CARESCAPE R860Recalls & reported problems

Datex-Ohmeda Inc · CARESCAPE R860 · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Datex-Ohmeda CARESCAPE R860 is a facility-use ventilator. Its 4,378 adverse event reports mostly describe insufficient flow, failure to run on battery and other power problems. The reports include 1 death and 17 injuries. 1 recall names this device: started in April 2022 and still open, it covers a battery backup that is too weak, so the ventilator can shut down early when not plugged into mains power.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
4,346
Deaths reported
1
Injuries reported
17
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4,346 adverse event reports that list this exact device (DI 00840682102346), received between Jan 30, 2018 and Sep 30, 2026 — classified as 1 death, 17 injuries, 4,327 malfunctions.

The most frequently reported specific device problem is “Insufficient Flow or Under Infusion” (1,914 reports).

1 FDA recall record name this device's identifier, 1 still open.

5 other recalls cover devices cleared under the same FDA submission (K142679, K210384); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 140 reportsDec 2023: 106 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 162 reportsApr 2024: 111 reportsMay 2024: 127 reportsJun 2024: 127 reportsJul 2024: 132 reportsAug 2024: 107 reportsSep 2024: 159 reportsOct 2024: 150 reportsNov 2024: 131 reportsDec 2024: 132 reportsJan 2025: 150 reportsFeb 2025: 108 reportsMar 2025: 153 reportsApr 2025: 97 reportsMay 2025: 121 reportsJun 2025: 134 reportsJul 2025: 112 reportsAug 2025: 137 reportsSep 2025: 96 reportsOct 2025: 129 reportsNov 2025: 135 reportsDec 2025: 203 reportsJan 2026: 199 reportsFeb 2026: 135 reportsMar 2026: 162 reportsApr 2026: 195 reportsMay 2026: 144 reportsJun 2026: 129 reportsJul 2026: 3 reportsAug 2026: 2 reportsSep 2026: 65 reportsOct 2026: 0 reports202420252026203
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 18, 2022Open, Classified

    Insufficient battery backup power resulting in premature shutdown of the ventilator when not connected to AC mains power supply.

    Z-1226-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Insufficient Flow or Under Infusion1,914
  • Failure to Run on Battery1,098
  • Power Problem340
  • Mechanical Problem292
  • Therapeutic or Diagnostic Output Failure142
  • No Apparent Adverse Event119
  • Unexpected Therapeutic Results106
  • Failure to Power Up95

Patient problems

  • No Clinical Signs, Symptoms or Conditions4,311
  • Low Oxygen Saturation16
  • No Patient Involvement12
  • Capsular Contracture2
  • Cardiac Arrest2
  • No Consequences Or Impact To Patient2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Muscle Hypotonia1

Recent adverse event reports

The 25 most recent of 4,346 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Failure to Run on Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50952203 ↗
  • MalfunctionSep 30, 2026Insufficient Flow or Under Infusion

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50952183 ↗
  • MalfunctionSep 30, 2026Measurement System Incompatibility

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951867 ↗
  • MalfunctionSep 30, 2026Insufficient Flow or Under Infusion

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949732 ↗
  • MalfunctionSep 30, 2026Failure to Power Up

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949635 ↗
  • MalfunctionSep 30, 2026Insufficient Flow or Under Infusion

    It was reported that there was a malfunction resulting in insufficient O2 during pre-use checkout. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946828 ↗
  • MalfunctionSep 30, 2026Failure to Run on Battery

    It was reported that there was a battery failure during pre-use checkout that could cause loss of mechanical ventilation. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945593 ↗
  • MalfunctionSep 30, 2026Mechanical Problem

    It was reported that there was a malfunction during pre-use checkout resulting in loss of ventilation.There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945540 ↗
  • MalfunctionSep 30, 2026Failure to Run on Battery

    It was reported that there was a battery failure that could cause loss of mechanical ventilation during pre-use checkout. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50944267 ↗
  • MalfunctionSep 30, 2026Failure to Run on Battery

    It was reported that there was a battery failure that could cause loss of mechanical ventilation during pre-use checkout. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50944102 ↗
  • MalfunctionSep 29, 2026Insufficient Flow or Under Infusion

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50940951 ↗
  • MalfunctionSep 29, 2026Insufficient Flow or Under Infusion

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50936810 ↗
  • MalfunctionSep 29, 2026Insufficient Flow or Under Infusion

    It was reported that there was a malfunction resulting in insufficient O2 during planned maintenance. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50934340 ↗
  • MalfunctionSep 29, 2026Failure to Run on Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50934124 ↗
  • MalfunctionSep 29, 2026Insufficient Flow or Under Infusion

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50934123 ↗
  • MalfunctionSep 29, 2026Insufficient Flow or Under Infusion

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50933099 ↗
  • MalfunctionSep 29, 2026Insufficient Flow or Under Infusion

    It was reported that there was a malfunction resulting in insufficient O2 during pre-use checkout. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50932972 ↗
  • MalfunctionSep 29, 2026Failure to Run on Battery

    It was reported that there was a battery failure that could cause loss of mechanical ventilation during pre-use checkout. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50932797 ↗
  • MalfunctionSep 29, 2026Failure to Run on Battery

    It was reported that there was a battery failure during pre-use checkout that could cause loss of mechanical ventilation. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50930553 ↗
  • MalfunctionSep 29, 2026Failure to Run on Battery

    It was reported that there was a battery failure during pre-use checkout that could cause loss of mechanical ventilation. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50930552 ↗
  • MalfunctionSep 29, 2026Failure to Run on Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50930547 ↗
  • MalfunctionSep 28, 2026Failure to Run on Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50926665 ↗
  • MalfunctionSep 28, 2026Insufficient Flow or Under Infusion

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50926509 ↗
  • MalfunctionSep 28, 2026Insufficient Flow or Under Infusion

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50917928 ↗
  • MalfunctionSep 28, 2026Improper Flow or Infusion

    It was reported that there was a malfunction resulting in insufficient O2. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50917871 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K142679, K210384. One submission often covers many models, so check the product and lot details in each record.

  • Z-1227-2022 · Apr 18, 2022 · Open, Classified
    Replacement back up batteries distributed on or after April 1, 2019, for CARESCAPE R860, Engstrom Carestation and Engstrom PRO Ventilators.
  • Z-2883-2020 · Jul 14, 2020 · Terminated
    CARESCAPE R860 Ventilator The CARESCAPE R860 ventilator is a microprocessor based, electronically controlled, pneumatically driven ventilator that includes integrated monitoring o
  • Z-0463-2019 · Oct 5, 2018 · Terminated
    GE HEALTHCARE CARESCAPE R860 Inspiratory Safety Guard, Part Numbers: a) 2066713-001 (single pack) b) 2083208-001 (box of ten singles) The CARESCAPE R860 ventilator is designed to
  • Z-2738-2018 · Jun 28, 2018 · Terminated
    The Neonatal Flow Sensor Cable used on devices CARESCAPE R860, ENGSTROM CARESTATION, ENGSTROM PRO and contained in Neonatal Software Upgrade Kit 2080496-010. The CARESCAPE R860 v
  • Z-2266-2018 · Dec 8, 2017 · Terminated
    CARESCAPE R860 ventilators with software version 10SP05 Product Usage: The CARESCAPE R860 ventilator is designed to provide mechanical ventilation or support to neonatal, pediatri

Other VENTILATOR, CONTINUOUS, FACILITY USE devices

All VENTILATOR, CONTINUOUS, FACILITY USE devices →

Sources and updates