LeMaitre Embolectomy Catheter e1651-88Recalls & reported problems
Lemaitre Vascular, Inc. · 5F Plus 80CM Over-the-Wire Embolectomy Catheter, eIFU · Class 2
All models of LeMaitre Embolectomy Catheter (47)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00840663100866), received between Apr 18, 2018 and Sep 1, 2026 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (2 reports).
2 FDA recall records name this device's identifier, 2 still open.
9 other recalls cover devices cleared under the same FDA submission (K022145); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Aug 28, 2026Class IIOpen, Classified
Devices failed in process testing related to the amount of pull force it can sustain prior to the balloon components failing.
Z-0117-2027 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Apr 11, 2025Class IIOpen, Classified
During internal product testing, it was observed that some packages to have incomplete seals (sterile barrier).
Z-1750-2025 · Root cause: Packaging process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Rupture2
- Deflation Problem1
- Material Fragmentation1
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 1, 2026Material Fragmentation; Material Separation
We have determined that (b)(4) units from TufTex® Over-the-Wire Embolectomy Catheter Lot number: XOT00002181 were inadvertently sold to 11 US customers prior to being released. The lot was on hold for an NCMR investigation, where 1 out of 8 samples failed pull testing during manufacturing.It is possible that products from this lot are at a higher risk of experiencing ligature slip resulting in balloon deflation issues or balloon detachment during use. This failure could lead to adverse events during use. However, no complaints have been received related to these devices.Based on the issue, a Field Corrective Action is being initiated by LeMaitre to recall these devices (Class II Recall).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50549488 ↗MalfunctionFeb 13, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT THE DOCTOR EXPERIENCED 3 RECENT BALLOON RUPTURES WITH THE TUFTEX OVER-THE-WIRE CATHETERS. AN OPERATION ON A PATIENT ON (B)(6) 2025 EXPERIENCED 2 RUPTURES AND AN OPERATION ON A PATIENT ON (B)(6) 2025 EXPERIENCED 1 RUPTURE. THIS IS REPORT 3 OF 3 RELATED TO THE CATHETER USED ON THE SECOND PATIENT. IT WAS REPORTED THAT THE TUFTEX OTW BALLOON RUPTURED DURING THE PROCEDURE. THE RUPTURES EITHER OCCUR ON AN INITIAL PULLBACK OR AFTER THE INITIAL PULLBACK WAS SUCCESSFUL AND A REPEATED PULLBACK IS REQUIRED. NO INJURY WAS REPORTED TO THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE DEVICES WAS DISCARDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21381718 ↗MalfunctionFeb 13, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT THE DOCTOR EXPERIENCED 3 RECENT BALLOON RUPTURES WITH THE TUFTEX OVER-THE-WIRE CATHETERS. AN OPERATION ON A PATIENT ON (B)(6) 2025 EXPERIENCED 2 RUPTURES AND AN OPERATION ON A PATIENT ON (B)(6) 2025 EXPERIENCED 1 RUPTURE. THIS IS REPORT 2 OF 3 RELATED TO THE SECOND CATHETER USED ON THE FIRST PATIENT. IT WAS REPORTED THAT THE SECOND TUFTEX OTW BALLOON RUPTURED DURING THE PROCEDURE. THE RUPTURES EITHER OCCUR ON AN INITIAL PULLBACK OR AFTER THE INITIAL PULLBACK WAS SUCCESSFUL AND A REPEATED PULLBACK IS REQUIRED. NO INJURY WAS REPORTED TO THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE DEVICES WERE DISCARDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21381701 ↗MalfunctionApr 18, 2018Deflation Problem; Deflation Problem
AFTER DECLOTTING THE AV FISTULA, THE PHYSICIAN INFLATED THE CATHETER ON THE ARTERIAL SIDE. HOWEVER,THE PHYSICIAN COULD NOT DEFLATE THE BALLOON. HE HAD TO DEFLATE THE BALLOON USING A NEEDLE IN ORDER TO REMOVE THE CATHETER OUT OF THE PATIENT'S ARTERY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7441565 ↗
FDA clearance and classification
Cleared through 510(k) K022145 on Jan 14, 2003 (decision: Substantially Equivalent). Submitted by LeMaitre Vascular, Inc. as “Over The Wire Embolectomy Catheter, Model 2302”. The FDA review took 196 days. FDA record ↗
The FDA regulates product code DXE (catheter, embolectomy) under 21 CFR 870.5150 as a Class II device. 21 CFR 870.5150 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K022145. One submission often covers many models, so check the product and lot details in each record.
- Z-1567-2020 · Mar 23, 2020 · Open, ClassifiedLeMaitre 5F Plus Over the Wire Embolectomy Catheter, Model Nos. 1651-84, e1651-84, 1651-88, e1651-88. The LeMaitre Over the Wire Embolectomy Catheter is indicated for use in the r
- Z-0383-2016 · Oct 21, 2015 · TerminatedLeMaitre Over-the-Wire Embolectomy Catheter 5F Plus, 40 cm, Model e1651-84. Indicated for the removal of arterial emboli and thrombi.
- Z-0384-2016 · Oct 21, 2015 · TerminatedLeMaitre Over-the-Wire Embolectomy Catheter 6F, 40 cm, Model e-1651-64. Indicated for the removal of arterial emboli and thrombi.
- Z-0381-2016 · Oct 21, 2015 · TerminatedLeMaitre Over-the-Wire Embolectomy Catheter 5F Plus, 80cm, Model 1651-88 and e1654-88. Indicated for the removal of arterial emboli and thrombi.
- Z-0382-2016 · Oct 21, 2015 · TerminatedLeMaitre Over-the-Wire Embolectomy Catheter 7F, 80cm, Model 1654-78 and e1654-78. Indicated for the removal of arterial emboli and thrombi.
- Z-0378-2016 · Oct 21, 2015 · TerminatedLeMaitre Over-the-Wire Embolectomy Catheter 3F, 40cm, Model 1651-34 and e1651-34. Indicated for the removal of arterial emboli and thrombi.
- Z-0385-2016 · Oct 21, 2015 · TerminatedLeMaitre Over-the-Wire Embolectomy Catheter 6F, 80 cm, Model e-1651-68. Indicated for the removal of arterial emboli and thrombi.
- Z-0380-2016 · Oct 21, 2015 · TerminatedLeMaitre Over-the-Wire Embolectomy Catheter 4F, 80cm, Model 1651-48 and e1654-48. Indicated for the removal of arterial emboli and thrombi.
- Z-0379-2016 · Oct 21, 2015 · TerminatedLeMaitre Over-the-Wire Embolectomy Catheter 3F, 80cm, Model 1651-38. Indicated for the removal of arterial emboli and thrombi.
Common questions about the LeMaitre Embolectomy Catheter e1651-88
Has the LeMaitre Embolectomy Catheter e1651-88 been recalled?
Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-0117-2027 (Class II), began on Aug 28, 2026: Devices failed in process testing related to the amount of pull force it can sustain prior to the balloon components failing. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the LeMaitre Embolectomy Catheter e1651-88?
The FDA's MAUDE database lists 4 adverse event reports for this device, received between Apr 18, 2018 and Sep 1, 2026. In those reports, the most common specific device problems are “Material Rupture” (2), “Deflation Problem” (1) and “Material Fragmentation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the LeMaitre Embolectomy Catheter e1651-88?
None of the reports are classified as deaths, 0 as injuries and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the LeMaitre Embolectomy Catheter e1651-88 FDA approved?
It was cleared through 510(k) K022145 on Jan 14, 2003. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the LeMaitre Embolectomy Catheter e1651-88, and is it still sold?
The FDA's device database lists Lemaitre Vascular, Inc. as the company responsible for it (brand LeMaitre Embolectomy Catheter, model e1651-88). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Catheter, Embolectomy devices
- Indigo System CAT6Penumbra, Inc. · CATHETER, EMBOLECTOMY2 deaths408 reports
- Indigo System CATRXKITPenumbra, Inc. · Catheter, Embolectomy1 death339 reports
- Fogarty 120804FPEdwards Lifesciences LLC · Catheter, Embolectomy1 recall1 death200 reports
- LeMaitre Embolectomy Catheter 1651-88Lemaitre Vascular, Inc. · Catheter, Embolectomy1 recall153 reports
- FOGARTY 120404FPEdwards Lifesciences LLC · Catheter, Embolectomy1 recall61 reports
- Indigo System CAT6KIT-APenumbra, Inc. · Catheter, Embolectomy109 reports
- Indigo System CAT8XTORQ115Penumbra, Inc. · CATHETER, EMBOLECTOMY2 deaths98 reports
- Fogarty 120805FPEdwards Lifesciences LLC · Catheter, Embolectomy1 recall45 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page