Skip to content

MERCURY 62075-050Recalls & reported problems

Spinal Elements, Inc. · Screw;Ø7.5 X 50MM MERCURY® CLASSIC POLYAXIAL SCREW · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00840606117791), received between Oct 22, 2019 and Oct 22, 2019 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • Failure of Implant1
  • Pain1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 22, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    IN LATE (B)(6) 2019, PATIENT (DIAGNOSED WITH PSEUDARTHROSIS) UNDERWENT REVISION SURGERY TO REMOVE (COMPETITIVE IMPLANT HARDWARE) LEFT FRACTURED BONE SCREW AND REPLACED WITH RIGHT SIDE, UNILATERAL HARDWARE FROM L4-S1. DURING ROUTINE TWO WEEK POST-OP FOLLOW UP, RADIOGRAPHIC IMAGING DEPICTED PEDICLE BONE SCREW HEAD SEPARATION FROM RIGHT SHANK AT S1. APPROXIMATELY TWO WEEKS LATER, REVISION SURGERY WAS PERFORMED TO REMOVE REPLACE THE RIGHT S1 SCREW AND ADD SUPPLEMENTAL FIXATION (ADJACENT ROD AND ILIAC SCREW). POST-REVISION PATIENT IS REPORTEDLY DOING WELL.

    Patient problems: Failure of Implant; Pain

    MAUDE report 9218773 ↗

Other Orthosis, spinal pedicle fixation, for degenerative disc disease devices

All Orthosis, spinal pedicle fixation, for degenerative disc disease devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page