ICU Medical CH3340Recalls & reported problems
Icu Medical, Inc. · Bag Spike w/Clave™ Additive Port, Clave™ · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00840619056216), received between Apr 25, 2025 and Apr 25, 2025 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (6 reports).
1 other recall cover devices cleared under the same FDA submission (K080989); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak6
- No Flow6
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 25, 2025Fluid/Blood Leak; No Flow; Fluid/Blood Leak; No Flow
THE COMPLAINT/EVENT OCCURRED ON AN UNSPECIFIED DATE AND INVOLVED A BAG SPIKE W/CLAVE® ADDITIVE PORT, CLAVE®. IT WAS REPORTED THAT FLUID WAS PREVENTED FROM PASSING THROUGH WHEN ATTACHING HAZARDOUS A INFUSIONS TO THE IV SET WHEN CONNECTED. THERE WAS ALSO POSSIBLE LEAKAGE UPON DISCONNECTION. THE SPIROS PORTS WERE STUCK OPEN ON THE CH3340 (EITHER SPIROS PORT) AFTER DISCONNECTING. THE INTERNAL STOPPER APPEARS TO BE TURNED/MISALIGNED AND IT DOESN¿T SPRING BACK TO CLOSE THE PORT. THEIR PROCESS FOR INFUSIONS IS SENDING THE IV BAG WITH THE CH3340 ATTACHED FROM PHARMACY. NURSING THEN ADDS A CH3034 BAG SPIKE ADAPTER ONTO THE SPIKE OF A REGULAR IV SET AND ADDS A SPINNING SPIROS CLOSED MALE LUER ADAPTER CH2000S TO THE LUER SIDE OF THE IV SET. THEY CONNECT THE CH3034 SPIROS TO THE IN-LINE BAG SPIKE AND THEN ATTACH THE SPINNING SPIROS TO THE ADDITIVE PORT ON THE BAG TO CIRCLE PRIME. THEY HAVE HAD MULTIPLES ISSUES WITH LEAKS FROM EITHER OF THE SPIROS PORTS ON THE CH3340 THAT WERE NOT REACTIVATING WHEN DISCONNECTING CAUSING UNSPECIFIED CHEMOTHERAPY TO SPILL. IN MOST OF THESE SITUATIONS, THE FLUID DOESN¿T PASS THROUGH TO THE LINE, SO THE NURSE GOES TO CHECK OR DETACH FROM THE BAG SPIKE SPILLING AFTE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21912871 ↗MalfunctionApr 25, 2025Fluid/Blood Leak; No Flow; Fluid/Blood Leak; No Flow
THE COMPLAINT/EVENT INVOLVED A BAG SPIKE W/CLAVE® ADDITIVE PORT, CLAVE®. IT WAS REPORTED THAT FLUID WAS PREVENTED FROM PASSING THROUGH WHEN ATTACHING HAZARDOUS A INFUSIONS TO THE IV SET WHEN CONNECTED. THERE WAS ALSO POSSIBLE LEAKAGE UPON DISCONNECTION. THE SPIROS PORTS WERE STUCK OPEN ON THE CH3340 (EITHER SPIROS PORT) AFTER DISCONNECTING. THE INTERNAL STOPPER APPEARS TO BE TURNED/MISALIGNED AND IT DOESN¿T SPRING BACK TO CLOSE THE PORT. THEIR PROCESS FOR INFUSIONS IS SENDING THE IV BAG WITH THE CH3340 ATTACHED FROM PHARMACY. NURSING THEN ADDS A CH3034 BAG SPIKE ADAPTER ONTO THE SPIKE OF A REGULAR IV SET AND ADDS A SPINNING SPIROS CLOSED MALE LUER ADAPTER CH2000S TO THE LUER SIDE OF THE IV SET. THEY CONNECT THE CH3034 SPIROS TO THE IN-LINE BAG SPIKE AND THEN ATTACH THE SPINNING SPIROS TO THE ADDITIVE PORT ON THE BAG TO CIRCLE PRIME. THEY HAVE HAD MULTIPLES ISSUES WITH LEAKS FROM EITHER OF THE SPIROS PORTS ON THE CH3340 THAT WERE NOT REACTIVATING WHEN DISCONNECTING CAUSING UNSPECIFIED CHEMOTHERAPY TO SPILL. IN MOST OF THESE SITUATIONS, THE FLUID DOESN¿T PASS THROUGH TO THE LINE, SO THE NURSE GOES TO CHECK OR DETACH FROM THE BAG SPIKE SPILLING AFTER DISCONNECTION, AND ADDITIONALLY TH
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21912673 ↗MalfunctionApr 25, 2025Fluid/Blood Leak; No Flow; Fluid/Blood Leak; No Flow
THE COMPLAINT/EVENT INVOLVED A BAG SPIKE W/CLAVE® ADDITIVE PORT, CLAVE®. IT WAS REPORTED THAT FLUID WAS PREVENTED FROM PASSING THROUGH WHEN ATTACHING HAZARDOUS A INFUSIONS TO THE IV SET WHEN CONNECTED. THERE WAS ALSO POSSIBLE LEAKAGE UPON DISCONNECTION. THE SPIROS PORTS WERE STUCK OPEN ON THE CH3340 (EITHER SPIROS PORT) AFTER DISCONNECTING. THE INTERNAL STOPPER APPEARS TO BE TURNED/MISALIGNED AND IT DOESN T SPRING BACK TO CLOSE THE PORT. THEIR PROCESS FOR INFUSIONS IS SENDING THE IV BAG WITH THE CH3340 ATTACHED FROM PHARMACY. NURSING THEN ADDS A CH3034 BAG SPIKE ADAPTER ONTO THE SPIKE OF A REGULAR IV SET AND ADDS A SPINNING SPIROS CLOSED MALE LUER ADAPTER CH2000S TO THE LUER SIDE OF THE IV SET. THEY CONNECT THE CH3034 SPIROS TO THE IN-LINE BAG SPIKE AND THEN ATTACH THE SPINNING SPIROS TO THE ADDITIVE PORT ON THE BAG TO CIRCLE PRIME. THEY HAVE HAD MULTIPLES ISSUES WITH LEAKS FROM EITHER OF THE SPIROS PORTS ON THE CH3340 THAT WERE NOT REACTIVATING WHEN DISCONNECTING CAUSING UNSPECIFIED CHEMOTHERAPY TO SPILL. IN MOST OF THESE SITUATIONS, THE FLUID DOESN¿T PASS THROUGH TO THE LINE, SO THE NURSE GOES TO CHECK OR DETACH FROM THE BAG SPIKE SPILLING AFTER DISCONNECTION, AND ADDITIONALLY TH
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21912376 ↗MalfunctionApr 25, 2025Fluid/Blood Leak; No Flow; Fluid/Blood Leak; No Flow
THE COMPLAINT/EVENT INVOLVED A BAG SPIKE W/CLAVE® ADDITIVE PORT, CLAVE®. IT WAS REPORTED THAT FLUID WAS PREVENTED FROM PASSING THROUGH WHEN ATTACHING HAZARDOUS A INFUSIONS TO THE IV SET WHEN CONNECTED. THERE WAS ALSO POSSIBLE LEAKAGE UPON DISCONNECTION. THE SPIROS PORTS WERE STUCK OPEN ON THE CH3340 (EITHER SPIROS PORT) AFTER DISCONNECTING. THE INTERNAL STOPPER APPEARS TO BE TURNED/MISALIGNED AND IT DOESN'T SPRING BACK TO CLOSE THE PORT. THEIR PROCESS FOR INFUSIONS IS SENDING THE IV BAG WITH THE CH3340 ATTACHED FROM PHARMACY. NURSING THEN ADDS A CH3034 BAG SPIKE ADAPTER ONTO THE SPIKE OF A REGULAR IV SET AND ADDS A SPINNING SPIROS CLOSED MALE LUER ADAPTER CH2000S TO THE LUER SIDE OF THE IV SET. THEY CONNECT THE CH3034 SPIROS TO THE IN-LINE BAG SPIKE AND THEN ATTACH THE SPINNING SPIROS TO THE ADDITIVE PORT ON THE BAG TO CIRCLE PRIME. THEY HAVE HAD MULTIPLES ISSUES WITH LEAKS FROM EITHER OF THE SPIROS PORTS ON THE CH3340 THAT WERE NOT REACTIVATING WHEN DISCONNECTING CAUSING UNSPECIFIED CHEMOTHERAPY TO SPILL. IN MOST OF THESE SITUATIONS, THE FLUID DOESN'T PASS THROUGH TO THE LINE, SO THE NURSE GOES TO CHECK OR DETACH FROM THE BAG SPIKE SPILLING AFTER DISCONNECTION, AND ADDITIONALLY TH
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21912170 ↗MalfunctionApr 25, 2025Fluid/Blood Leak; No Flow; Fluid/Blood Leak; No Flow
THE COMPLAINT/EVENT INVOLVED A BAG SPIKE W/CLAVE® ADDITIVE PORT, CLAVE®. IT WAS REPORTED THAT FLUID WAS PREVENTED FROM PASSING THROUGH WHEN ATTACHING HAZARDOUS A INFUSIONS TO THE IV SET WHEN CONNECTED. THERE WAS ALSO POSSIBLE LEAKAGE UPON DISCONNECTION. THE SPIROS PORTS WERE STUCK OPEN ON THE CH3340 (EITHER SPIROS PORT) AFTER DISCONNECTING. THE INTERNAL STOPPER APPEARS TO BE TURNED/MISALIGNED AND IT DOESN¿T SPRING BACK TO CLOSE THE PORT. THEIR PROCESS FOR INFUSIONS IS SENDING THE IV BAG WITH THE CH3340 ATTACHED FROM PHARMACY. NURSING THEN ADDS A CH3034 BAG SPIKE ADAPTER ONTO THE SPIKE OF A REGULAR IV SET AND ADDS A SPINNING SPIROS CLOSED MALE LUER ADAPTER CH2000S TO THE LUER SIDE OF THE IV SET. THEY CONNECT THE CH3034 SPIROS TO THE IN-LINE BAG SPIKE AND THEN ATTACH THE SPINNING SPIROS TO THE ADDITIVE PORT ON THE BAG TO CIRCLE PRIME. THEY HAVE HAD MULTIPLES ISSUES WITH LEAKS FROM EITHER OF THE SPIROS PORTS ON THE CH3340 THAT WERE NOT REACTIVATING WHEN DISCONNECTING CAUSING UNSPECIFIED CHEMOTHERAPY TO SPILL. IN MOST OF THESE SITUATIONS, THE FLUID DOESN¿T PASS THROUGH TO THE LINE, SO THE NURSE GOES TO CHECK OR DETACH FROM THE BAG SPIKE SPILLING AFTER DISCONNECTION, AND ADDITIONALLY TH
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21912014 ↗MalfunctionApr 25, 2025Fluid/Blood Leak; No Flow; Fluid/Blood Leak; No Flow
THE COMPLAINT/EVENT INVOLVED A BAG SPIKE W/CLAVE® ADDITIVE PORT, CLAVE®. IT WAS REPORTED THAT FLUID WAS PREVENTED FROM PASSING THROUGH WHEN ATTACHING HAZARDOUS A INFUSIONS TO THE IV SET WHEN CONNECTED. THERE WAS ALSO POSSIBLE LEAKAGE UPON DISCONNECTION. THE SPIROS PORTS WERE STUCK OPEN ON THE CH3340 (EITHER SPIROS PORT) AFTER DISCONNECTING. THE INTERNAL STOPPER APPEARS TO BE TURNED/MISALIGNED AND IT DOESN¿T SPRING BACK TO CLOSE THE PORT. THEIR PROCESS FOR INFUSIONS IS SENDING THE IV BAG WITH THE CH3340 ATTACHED FROM PHARMACY. NURSING THEN ADDS A CH3034 BAG SPIKE ADAPTER ONTO THE SPIKE OF A REGULAR IV SET AND ADDS A SPINNING SPIROS CLOSED MALE LUER ADAPTER CH2000S TO THE LUER SIDE OF THE IV SET. THEY CONNECT THE CH3034 SPIROS TO THE IN-LINE BAG SPIKE AND THEN ATTACH THE SPINNING SPIROS TO THE ADDITIVE PORT ON THE BAG TO CIRCLE PRIME. THEY HAVE HAD MULTIPLES ISSUES WITH LEAKS FROM EITHER OF THE SPIROS PORTS ON THE CH3340 THAT WERE NOT REACTIVATING WHEN DISCONNECTING CAUSING UNSPECIFIED CHEMOTHERAPY TO SPILL. IN MOST OF THESE SITUATIONS, THE FLUID DOESN¿T PASS THROUGH TO THE LINE, SO THE NURSE GOES TO CHECK OR DETACH FROM THE BAG SPIKE SPILLING AFTER DISCONNECTION, AND ADDITIONALLY TH
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21911751 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K080989. One submission often covers many models, so check the product and lot details in each record.
- Z-0710-2021 · Aug 31, 2020 · Open, ClassifiedAppx 1.7 ml Bag Spike w/Integrated Clave, 2 Drop-In Spiros, REF: CH3260
Other Syringe, piston devices
- BD Luer-Lok 309657Becton, Dickinson And Company · Syringe, piston1,266 reports
- BD Luer-Lok 309653Becton, Dickinson And Company · Syringe, piston948 reports
- BD Luer-Lok 309628Becton, Dickinson And Company · Syringe, piston862 reports
- N/A 302995Becton, Dickinson And Company · Syringe, piston1 recall537 reports
- BD Plastipak Syringe 300865Becton, Dickinson And Company · Syringe, piston515 reports
- N/A 302830Becton, Dickinson And Company · Syringe, piston462 reports
- N/A 302832Becton, Dickinson And Company · Syringe, piston424 reports
- BD Ultra-Fine 328468Becton, Dickinson and Company Inc · Syringe, piston323 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page