ICU Medical 42646-06Recalls & reported problems

Icu Medical, Inc. · Transpac™ IV MONITORING KIT WITH Safeset™ RESERVOIR AND 2 BLOOD SAMPLING PORTS, 84" TUBING, DISPOSABLE TRANSDUCER, 3 ML SQUEEZE FLUSH, MACRODRIP (POLE MOUNT) · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
1
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00840619028053), received between Oct 26, 2016 and Aug 21, 2024 — classified as 1 death, 1 injury, 6 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Material Separation3
  • Detachment Of Device Component1
  • Device Inoperable1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Consequences Or Impact To Patient2
  • Blood Loss1
  • Hemorrhage/Blood Loss/Bleeding1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 21, 2024Break; Break

    AN EVENT INVOLVING A TRANSPAC® IV MONITORING KIT WITH SAFESET¿ RESERVOIR AND 2 BLOOD SAMPLING PORTS, 84" TUBING, DISPOSABLE TRANSDUCER, 3 ML SQUEEZE FLUSH, MACRODRIP (POLE MOUNT) EXPERIENCED BREAKAGE. IT WAS REPORTED THAT THE CUSTOMER HAD A FAULTY TUBING. THE EVENT WAS NOTED ON 22-JUL-2024. THERE WAS UNKNOWN PATIENT INVOLVEMENT, AND UNKNOWN HUMAN HARM. IT WAS REPORTED THAT THE PRODUCT HAD A BROKEN ART LINE KIT. UPDATE: IT WAS NOTED THAT THERE WERE NO VISIBLE DEFECTS SEEN PRIOR TO THE BRAKEAGE. THERE WAS PATIENT INVOLVEMENT, NO HUMAN HARM, AND THERE WAS DELAY IN THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20017343 ↗
  • InjuryJun 28, 2024Break; Break

    THE EVENT INVOLVED A TRANSPAC IV MONITORING KIT WITH SAFESET¿ RESERVOIR AND 2 BLOOD SAMPLING PORTS, 84" TUBING, DISPOSABLE TRANSDUCER, 3 ML SQUEEZE FLUSH, MACRODRIP (POLE MOUNT) WHERE THE CUSTOMER REPORTED THAT THE DEVICE HAD A "BROKEN ART LINE KIT". THE PATIENT WAS FOUND BLEEDING PROFUSELY ON HIS ARM. NO INTERVENTION WAS REPORTED. NO OTHER INFORMATION AVAILABLE AT THIS TIME.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions

    MAUDE report 19635429 ↗
  • MalfunctionJul 13, 2023Material Separation; Material Separation

    THE EVENT INVOLVED A TRANSPAC® IV MONITORING KIT WITH SAFESET¿ RESERVOIR AND 2 BLOOD SAMPLING PORTS, 84" TUBING, DISPOSABLE TRANSDUCER, 3 ML SQUEEZE FLUSH, MACRODRIP (POLE MOUNT).IT WAS REPORTED THAT THE TUBING HAS BEEN DISCONNECTED FROM THE SOURCE AND IT WAS DETECTED BY THE NURSE RIGHT AWAY. THE TUBING WAS HOOKED UP TO A CENTRAL LINE FOR CENTRAL VENUS PRESSURE MONITORING AND CAME APART. THE TUBING WAS PROPERLY SECURED WITH A LINE HOLDER ABOUT 10 MINUETS PRIOR, THE PRESSURE TUBING WAS USED FOR A BLOOD DRAW FROM CENTRAL VENUS LINE WITHOUT ANY ISSUE. LAB CAME BACK AND NEEDED ADDITIONAL BLOOD AND THE PATIENT WAS NOT REPOSITIONED AND THE TUBING WAS NOT TOUCHED SINCE THE PAST DRAW AND NEEDED ADDITIONAL BLOOD AND THE PATIENT WAS ALSO NOT REPOSISTIONED. WHEN THE VAMP WAS LIFTED OFF THE BED, A PORTION OF THE TUBING CAME APART WITH VERY LITTLE FORCE. A SMALL PORTION OF THE PRESSURE TUBING WAS LEFT ATTACHED TO THE CVL AND THE TUBING WAS IMMEDIATLEY CLAMPED PROXIMAL TO THE PATIENT. THE TUBING WAS UNHOOKED FROM THE CVL. THE TUBING WAS NOT RIPPED AND THE END DUD NOT HAVE A LUER LOCK OR ANY ATTACHMENT WHERE IT COULD BE INTENTIONALLY DISCONNECTED. THE TUBING SEEMED ABNORMALLY WEAK THERE WAS PATIE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17318330 ↗
  • MalfunctionApr 1, 2022Material Separation; Material Separation

    THE EVENT INVOLVED A TRANSPAC® IV MONITORING KIT WITH SAFESET¿ RESERVOIR AND 2 BLOOD SAMPLING PORTS, 84" TUBING, DISPOSABLE TRANSDUCER, 3 ML SQUEEZE FLUSH, MACRODRIP (POLE MOUNT). IT WAS STATED THAT THE A-LINE SEPARATED IN HALF DURING A PROCEDURE. THERE WAS PATIENT INVOLVEMENT, HOWEVER NO HUMAN HARM AND NO DELAY IN THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13980787 ↗
  • MalfunctionMar 29, 2022Material Separation; Material Separation

    THE EVENT INVOLVED A TRANSPAC® IV MONITORING KIT WITH SAFESET¿ RESERVOIR AND 2 BLOOD SAMPLING PORTS, 84" TUBING, DISPOSABLE TRANSDUCER, 3 ML SQUEEZE FLUSH, MACRODRIP (POLE MOUNT) IN WHICH THE CUSTOMER REPORTED THAT THE ARTERIAL SYRINGE BECAME SEPARATED FROM THE RUBBER. THE EVENT HAPPENED DURING PROCEDURE ON A PATIENT. THERE WAS PATIENT INVOLVEMENT, HOWEVER, NO HARM WAS REPORTED AS A CONSEQUENCE OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13941160 ↗
  • DeathDec 11, 2020Break; Break

    THE EVENT INVOLVED A TRANSPAC IV MONITORING KIT THAT THE CUSTOMER REPORTED THE TRANSPAC BROKE AT THE CONNECTION OF THE SAFESET AND THE TUBING NEAR THE WHITE ONE WAY STOPCOCK DURING PATIENT USE. SOMEONE BECAME AWARE OF THE ISSUE SINCE THE PATIENT BLED OUT FROM A MAJOR ARTERY. THE CUSTOMER WAS UNABLE TO QUANTIFY THE BLOOD LOSS. THE PATIENT WAS A COVID POSITIVE PATIENT. THE CUSTOMER STATED THE BREAK HAPPENED DURING A CODE. PRESSURE WAS HELD AT THE SITE AND IMMEDIATELY CHANGED OUT TO ANOTHER SET. THE CUSTOMER STATED THERE WERE NO OTHER MEDICAL INTERVENTIONS REQUIRED RELATED TO THE TUBING BREAK. THE CUSTOMER REPORTED THE PATIENT DECEASED ON A UNSPECIFIED DATE. THE CUSTOMER DECLINED TO PROVIDE ADDITIONAL INFORMATION UPON FOLLOW-UP.

    Patient problems: Blood Loss

    MAUDE report 10990734 ↗
  • MalfunctionAug 9, 2017Leak/Splash; Leak/Splash

    COMPLAINT RECEIVED REGARDING ONE 42646-06, TRANSPAC IV MONITORING KIT. REPORT STATES: PATIENT HAD L. FEMORAL ARTERIAL LINE THAT BROKE OFF AT 3-WAY STOPCOCK, WHERE THE TUBING FITS INTO THE LUER. PRESSURE BAG BECAME INEFFECTIVE. CLAMPED OFF THE LINE AND REPLACED IT WITH A NEW ARTERIAL LINE WITH NO FURTHER ISSUES. NO ADVERSE PATIENT CONSEQUENCES REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6778249 ↗
  • MalfunctionOct 26, 2016Detachment Of Device Component; Device Inoperable

    ARTERIAL LINE SAFESET SYRINGE PLUNGER CAME APART AND WAS NON-FUNCTIONAL. BLACK ROBBER STOPPER WAS DETACHED FROM SYRINGE. BLOOD WAS BACKED UP IN THE SYRINGE AND UNSURE IF AIR WAS PRESENT. SAFESET CHANGED. NO HARM TO PT. ALSO NOTED IN OTHER SAFESETS THAT THE RUBBER SLEEVE INSIDE THE SYRINGE DETACHED FROM ITS ANCHOR POINT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6063184 ↗

Other Transducer, blood-pressure, extravascular devices

All Transducer, blood-pressure, extravascular devices →