AW Server 3.2Recalls & reported problems

Ge Medical Systems · AW Server 3.2 Medical Device · Class 2

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

2 FDA recall records name this device's identifier, 2 still open.

3 other recalls cover devices cleared under the same FDA submission (K081985); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. May 8, 2026Open, Classified

    GE HealthCare has become aware of a context synchronization issue in AW Server 3.2 ext. 6.5. When a user selects a patient or exam in the AW Server Web Client worklist and launches an interactive application (e.g. Volume Viewer), the application may open the previous patient's exam instead of the intended one. When this issue occurs, there is no system warning or error notification. If the error is not recognized, a clinical user could review, interpret, or report images for the wrong patient which could lead to misdiagnosis, incorrect clinical decisions, resulting in delayed or incorrect treatment.

    Z-2553-2026 · Root cause: Software design · FDA record ↗

  2. Sep 5, 2025Open, Classified

    Firm has identified a security vulnerability in AW Server products. If exploited, a malicious actor could compromise the confidentiality, integrity, and availability of patient data.

    Z-0068-2026 · Root cause: Software in the Use Environment · FDA record ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K081985. One submission often covers many models, so check the product and lot details in each record.

  • Z-0270-2014 · Sep 19, 2013 · Terminated
    AW VolumeShare 4 (ADVANTAGE WORKSTATION 4.3.), AW VolumeShare 5 (AW VOLUMESHARE 5 WITH ANGIOVIZ OPTION), AW Server. K052995: Advantage Workstation 4.3 is a review station, which
  • Z-1137-2013 · Jun 28, 2012 · Terminated
    GE Healthcare, AW Server. Product Usage: . AW Server is a medical software system that allows multiple users to remotely access AW applications from compatible computers on a net
  • Z-0486-2012 · Nov 17, 2011 · Terminated
    GE Healthcare AW Server version 2.0 with release 2.0 - 4.0 A medical software system that allows multiple users to remotely access AW applications from compatible computers on a

Other System, image processing, radiological devices

All System, image processing, radiological devices →