Skip to content

Discovery NM 530c H3000SARecalls & reported problems

G E Medical Systems Israel Ltd. · UFC GANTRY FINAL ASSEMBLY · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

2 other recalls cover devices cleared under the same FDA submission (K080124); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Jun 2, 2022Class IIOpen, Classified

    As a result of unclear guidance in the service manual, the table can drop during a specific service activity if the wrong screws are removed by the service personnel without the appropriate table supports in place resulting in injury to service personnel.

    Z-1511-2022 · Root cause: Error in labeling · Recall details · FDA record ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K080124. One submission often covers many models, so check the product and lot details in each record.

  • Z-2259-2020 · Nov 30, 2018 · Open, Classified
    System, Tomography Computed Emmission
  • Z-1168-2013 · Jan 9, 2013 · Terminated
    Ventri Discovery NM 530c, Ventri with Premium Table Discovery NM 530c, models H3000ZW & H3000SA. The system is used to perform nuclear imaging procedures.

Other System, tomography, computed, emission devices

All System, tomography, computed, emission devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page