Skip to content

EVO/EVO+VISIAN Implantable Collamer Lens VICM5 13.2 (-12.5) (REF VICM5 13.2)Recalls & reported problems

Staar Surgical Company · VICM5 - Diopter: -12.5; Overall Diameter (Length): 13.2 · Class 3 · Implant

All models of EVO/EVO+VISIAN Implantable Collamer Lens (1030)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00840311300983), received between Jul 25, 2023 and Jun 9, 2026 — classified as 6 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Inadequacy of Device Shape and/or Size” (1 report).

8 other recalls cover devices cleared under the same FDA submission (P030016); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 2 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem4
  • No Apparent Adverse Event3
  • Inadequacy of Device Shape and/or Size1
  • Misfocusing1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available3
  • Blurred Vision3
  • No Clinical Signs, Symptoms or Conditions3
  • Corneal Edema1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 9, 2026Adverse Event Without Identified Device or Use Problem

    A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED OVER/UNDER CORRECTION. LENS EXCHANGED WITH DIFFERENT DIOPTER LENS. PROBLEM RESOLVED. THERE ARE MULTIPLE MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS PATIENT. THIS REPORT IS 1 OF 2 TOTAL REPORTS. BASED ON THE INFORMATION PROVIDED, THE RISK ASSESSMENT FOR OUR PRODUCT, AND OUR EXPERIENCE, WE HAVE DETERMINED THE CAUSE OR CONTRIBUTION TO THE REPORTED ISSUE IS NOT INDICATIVE OF A MANUFACTURING RELATED ISSUE OR PRODUCT DEFICIENCY. CAUSE OF THE EVENT WAS REPORTED AS UNKNOWN.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 25439025 ↗
  • InjuryMar 19, 2026Adverse Event Without Identified Device or Use Problem

    A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED EXCESSIVE VAULT, SIGNIFICANT REDUCTION OF IRIDOCORNEAL ANGLES AND SIGNIFICANT SHALLOWING OF AC. IN A SEPARATE SURGERY THE LENS WAS EXCHANGED WITH A SHORTER LENGTH AND THE PROBLEM WAS RESOLVED. THE CAUSE OF THE EVENT WAS REPORTED AS A PATIENT RELATED FACTOR.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 24640871 ↗
  • InjuryDec 4, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A HEALTHCARE PROFESSIONAL REPORTED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED OVER/UNDER CORRECTION. OTHER SURGICAL INTERVENTION WAS PERFORMED: LRIS. CAUSE OF THE EVENT IS UNKNOWN.

    Patient problems: Blurred Vision

    MAUDE report 23722999 ↗
  • MalfunctionNov 8, 2024No Apparent Adverse Event; No Apparent Adverse Event

    A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED EXCESSIVE VAULT, PIGMENT DISPERSION AND SIGNIFICANT REDUCTION OF IRIDO-CORNEAL ANGLES. DUE TO EXCESSIVE VAULT, THE LENS WAS REMOVED AND REPLACED INTRAOPERATIVELY FOR A SHORTER LENGTH LENS. THE PROBLEM WAS RESOLVED. CAUSE OF EVENT WAS REPORTED AS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20644810 ↗
  • InjuryMay 22, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE REPORTER INDICATED THAT THE SURGEON IMPLANTED A 13.2MM VICM5_13.2 -12.50D IMPLANTABLE COLLAMER LENS INTO THE PATIENT'S LEFT EYE (OS) ON (B)(6) 2024. PRODUCT NONCONFORMITY OBSERVED; LENS POWER DIFFERENCE SUSPECTED. LENS WAS EXCHANGED ON (B)(6) 2024 BUT PROBLEM WAS NOT RESOLVED. CAUSE OF EVENT WAS REPORTED AS DEVICE; DEVICE DID NOT FAIL TO PERFORM AS INTENDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19372622 ↗
  • InjuryJan 8, 2024Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    THE REPORTER INDICATED THAT A 13.7MM VICM5_13.7 IMPLANTABLE COLLAMER LENS OF -12.50 DIOPTER WAS IMPLANTED INTO THE PATIENT'S LEFT EYE (OS) ON (B)(6) 2023. ON (B)(6) 2023 THE LENS WAS EXCHANGED WITH A SHORTER LENGTH LENS DUE TO EXCESSIVE VAULT AND SIGNIFICANT REDUCTION OF IRIDOCORNEAL ANGLES. THE EXCHANGE RESOLVED THE PROBLEM, THE REPORTER STATED " VAULT REDUCED TO NORMAL AND PATIENT IS DOING GOOD". CAUSE OF EVENT WAS UNKNOWN.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 18472104 ↗
  • MalfunctionDec 29, 2023No Apparent Adverse Event; No Apparent Adverse Event

    THE REPORTER INDICATED THAT ON A 13.2MM VICM513.2 IMPLANTABLE COLLAMER LENS OF A -12.5 DIOPTER TORE/BROKE DURING LOADING AFTER OPENING THE VIAL. A REPLACEMENT LENS WAS IMPLANTED INTO THE PATIENT'S RIGHT EYE (OD) ON 19-JUL-2023 AND THE PROBLEM WAS NOT RESOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18421055 ↗
  • InjuryDec 29, 2023Misfocusing; Misfocusing

    THE REPORTER INDICATED THAT ON A 13.2MM VICM513.2 IMPLANTABLE COLLAMER LENS OF A -12.5 WAS IMPLANTED AS A REPLACEMENT INTO THE PATIENT'S RIGHT EYE (OD) ON (B)(6) 2023. ON (B)(6) 2023, THE LENS WAS EXCHANGED FOR A TORIC LENS DUE TO REFRACTIVE SURPRISE. THE PROBLEM WAS RESOLVED. THE CAUSE OF THE EVENT IS UNKNOWN.

    Patient problems: Blurred Vision

    MAUDE report 18421048 ↗
  • MalfunctionJul 25, 2023No Apparent Adverse Event; No Apparent Adverse Event

    THE REPORTER INDICATED THAT A 13.2MM VICM5_13.2 IMPLANTABLE COLLAMER LENS OF -12.50 DIOPTER TORE DURING INJECTION, INTRAOPERATIVELY THE LENS WAS REMOVED WITH PATIENT INJURY OF POSTOP CORNEAL EDEMA. SURGEON ALSO NOTED PATIENT HAD BLURRED VISION. A REPLACEMENT LENS OF THE SAME LENGTH WAS IMPLANTED ON THIS SAME DATE. CAUSE OF THE EVENT WAS REPORTED AS THE DEVICE AND THERE IS DEVICE CAUSALITY.

    Patient problems: Blurred Vision; Corneal Edema

    MAUDE report 17393103 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P030016. One submission often covers many models, so check the product and lot details in each record.

  • Z-1480-2026 · Jan 21, 2026 · Open, Classified
    EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5
  • Z-0464-2024 · Sep 29, 2023 · Open, Classified
    EVO+VISIAN Implantable Collamer Lens, REF:VICM5_12.1. Phakic Intraocular lens.
  • Z-0466-2024 · Sep 29, 2023 · Open, Classified
    EVO+VISIAN Implantable Collamer Lens, REF: VICM5_13.2. Phakic Intraocular lens.
  • Z-0469-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens REF: VICMO13.2. Phakic Intraocular lens.
  • Z-0467-2024 · Sep 29, 2023 · Open, Classified
    EVO +VISIAN Implantable Collamer Lens, REF: VICM5_13.7.Phakic Intraocular lens.
  • Z-0465-2024 · Sep 29, 2023 · Open, Classified
    EVO+VISIAN Implantable Collamer Lens, REF: VICM5_12.6. Phakic Intraocular lens.
  • Z-0468-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens, REF: VICMO12.1. Phakic Intraocular lens.
  • Z-0474-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens, REF: VICMO12.6. Phakic Intraocular lens.