LeMaitre Embolectomy Catheter 1651-88Recalls & reported problems

Lemaitre Vascular, Inc. · 5F Plus 80CM Over-the-Wire Embolectomy Catheter · Class 2

All models of LeMaitre Embolectomy Catheter (47)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This LeMaitre embolectomy catheter is a balloon catheter used to remove clots from blood vessels. The 162 reports are all malfunctions, mostly describing the balloon rupturing during procedures, and include no deaths or injuries. There is 1 open recall, started in April 2025, over packages that may have incomplete seals found in internal testing.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
153
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 153 adverse event reports that list this exact device (DI 00840663100750), received between May 22, 2018 and Sep 29, 2026 — classified as 153 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (141 reports).

1 FDA recall record name this device's identifier, 1 still open.

10 other recalls cover devices cleared under the same FDA submission (K022145); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 10 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 4 reportsMay 2024: 4 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 5 reportsOct 2024: 4 reportsNov 2024: 9 reportsDec 2024: 0 reportsJan 2025: 11 reportsFeb 2025: 4 reportsMar 2025: 1 reportsApr 2025: 8 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 3 reportsAug 2025: 2 reportsSep 2025: 5 reportsOct 2025: 4 reportsNov 2025: 5 reportsDec 2025: 1 reportsJan 2026: 2 reportsFeb 2026: 3 reportsMar 2026: 4 reportsApr 2026: 19 reportsMay 2026: 7 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 2 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202619
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 11, 2025Class IIOpen, Classified

    During internal product testing, it was observed that some packages to have incomplete seals (sterile barrier).

    Z-1750-2025 · Root cause: Packaging process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Rupture141
  • Deflation Problem4
  • Failure to Deflate3
  • Burst Container or Vessel2
  • Material Integrity Problem2
  • Material Puncture/Hole1
  • Material Separation1
  • Unraveled Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions144
  • No Consequences Or Impact To Patient5
  • No Known Impact Or Consequence To Patient3
  • Administration Site Reaction1

Recent adverse event reports

The 25 most recent of 153 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Material Rupture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50935643 ↗
  • MalfunctionSep 28, 2026Material Rupture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50919477 ↗
  • MalfunctionMay 26, 2026Material Rupture

    IT WAS REPORTED THAT THREE (3) TUFTEX OTW BALLOON RUPTURED WHILE REMOVING A BLOOD CLOT DURING A PROCEDURE. THE DEVICES WERE CHECKED PRIOR TO USE. NO PATIENT INJURY WAS REPORTED. NOTE: THIS IS REPORT 3 OF 3 RELATED TO THE THIRD CATHETER USED IN THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25280529 ↗
  • MalfunctionMay 26, 2026Material Rupture

    IT WAS REPORTED THAT THREE (3) TUFTEX OTW BALLOON RUPTURED WHILE REMOVING A BLOOD CLOT DURING A PROCEDURE. THE DEVICES WERE CHECKED PRIOR TO USE. NO PATIENT INJURY WAS REPORTED. NOTE: THIS IS REPORT 2 OF 3 RELATED TO THE SECOND CATHETER USED IN THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25279476 ↗
  • MalfunctionMay 26, 2026Material Rupture

    IT WAS REPORTED THAT THREE (3) TUFTEX OTW BALLOON RUPTURED WHILE REMOVING A BLOOD CLOT DURING A PROCEDURE. THE DEVICES WERE CHECKED PRIOR TO USE. NO PATIENT INJURY WAS REPORTED. NOTE: THIS IS REPORT 1 OF 3 RELATED TO THE FIRST CATHETER USED IN THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25278313 ↗
  • MalfunctionMay 19, 2026Material Rupture

    IT WAS REPORTED THAT THE TUFTEX OTW BALLOON RUPTURED DURING USE IN A PROCEDURE. THE DEVICE WAS CHECKED PRIOR TO USE. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25225196 ↗
  • MalfunctionMay 19, 2026Material Rupture; Unraveled Material

    IT WAS REPORTED THAT THE TUFTEX OTW BALLOON RUPTURED AND THE LIGATURE UNRAVELED DURING USE IN A PROCEDURE. THE DEVICE WAS CHECKED PRIOR TO USE. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25225081 ↗
  • MalfunctionMay 19, 2026Material Rupture

    IT WAS REPORTED THAT THE TUFTEX OTW BALLOON RUPTURED DURING USE IN A PROCEDURE. THE DEVICE WAS CHECKED PRIOR TO USE. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25224801 ↗
  • MalfunctionMay 7, 2026Material Rupture

    IT WAS REPORTED THAT THE TUFTEX OTW BALLOON RUPTURED DURING USE IN A PROCEDURE. THE DEVICE WAS CHECKED PRIOR TO USE. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25110348 ↗
  • MalfunctionApr 29, 2026Material Rupture

    IT WAS REPORTED THAT THE BALLOON RUPTURED DURING PRE-USE TESTING. THE PROCEDURE WAS COMPLETED USING A REPLACEMENT PRODUCT. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25026764 ↗
  • MalfunctionApr 28, 2026Material Rupture

    IT WAS REPORTED THAT THE TUFTEX OTW BALLOON RUPTURED WHILE REMOVING A BLOOD CLOT DURING A PROCEDURE. THE DEVICE WAS CHECKED PRIOR TO USE. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25017008 ↗
  • MalfunctionApr 28, 2026Material Rupture

    IT WAS REPORTED THAT THE TUFTEX OTW BALLOON RUPTURED DURING USE IN A PROCEDURE. IT WAS CHECKED PRIOR TO USE. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25016983 ↗
  • MalfunctionApr 28, 2026Material Rupture

    IT WAS REPORTED THAT TWO (2) TUFTEX OTW CATHETERS EXPERIENCED BALLOON RUPTURE WHILE REMOVING A BLOOD CLOT DURING A PROCEDURE. THE DEVICES WERE CHECKED PRIOR TO USE. NO PATIENT INJURY WAS REPORTED. NOTE: THIS IS REPORT 2 OF 2 RELATED TO THE SECOND CATHETER USED IN THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25016416 ↗
  • MalfunctionApr 28, 2026Material Rupture

    IT WAS REPORTED THAT TWO (2) TUFTEX OTW CATHETERS EXPERIENCED BALLOON RUPTURE WHILE REMOVING A BLOOD CLOT DURING A PROCEDURE. THE DEVICES WERE CHECKED PRIOR TO USE. NO PATIENT INJURY WAS REPORTED. NOTE: THIS IS REPORT 1 OF 2 RELATED TO THE FIRST CATHETER USED IN THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25016180 ↗
  • MalfunctionApr 21, 2026Material Rupture

    IT WAS REPORTED THAT THE TUFTEX OTW BALLOON RUPTURED DURING A PROCEDURE. THE DEVICE WAS CHECKED PRIOR TO USE. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24940136 ↗
  • MalfunctionApr 21, 2026Material Rupture

    IT WAS REPORTED THAT THE TUFTEX OTW BALLOON RUPTURED DURING A PROCEDURE. THE DEVICE WAS CHECKED PRIOR TO USE. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24940116 ↗
  • MalfunctionApr 16, 2026Material Rupture

    IT WAS REPORTED THAT THE TUFTEX OTW BALLOON RUPTURED WHILE REMOVING A BLOOD CLOT DURING A PROCEDURE. THE DEVICE WAS CHECKED PRIOR TO USE. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24901901 ↗
  • MalfunctionApr 16, 2026Material Rupture

    IT WAS REPORTED THAT TWO (2) TUFTEX OTW BALLOON RUPTURED DURING USE IN A PROCEDURE. THE DEVICE WAS CHECKED PRIOR TO USE. NO PATIENT INJURY WAS REPORTED. NOTE: THIS IS REPORT 2 OF 2 RELATED TO THE SECOND CATHETER USED IN THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24901523 ↗
  • MalfunctionApr 16, 2026Material Rupture

    IT WAS REPORTED THAT TWO (2) TUFTEX OTW BALLOON RUPTURED DURING USE IN A PROCEDURE. THE DEVICE WAS CHECKED PRIOR TO USE. NO PATIENT INJURY WAS REPORTED. NOTE: THIS IS REPORT 1 OF 2 RELATED TO THE FIRST CATHETER USED IN THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24901513 ↗
  • MalfunctionApr 14, 2026Material Rupture

    IT WAS REPORTED THAT TWO (2) TUFTEX OTW BALLOONS RUPTURED DURING BLOOD CLOT REMOVAL. THE DEVICES WERE CHECKED PRIOR TO USE. NO PATIENT INJURY WAS REPORTED. NOTE: THIS IS REPORT 2 OF 2 RELATED TO THE SECOND CATHETER USED IN THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24873664 ↗
  • MalfunctionApr 14, 2026Material Rupture

    IT WAS REPORTED THAT TWO (2) TUFTEX OTW BALLOONS RUPTURED DURING BLOOD CLOT REMOVAL. THE DEVICES WERE CHECKED PRIOR TO USE. NO PATIENT INJURY WAS REPORTED. NOTE: THIS IS REPORT 1 OF 2 RELATED TO THE FIRST CATHETER USED IN THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24873646 ↗
  • MalfunctionApr 9, 2026Material Rupture

    IT WAS REPORTED THAT TWO TUFTEX OTW BALLOON RUPTURED DURING A PROCEDURE. THE DEVICES WERE CHECKED PRIOR TO USE. NO PATIENT INJURY WAS REPORTED. NOTE: THIS IS REPORT 2 OF 2 RELATED TO THE SECOND CATHETER USED IN THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24828932 ↗
  • MalfunctionApr 9, 2026Material Rupture

    IT WAS REPORTED THAT TWO TUFTEX OTW BALLOON RUPTURED DURING A PROCEDURE. THE DEVICES WERE CHECKED PRIOR TO USE. NO PATIENT INJURY WAS REPORTED. NOTE: THIS IS REPORT 1 OF 2 RELATED TO THE FIRST CATHETER USED IN THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24828873 ↗
  • MalfunctionApr 9, 2026Material Rupture

    IT WAS REPORTED THAT THE TUFTEX OTW BALLOON RUPTURED DURING A PROCEDURE. THE DEVICE WAS CHECKED PRIOR TO USE. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24828306 ↗
  • MalfunctionApr 6, 2026Material Rupture

    IT WAS REPORTED THAT THE TUFTEX OTW BALLOON RUPTURED DURING INSERTION FOR A CATHETERIZATION PROCEDURE. A REPLACEMENT DEVICE WAS USED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24791628 ↗

FDA clearance and classification

Cleared through 510(k) K022145 on Jan 14, 2003 (decision: Substantially Equivalent). Submitted by LeMaitre Vascular, Inc. as “Over The Wire Embolectomy Catheter, Model 2302”. The FDA review took 196 days. FDA record ↗

The FDA regulates product code DXE (catheter, embolectomy) under 21 CFR 870.5150 as a Class II device. 21 CFR 870.5150 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K022145. One submission often covers many models, so check the product and lot details in each record.

  • Z-0117-2027 · Aug 28, 2026 · Open, Classified
    TufTex Over-the-Wire Embolectomy Catheter; Catalog Number: E1651-88;
  • Z-1567-2020 · Mar 23, 2020 · Open, Classified
    LeMaitre 5F Plus Over the Wire Embolectomy Catheter, Model Nos. 1651-84, e1651-84, 1651-88, e1651-88. The LeMaitre Over the Wire Embolectomy Catheter is indicated for use in the r
  • Z-0380-2016 · Oct 21, 2015 · Terminated
    LeMaitre Over-the-Wire Embolectomy Catheter 4F, 80cm, Model 1651-48 and e1654-48. Indicated for the removal of arterial emboli and thrombi.
  • Z-0381-2016 · Oct 21, 2015 · Terminated
    LeMaitre Over-the-Wire Embolectomy Catheter 5F Plus, 80cm, Model 1651-88 and e1654-88. Indicated for the removal of arterial emboli and thrombi.
  • Z-0385-2016 · Oct 21, 2015 · Terminated
    LeMaitre Over-the-Wire Embolectomy Catheter 6F, 80 cm, Model e-1651-68. Indicated for the removal of arterial emboli and thrombi.
  • Z-0384-2016 · Oct 21, 2015 · Terminated
    LeMaitre Over-the-Wire Embolectomy Catheter 6F, 40 cm, Model e-1651-64. Indicated for the removal of arterial emboli and thrombi.
  • Z-0379-2016 · Oct 21, 2015 · Terminated
    LeMaitre Over-the-Wire Embolectomy Catheter 3F, 80cm, Model 1651-38. Indicated for the removal of arterial emboli and thrombi.
  • Z-0383-2016 · Oct 21, 2015 · Terminated
    LeMaitre Over-the-Wire Embolectomy Catheter 5F Plus, 40 cm, Model e1651-84. Indicated for the removal of arterial emboli and thrombi.
  • Z-0382-2016 · Oct 21, 2015 · Terminated
    LeMaitre Over-the-Wire Embolectomy Catheter 7F, 80cm, Model 1654-78 and e1654-78. Indicated for the removal of arterial emboli and thrombi.
  • Z-0378-2016 · Oct 21, 2015 · Terminated
    LeMaitre Over-the-Wire Embolectomy Catheter 3F, 40cm, Model 1651-34 and e1651-34. Indicated for the removal of arterial emboli and thrombi.

Common questions about the LeMaitre Embolectomy Catheter 1651-88

Has the LeMaitre Embolectomy Catheter 1651-88 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1750-2025 (Class II), began on Apr 11, 2025: During internal product testing, it was observed that some packages to have incomplete seals (sterile barrier). If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the LeMaitre Embolectomy Catheter 1651-88?

The FDA's MAUDE database lists 153 adverse event reports for this device, received between May 22, 2018 and Sep 29, 2026. In those reports, the most common specific device problems are “Material Rupture” (141), “Deflation Problem” (4) and “Failure to Deflate” (3); the most common patient problem is “Administration Site Reaction” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the LeMaitre Embolectomy Catheter 1651-88?

None of the reports are classified as deaths, 0 as injuries and 153 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the LeMaitre Embolectomy Catheter 1651-88 FDA approved?

It was cleared through 510(k) K022145 on Jan 14, 2003. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the LeMaitre Embolectomy Catheter 1651-88, and is it still sold?

The FDA's device database lists Lemaitre Vascular, Inc. as the company responsible for it (brand LeMaitre Embolectomy Catheter, model 1651-88). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Catheter, Embolectomy devices

All Catheter, Embolectomy devices →