ICU Medical 42584-05Recalls & reported problems

Icu Medical, Inc. · Transpac™ IV Monitoring Kit, 60", Disposable Transducer, 3 ml Squeeze Flush Device, Macrodrip · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00840619027414), received between Jun 26, 2024 and Aug 13, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Crack1
  • Failure to Disconnect1
  • Imprecision1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 13, 2024Imprecision; Imprecision

    AN EVENT INVOLVING A TRANSDUCER, 60 INCH (152 CM), 3 ML/HR, MICRODRIP EXPERIENCED INACCURATE READINGS. THE FOLLOWING ISSUE WAS REPORTED BY THE CUSTOMER: ¿ AN ARTERIAL CATHETER WAS PLACED USING THE KIT FOR MEASURING THE PRESSURE. IT WAS OBSERVED THAT IT INDICATED 45 SYSTOLE ( WHICH DOES NOT MATCH THE PATIENT'S CLINICAL CONDITION ). IT WAS AFTERWARDS OBSERVED THAT IT COULD NOT BE POSSIBLE TO ZERO THE CALIBRATION. THE CATHETER WORKED NORMALLY DURING SAMPLING. ACTIONS TAKEN: A BLOOD PRESSURE MEASUREMENT WITH CUFF SHOWED A DIFFERENCE OF 4 POINTS IN THE PRESSURE. THE TEAM PERFORMED A MODULE AND CABLE CHANGE, WHICH WAS INEFFECTIVE. FINALLY, THE TEAM CHANGED COMPLETELY THE MEASUREMENT KIT, AFTER THE CHANGE THE KIT WAS FUNCTIONAL. THERE WERE NO SERIOUS CONSEQUENCES FOR THE PATIENT, HOWEVER, THE PATIENT APPEARED HYPOTENSIVE WHEN SHE WAS ACTUALLY NORMOTENSIVE, RESULTING IN A LOSS OF TIME. ".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19977735 ↗
  • MalfunctionJul 3, 2024Break; Failure to Disconnect; Break; Failure to Disconnect

    THE EVENT OCCURRED ON AN UNSPECIFIED DATE INVOLVING A TRANSPAC® IV MONITORING KIT, 60", DISPOSABLE TRANSDUCER, 3 ML SQUEEZE FLUSH DEVICE, MACRODRIP WHERE THE CUSTOMER REPORTED THAT THE CURRENT ARTERIAL LINE IS DEFECTIVE. IT IS DIFFICULT TO DISCONNECT AT PATIENT'S SIDE AND FLUSHING DEVICE BREAKS EASILY. THERE WAS PATIENT INVOLVEMENT AND DELAY IN CRITICAL THERAPY. THE CAREGIVER CHANGED TO ANOTHER ARTERIAL LINE TUBING TO ADDRESS THE ISSUE. THERE WAS NO SERIOUS DETERIORATION IN THE STATE OF HEALTH OF A PATIENT, USER, OR OTHER PERSON. THERE WAS NO DEATH OF A PATIENT, USER, OR OTHER PERSON. THERE WAS NO POTENTIAL FOR DEATH OR SERIOUS DETERIORATION IN HEALTH OF A PATIENT, USER, OR OTHER PERSON.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19672170 ↗
  • MalfunctionJun 26, 2024Crack; Crack

    THE EVENT OCCURRED ON AN UNSPECIFIED DATE AND INVOLVED A TRANSPAC® IV MONITORING KIT, 60", DISPOSABLE TRANSDUCER, 3 ML SQUEEZE FLUSH DEVICE, MACRODRIP. THE CUSTOMER STATED THAT THE CURRENT ARTERIAL LINE IS DEFECTIVE. THE FLUSHING DEVICE BREAKS EASILY. THERE WAS UNKNOWN PATIENT INVOLVEMENT AND UNKNOWN PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19620069 ↗

Other Transducer, blood-pressure, extravascular devices

All Transducer, blood-pressure, extravascular devices →