXCHANGED FOR A SHORTER LENGTH LENS DUE TO EXCESSIVE VAULTING AND THE PROBLEM RESOLVED. THE CAUSE OF THE EVENT WAS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19640500 ↗InjuryMay 2, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE REPORTER INDICATED THAT A 13.2MM VICM5 13.2 IMPLANTABLE COLLAMER LENS OF -5.50 DIOPTER WAS IMPLANTED INTO THE PATIENTS LEFT EYE (OS) ON (B)(6) 2024. EXCESSIVE VAULT AND PIGMENT DISPERSION WAS REPORTED. ON (B)(6) 2024 THE LENS WAS EXPLANTED AND THE PROBLEM WAS RESOLVED.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 19234501 ↗InjuryFeb 5, 2024Inadequacy of Device Shape and/or Size; Device Dislodged or Dislocated; Inadequacy of Device Shape and/or Size; Device Dislodged or Dislocated
THE REPORTER INDICATED THAT A 13.2MM VICM5_13.2 IMPLANTABLE COLLAMER LENS OF -5.50 DIOPTER WAS IMPLANTED INTO THE PATIENT'S LEFT EYE (OS) ON (B)(6) 2023. THE REPORTER REPORTS OBSERVATION OF EXCESSIVE VAULT AND ELEVATED INTRAOCULAR PRESSURE. LENS REMAINS IMPLANTED. IF ADDITIONAL INFORMATION IS RECEIVED A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED.
Patient problems: Intraocular Pressure Increased
MAUDE report 18645384 ↗InjuryDec 29, 2023Misfocusing; Misfocusing
THE REPORTER INDICATED THAT A 13.2MM VICM513.2 IMPLANTABLE COLLAMER LENS OF -5.50 DIOPTER WAS IMPLANTED INTO THE PATIENT'S RIGHT EYE (OD) ON (B)(6) 2023. ON 16-SEP-2023, THE LENS WAS REMOVED DUE TO TOXIC ANTERIOR SEGMENT SYNDROME (TASS). THE PROBLEM WAS RESOLVED. CAUSE OF THE EVENT IS UNKNOWN.
Patient problems: Toxic Anterior Segment Syndrome (TASS)
MAUDE report 18421051 ↗InjuryOct 5, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE SURGEON REPORTS A CASE OF UVEITIS THREE WEEKS AFTER SURGERY. PER UPDATED INFORMATION THE PATIENT IS DOING WELL AND THE TREATMENT ( EYE DROPS) HAS SOLVED THE ISSUE. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NONE HAS BEEN FORTHCOMING. IF ADDITIONAL INFORMATION IS RECEIVED A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED.
Patient problems: Uveitis
MAUDE report 17877885 ↗InjuryOct 5, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE SURGEON REPORTS A CASE OF UVEITIS THREE WEEKS AFTER SURGERY. PER UPDATED INFORMATION THE PATIENT IS DOING WELL AND THE TREATMENT ( EYE DROPS) HAS SOLVED THE ISSUE. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NONE HAS BEEN FORTHCOMING. IF ADDITIONAL INFORMATION IS RECEIVED A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED.
Patient problems: Uveitis
MAUDE report 17877870 ↗InjurySep 6, 2023Misfocusing; Adverse Event Without Identified Device or Use Problem; Misfocusing; Adverse Event Without Identified Device or Use Problem
THE REPORTER INDICATED THAT THE SURGEON IMPLANTED A 13.2MM VICM5_13.2, -5.50 DIOPTER IMPLANTABLE COLLAMER LENS INTO THE PATIENT'S LEFT EYE (OS) ON 20-JUL-2023. IT WAS REPORTED THAT THE PATIENT HAS REFRACTIVE SURPRISE. THE SURGEON PERFORMED LRI SURGERY. LENS REMAINS IMPLANTED. IT WAS REPORTED THAT THE CAUSE OF THE EVENT WAS REPORTED AS A USER ERROR WITH NO DEVICE CAUSALITY AND NOTED "SURGEON PLANNED ASTIGMATISM CORRECTION WITH LRI WHICH DOESN'T WORK". IF ADDITIONAL INFORMATION IS RECEIVED A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED.
Patient problems: Blurred Vision
MAUDE report 17694989 ↗MalfunctionJul 13, 2023Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
THE REPORTER INDICATED THE SURGEON IMPLANTED A 13.2MM ICL IMPLANTABLE COLLAMER LENS, INTO THE PATIENTS LEFT EYE (OS). THE LENS WAS REPORTED AS HIGH VAULTING AND REMAINED IMPLANTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NONE HAS BEEN FORTHCOMING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17322120 ↗MalfunctionJun 27, 2023Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
THE REPORTER INDICATED THAT A 13.2MM VICM513.2 IMPLANTABLE COLLAMER LENS OF 5.5 DIOPTER WAS IMPLANTED INTO THE PATIENT'S RIGHT EYE (OD) ON (B)(6)2023. THE LENS WAS INTRAOPERATIVELY IMPLANTED, REMOVED, AND REPLACED FOR A SHORTER LENGTH LENS DUE TO EXCESSIVE VAULT. THE CAUSE OF THE EVENT IS UNKNOWN AND THE PROBLEM WAS RESOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17213815 ↗InjuryMar 10, 2023Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
THE REPORTER INDICATED THAT THE SURGEON IMPLANTED A 13.2MM VICM5_13.2 IMPLANTABLE COLLAMER LENS OF DIOPTER -5.50 INTO THE PATIENT'S RIGHT EYE (OD) ON (B)(6) 2022. THE LENS WAS EXCHANGED ON (B)(6) 2022 FOR A SHORTER LENGTH LENS DUE TO EXCESSIVE VAULT. THIS RESOLVED THE PROBLEM. THE REPORTER DID NOT INDICATE THE CAUSE OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16520001 ↗InjuryMar 10, 2023Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
THE REPORTER INDICATED THAT THE SURGEON IMPLANTED A 13.2MM VICM5 13.2 IMPLANTABLE COLLAMER LENS OF DIOPTER -5.50 INTO THE PATIENT'S LEFT EYE (OS) ON (B)(6) 2022. THE LENS WAS EXCHANGED ON (B)(6) 2022 FOR A SHORTER LENGTH LENS DUE TO EXCESSIVE VAULT. THIS RESOLVED THE PROBLEM. THE REPORTER DID NOT INDICATE THE CAUSE OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16519997 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P030016. One submission often covers many models, so check the product and lot details in each record.
- Z-1480-2026 · Jan 21, 2026 · Open, ClassifiedEVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5
- Z-0464-2024 · Sep 29, 2023 · Open, ClassifiedEVO+VISIAN Implantable Collamer Lens, REF:VICM5_12.1. Phakic Intraocular lens.
- Z-0474-2024 · Sep 29, 2023 · Open, ClassifiedEVO VISIAN Implantable Collamer Lens, REF: VICMO12.6. Phakic Intraocular lens.
- Z-0467-2024 · Sep 29, 2023 · Open, ClassifiedEVO +VISIAN Implantable Collamer Lens, REF: VICM5_13.7.Phakic Intraocular lens.
- Z-0468-2024 · Sep 29, 2023 · Open, ClassifiedEVO VISIAN Implantable Collamer Lens, REF: VICMO12.1. Phakic Intraocular lens.
- Z-0466-2024 · Sep 29, 2023 · Open, ClassifiedEVO+VISIAN Implantable Collamer Lens, REF: VICM5_13.2. Phakic Intraocular lens.
- Z-0469-2024 · Sep 29, 2023 · Open, ClassifiedEVO VISIAN Implantable Collamer Lens REF: VICMO13.2. Phakic Intraocular lens.
- Z-0465-2024 · Sep 29, 2023 · Open, ClassifiedEVO+VISIAN Implantable Collamer Lens, REF: VICM5_12.6. Phakic Intraocular lens.
Other Lens, Intraocular, Phakic devices
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 13.2 (-8.0)Staar Surgical Company · Lens, Intraocular, Phakic1 recall12 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 13.2 (-11.0)Staar Surgical Company · Lens, Intraocular, Phakic1 recall11 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 13.2 (-5.5)Staar Surgical Company · Lens, Intraocular, Phakic1 recall11 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 13.2 (-8.5)Staar Surgical Company · Lens, Intraocular, Phakic1 recall11 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 12.6 (-8.0)Staar Surgical Company · Lens, Intraocular, Phakic1 recall9 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 13.2 (-9.0)Staar Surgical Company · Lens, Intraocular, Phakic1 recall9 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 12.6 (-9.0)Staar Surgical Company · Lens, Intraocular, Phakic1 recall9 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 12.6 (-9.5)Staar Surgical Company · Lens, Intraocular, Phakic1 recall8 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page