ICU Medical CH-74-10Recalls & reported problems
Icu Medical, Inc. · Locking Universal Vented Vial Spike, Clave™, 10 Units · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00840619065706), received between Sep 14, 2020 and Sep 14, 2020 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (4 reports).
1 other recall cover devices cleared under the same FDA submission (K080989); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fluid/Blood Leak4
Patient problems
- Chemical Exposure4
- No Consequences Or Impact To Patient4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 14, 2020Fluid/Blood Leak; Fluid/Blood Leak
THE EVENT OCCURRED ON AN UNKNOWN DATE AND INVOLVED A LOCKING UNIVERSAL VENTED VIAL SPIKE, CLAVE, 10 UNITS THAT LEAKED CHEMOTHERAPY DURING PREPARATION. THE CUSTOMER REPORTED THAT SEVERAL PREPARERS HAD LEAKAGE ISSUES. THEY FOUND THE FILTER ASSOCIATED WITH THE SPIKE WAS NOT WORKING. AS A RESULT, LEAKS WERE OBSERVED AS WELL AS OVERPRESSURE INSIDE THE BOTTLE CAUSING PROJECTION INSIDE THE ENCLOSURE WHEN THE SPIKE IS REMOVED FROM THE BOTTLE. THE DRUGS INVOLVED WERE PACLITAXEL, ETOPOSIDE, AND DOCETAXEL (FAIRLY VISCOUS CHEMOTHERAPY). ADDITIONALLY, THE CUSTOMER STATED THERE WAS AN UNPROTECTED EXPOSURE TO CHEMOTHERAPY CAUSING THE CONTAMINATION OF THE PREPARER (GLOVES, JACKET), THE BOTTLE TO BE DISCARDED EACH TIME (FINANCIAL LOSS), CHANGING OF GLOVES, CHANGING THE STERILE FIELD AND CLEANING THE LAMINAR FLOW CHAMBER. THE SPILL WAS CLEANED PER PROTOCOL. NO FURTHER INFORMATION WAS PROVIDED. THE CUSTOMER REPORTED FOUR INCIDENTS. THIS REPORT CAPTURES THE FOURTH OF FOUR INCIDENTS.
Patient problems: Chemical Exposure; No Consequences Or Impact To Patient
MAUDE report 10529001 ↗MalfunctionSep 14, 2020Fluid/Blood Leak; Fluid/Blood Leak
THE EVENT OCCURRED ON AN UNKNOWN DATE AND INVOLVED A LOCKING UNIVERSAL VENTED VIAL SPIKE, CLAVE, 10 UNITS THAT LEAKED CHEMOTHERAPY DURING PREPARATION. THE CUSTOMER REPORTED THAT SEVERAL PREPARERS HAD LEAKAGE ISSUES. THEY FOUND THE FILTER ASSOCIATED WITH THE SPIKE WAS NOT WORKING. AS A RESULT, LEAKS WERE OBSERVED AS WELL AS OVERPRESSURE INSIDE THE BOTTLE CAUSING PROJECTION INSIDE THE ENCLOSURE WHEN THE SPIKE IS REMOVED FROM THE BOTTLE. THE DRUGS INVOLVED WERE PACLITAXEL, ETOPOSIDE, AND DOCETAXEL (FAIRLY VISCOUS CHEMOTHERAPY). ADDITIONALLY, THE CUSTOMER STATED THERE WAS AN UNPROTECTED EXPOSURE TO CHEMOTHERAPY CAUSING THE CONTAMINATION OF THE PREPARER (GLOVES, JACKET), THE BOTTLE TO BE DISCARDED EACH TIME (FINANCIAL LOSS), CHANGING OF GLOVES, CHANGING THE STERILE FIELD AND CLEANING THE LAMINAR FLOW CHAMBER. THE SPILL WAS CLEANED PER PROTOCOL. NO FURTHER INFORMATION WAS PROVIDED. THE CUSTOMER REPORTED FOUR INCIDENTS. THIS REPORT CAPTURES THE SECOND OF FOUR INCIDENTS.
Patient problems: Chemical Exposure; No Consequences Or Impact To Patient
MAUDE report 10528998 ↗MalfunctionSep 14, 2020Fluid/Blood Leak; Fluid/Blood Leak
THE EVENT OCCURRED ON AN UNKNOWN DATE AND INVOLVED A LOCKING UNIVERSAL VENTED VIAL SPIKE, CLAVE, 10 UNITS THAT LEAKED CHEMOTHERAPY DURING PREPARATION. THE CUSTOMER REPORTED THAT SEVERAL PREPARERS HAD LEAKAGE ISSUES. THEY FOUND THE FILTER ASSOCIATED WITH THE SPIKE WAS NOT WORKING. AS A RESULT, LEAKS WERE OBSERVED AS WELL AS OVERPRESSURE INSIDE THE BOTTLE CAUSING PROJECTION INSIDE THE ENCLOSURE WHEN THE SPIKE IS REMOVED FROM THE BOTTLE. THE DRUGS INVOLVED WERE PACLITAXEL, ETOPOSIDE, AND DOCETAXEL (FAIRLY VISCOUS CHEMOTHERAPY). ADDITIONALLY, THE CUSTOMER STATED THERE WAS AN UNPROTECTED EXPOSURE TO CHEMOTHERAPY CAUSING THE CONTAMINATION OF THE PREPARER (GLOVES, JACKET), THE BOTTLE TO BE DISCARDED EACH TIME (FINANCIAL LOSS), CHANGING OF GLOVES, CHANGING THE STERILE FIELD AND CLEANING THE LAMINAR FLOW CHAMBER. THE SPILL WAS CLEANED PER PROTOCOL. NO FURTHER INFORMATION WAS PROVIDED. THE CUSTOMER REPORTED FOUR INCIDENTS. THIS REPORT CAPTURES THE THIRD OF FOUR INCIDENTS.
Patient problems: Chemical Exposure; No Consequences Or Impact To Patient
MAUDE report 10528997 ↗MalfunctionSep 14, 2020Fluid/Blood Leak; Fluid/Blood Leak
THE EVENT OCCURRED ON AN UNKNOWN DATE AND INVOLVED A LOCKING UNIVERSAL VENTED VIAL SPIKE, CLAVE, 10 UNITS THAT LEAKED CHEMOTHERAPY DURING PREPARATION. THE CUSTOMER REPORTED THAT SEVERAL PREPARERS HAD LEAKAGE ISSUES. THEY FOUND THE FILTER ASSOCIATED WITH THE SPIKE WAS NOT WORKING. AS A RESULT, LEAKS WERE OBSERVED AS WELL AS OVERPRESSURE INSIDE THE BOTTLE CAUSING PROJECTION INSIDE THE ENCLOSURE WHEN THE SPIKE IS REMOVED FROM THE BOTTLE. THE DRUGS INVOLVED WERE PACLITAXEL, ETOPOSIDE, AND DOCETAXEL (FAIRLY VISCOUS CHEMOTHERAPY). ADDITIONALLY, THE CUSTOMER STATED THERE WAS AN UNPROTECTED EXPOSURE TO CHEMOTHERAPY CAUSING THE CONTAMINATION OF THE PREPARER (GLOVES, JACKET), THE BOTTLE TO BE DISCARDED EACH TIME (FINANCIAL LOSS), CHANGING OF GLOVES, CHANGING THE STERILE FIELD AND CLEANING THE LAMINAR FLOW CHAMBER. THE SPILL WAS CLEANED PER PROTOCOL. NO FURTHER INFORMATION WAS PROVIDED. THE CUSTOMER REPORTED FOUR INCIDENTS. THIS REPORT CAPTURES THE FIRST OF FOUR INCIDENTS.
Patient problems: Chemical Exposure; No Consequences Or Impact To Patient
MAUDE report 10528992 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K080989. One submission often covers many models, so check the product and lot details in each record.
- Z-0710-2021 · Aug 31, 2020 · Open, ClassifiedAppx 1.7 ml Bag Spike w/Integrated Clave, 2 Drop-In Spiros, REF: CH3260
Other Syringe, piston devices
- BD Luer-Lok 309657Becton, Dickinson And Company · Syringe, piston1,266 reports
- BD Luer-Lok 309653Becton, Dickinson And Company · Syringe, piston948 reports
- BD Luer-Lok 309628Becton, Dickinson And Company · Syringe, piston862 reports
- N/A 302995Becton, Dickinson And Company · Syringe, piston1 recall537 reports
- BD Plastipak Syringe 300865Becton, Dickinson And Company · Syringe, piston515 reports
- N/A 302830Becton, Dickinson And Company · Syringe, piston462 reports
- N/A 302832Becton, Dickinson And Company · Syringe, piston424 reports
- BD Ultra-Fine 328468Becton, Dickinson and Company Inc · Syringe, piston323 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page