Skip to content

Discovery PETCT 610Recalls & reported problems

GE Medical Systems, L.L.C. · Discovery 610 PET Gantry with Labels, Eagle · Class 2

All models of Discovery (8)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Jun 2, 2022Class IIOpen, Classified

    As a result of unclear guidance in the service manual, the table can drop during a specific service activity if the wrong screws are removed by the service personnel without the appropriate table supports in place resulting in injury to service personnel.

    Z-1511-2022 · Root cause: Error in labeling · Recall details · FDA record ↗

FDA clearance and classification

The FDA regulates product code KPS (system, tomography, computed, emission) under 21 CFR 892.1200 as a Class II device. 21 CFR 892.1200 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Discovery PETCT 610

Has the Discovery PETCT 610 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1511-2022 (Class II), began on Jun 2, 2022: As a result of unclear guidance in the service manual, the table can drop during a specific service activity if the wrong screws are removed by the service personnel without the appropriate table supports in place… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

Who makes the Discovery PETCT 610, and is it still sold?

The FDA's device database lists GE Medical Systems, L.L.C. as the company responsible for it (brand Discovery, model PETCT 610). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other System, tomography, computed, emission devices

All System, tomography, computed, emission devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page