Optima XR220amxRecalls & reported problems
GE Medical Systems, L.L.C. · Optima XR220 Designated Medical Device Assembly · Class 2
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
2 FDA recall records name this device's identifier, 2 still open.
6 other recalls cover devices cleared under the same FDA submission (K142383); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Jun 5, 2026Open, Classified
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.
Z-2727-2026 · Root cause: Process control · FDA record ↗
- Nov 25, 2024Open, Classified
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work properly causing the arm with X-ray tube and collimator to drop downward on the column track.
Z-0812-2025 · Root cause: Component design/selection · FDA record ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K142383. One submission often covers many models, so check the product and lot details in each record.
- Z-2726-2026 · Jun 5, 2026 · Open, ClassifiedOptima XR200amx, Mobile Digital-Ready Radiographic System
- Z-0811-2025 · Nov 25, 2024 · Open, ClassifiedGE Healthcare Optima XR200amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-3; 2) 5555000-4: 3) A0659MB
- Z-0770-2016 · Dec 23, 2015 · TerminatedGE Healthcare, Optima XR200amx, Mobile Digital Ready Radiographic System Designed to generate diagnostic radiographic images.
- Z-0768-2016 · Dec 23, 2015 · TerminatedGE Healthcare, Optima XR220amx, Mobile Digital Ready Radiographic System Designed to generate diagnostic radiographic images.
- Z-1406-2015 · Dec 19, 2014 · TerminatedGE Healthcare Automatic Mobile X-Ray (AMX) Series: Optima XR200 amx
- Z-1407-2015 · Dec 19, 2014 · TerminatedGE Healthcare Automatic Mobile X-Ray (AMX) Series: Optima XR220 amx
Other System, x-ray, mobile devices
- Philips MobileDiagnost wDR 2.2Sedecal · System, X-Ray, Mobile3 recalls7 reports
- DRX-Revolution Mobile X-ray SystemCarestream Health, Inc. · System, x-ray, mobile2 recalls12 reports
- Optima XR200amxGE Medical Systems, L.L.C. · System, x-ray, mobile2 recalls1 report
- Optima XR240amxGE Medical Systems, L.L.C. · System, x-ray, mobile2 recalls0 reports
- DRX-Revolution Mobile X-ray SystemCarestream Health, Inc. · System, x-ray, mobile1 recall16 reports
- SEDECAL Soltus 500Sedecal · System, X-Ray, Mobile1 recall4 reports
- MobileDaRt Evolution 503-79999-01Shimadzu Corporation · System, X-Ray, Mobile1 recall0 reports
- FDR AQRO DR-XD 1000Fujifilm Corporation · System, x-ray, mobile1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.