ICU Medical CH2000S-10Recalls & reported problems
Icu Medical, Inc. · Spinning Spiros™ Closed Male Luer, 10 units · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00840619033910), received between Sep 23, 2020 and Jul 28, 2026 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (4 reports).
5 other recalls cover devices cleared under the same FDA submission (K082806); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak4
- Disconnection2
- Separation Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Chemical Exposure3
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 28, 2026Disconnection
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50059740 ↗MalfunctionApr 21, 2026Separation Problem
EVENT OCCURRED REGARDING A SPINNING SPIROS CLOSED MALE LUER, 10 UNITS WHERE IT WAS REPORTED THAT DURING A 5-FU PUMP INFUSION, THE SPIROS DEVICE CAME UNSCREWED FROM THE PUMP TUBING. IT DID NOT UNSCREW FROM THE BLUE CLAVE BUT FROM THE PUMP TUBING ITSELF. THEY REPORTED THAT ONCE SCREWED ON, IT SHOULD JUST SPIN AND NOT BE ABLE TO UNSCREW. THIS EVENT OCCURRED IN THE TOPSHAM OFFICE. THERE WAS PATIENT INVOLVEMENT BUT NO REPORTED HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24941717 ↗MalfunctionJul 8, 2021Fluid/Blood Leak; Fluid/Blood Leak
THE CUSTOMER REPORTED THAT THE SPIROS LEAKED LEAKED WHILE PUSHING DOXORUBICIN. THE LEAK OCCURRED ON THE LUER CONNECTION BETWEEN SPIROS AND SYRINGE. THERE WAS PATIENT INVOLVEMENT BUT NO REPORTED HARM, THE ONLY PATIENT EXPOSURE WAS 2 DROPS ON PATIENT'S SWEATER. THE CHEMOTHERAPY SPILL WAS CLEANED UP PER PROTOCOL; HOWEVER, A SPILL KIT WAS NOT USED. THERE WAS NO ADVERSE EVENT OR DELAY IN THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12134350 ↗MalfunctionMay 25, 2021Fluid/Blood Leak; Fluid/Blood Leak
THE EVENT INVOLVED A SPINNING SPIROS® CLOSED MALE LUER WHICH DISCONNECTED FROM THE LUER OF THE SYRINGE WHILE THE NURSE WAS BEGINNING TO ADMINISTER THE CHEMOTHERAPY (DOXORUBICIN) TO THE PATIENT. THE DRUG DID DRIP ONTO PATIENT AND PATIENT'S SHIRT; THE SHIRT WAS REMOVED, AND THE AREA WAS WASHED WITH SOAP AND WATER. THERE WAS PATIENT INVOLVEMENT, HOWEVER, THERE WAS NO HARM REPORTED AS A CONSEQUENCE OF THIS EVENT.
Patient problems: Chemical Exposure; No Clinical Signs, Symptoms or Conditions
MAUDE report 11881463 ↗MalfunctionOct 2, 2020Fluid/Blood Leak; Fluid/Blood Leak
THE EVENT INVOLVED A SPINNING SPIROS® CLOSED MALE LUER, 10 UNITS THAT LEAKED BETWEEN THE SYRINGE AND THE SPIROS LUER LOCK WHEN THE NURSE WAS GETTING READY TO ADMINISTER A DOXORUBICIN INFUSION. IT WAS REPORTED THAT DILUTED CHEMO TOUCHED THE SKIN OF THE PATIENT¿S FOREARM WHEN THE SPIROS CAME DISCONNECTED FROM THE SYRINGE. THE AREA WAS WASHED IMMEDIATELY (SKIN REMAINED INTACT) AND THE CHEMO SPILL WAS CLEANED UP PER FACILITY PROTOCOL. THE TUBING WAS SET UP WITH INTRAVENOUS PUSH (IVP) ADMINISTERED AT LOWEST HUB OF THE UNKNOWN TUBING SET THAT WAS INFUSING NORMAL SALINE TO GRAVITY. THE SET UP AND DRUG WAS REPLACED AND ADMINISTERED SUCCESSFULLY WITHOUT FURTHER INCIDENT. THERE WERE NO HOLES, CUTS, TEARS OR ANY DEFECTS NOTED. THERE WAS PATIENT INVOLVEMENT AND NO ADVERSE EVENT REPORTED.
Patient problems: Chemical Exposure; No Consequences Or Impact To Patient
MAUDE report 10623275 ↗MalfunctionSep 23, 2020Disconnection; Fluid/Blood Leak; Disconnection; Fluid/Blood Leak
THE EVENT INVOLVED A SPINNING SPIROS® CLOSED MALE LUER, 10 UNITS THAT THE CUSTOMER REPORTED A LEAK OF DOXORUBICIN. A BOLUS DOSE OF DOXORUBICIN WAS BEING ADMINISTERED AT THE LOWEST HUB OF THE NORMAL SALINE TUBING THAT WAS RUNNING TO GRAVITY. THE SPIROS CAME DISCONNECTED FROM THE SYRINGE, DILUTED CHEMOTHERAPY TOUCHED THE PATIENT'S FOREARM, AND THE AREA WAS WASHED IMMEDIATELY. THERE WAS PATIENT INVOLVEMENT, HOWEVER, NO DELAY IN DELIVERY DUE TO THE EVENT AND NO REPORT OF ADVERSE EVENT.
Patient problems: Chemical Exposure; No Consequences Or Impact To Patient
MAUDE report 10572526 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K082806. One submission often covers many models, so check the product and lot details in each record.
- Z-1723-2022 · Aug 1, 2022 · Open, Classified6.5 IN (17 cm) APPX 0.78 ml, EXT SET w/Clave CLEAR, Clave 4-WAY Stopcock, Spiros, RED CAP. Accessory for intravascular administration set.
- Z-1567-2021 · Mar 16, 2021 · Open, ClassifiedREF 20124-01 30" (76 cm) Appx 3.2 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, Spiros, Bag Hanger LOT 5142971 - Product Usage: provides access for the administration of
- Z-1568-2021 · Mar 16, 2021 · Open, ClassifiedREF 20130-01 Spinning Spiros Closed Male Luer, Red Cap - Product Usage: provides access for the administration of fluids from a container to a patients vascular system through the
- Z-1569-2021 · Mar 16, 2021 · Open, ClassifiedREF Spinning Spiros Closed Male Luer - Product Usage: provides access for the administration of fluids from a container to a patients vascular system through the administration se
- Z-0702-2021 · Aug 31, 2020 · Open, Classified30" (76 cm) Appx 3.2 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, Spiros, Bag Hanger, REF: 20124-01; Spinning Spiros Closed Male Luer, Red Cap, REF: 20130-01; Spinning Sp
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page